Fiber Collimators Series 60FC-LSA
Fiber Collimators Series 60FC-LSA
This document provides assistance in installing the fiber collimators series 60FC-LSA. It describes how a fiber
cable is attached and how the collimation setting is adjusted.
1.
Before You Start
The laser beam coupler is shipped with a protection cap for the fiber receptacle and with a front cap.
Notice:
There is a polymer cap on the receptacle. Please perform the following steps in order to remove this cap:
Now, make sure that the connector of the fiber you want to attach matches the receptacle type of the laser
beam coupler:
• Use fibers with APC (8°-polish) connectors for laser beam couplers with an inclined coupling axis
(60SMF-LSA-4).
• Use fibers with PC (0°-polish) connectors for laser beam couplers with a coaxial axis
(60SMF-LSA-0).
The laser beam couplers are compatible to all fiber connectors of type LSA, DIN, AVIO and AVIM.
Notice:
Adjustment_60FC-LSA
20.05.2021
Schäfter+Kirchhoff GmbH Kieler Str. 212 22525 Hamburg, Germany
Please remove all the protection caps first and do not use them as beam
dumps (risk of photo contamination).
• Do not touch either the optical surface of the lens or the fiber end-face.
• If the coupler is not in use, reattach both rear and front protection caps.
© Schäfter+Kirchhoff GmbH
Figure 1:
First, remove the threaded cap from
the fiber receptacle of the fiber
collimator.
Figure 2:
Then, slightly loosen the small pin
screw. Use the screwdriver 9D-12.
Make sure to not loosen it too far, as it
is small and easily lost.
+49 40 85 39 97-0 info@sukhamburg.de www.sukhamburg.com
1
Need help?
Do you have a question about the 60FC-LSA Series and is the answer not in the manual?
Questions and answers