Noise Sensor V3.0/NBL-W-NS
Product Description
Noise transducers are complex and sophisticated
devices that enable precision measurements of
sound. This instrument has a durable construction
and will last for several years of normal use when
operated correctly. Please read this manual carefully
and do not lose it.
Technical parameters
Measurement range: Range: 30~130dB.
Output: 485 output
Accuracy (23±5°C):
Frequency weighting in accordance with IEC 61672 type 2,
calibrated with an input signal of 94dB (31.5Hz - 8kHz).
Frequency range: 31.5Hz to 8kHz.
Time response: Fast response: T = 200ms.
The time response simulates the time response of the
human ear.
http://www.niubol.com
Calibrator: B&K (Bruel & kjaer), Multifunction Sound
Calibrator, type 4226.
Microphone: Condenser microphone.
Microphone size: 0.5 inch.
Range Selection: 30~130dB
Maximum output impedance: 200 ohms.
Power Supply: DC 12V
Power Loss: DC 12V: ≈ 20mW
Packaging: ABS plastic
Operating temperature: -15-50°C (5-122°F).
Operating humidity: <80% relative humidity.
Wiring method
(1) If equipped with our collector, directly use the
sensor cable to connect the sensor to the
corresponding interface on the collector. cable to
connect the sensor to the corresponding interface on
the collector.
(2) If the transmitter is purchased separately, the
wiring sequence of the matching wires of the
transmitter are as follows for:
Thread
Colour
Red
Black
(Green)
Yellow
Blue
1
output signal
Communicatio
n Type
Power Positive
Power Ground
A+/TX
B-/RX

Advertisement

loading

Summary of Contents for NiuBol 3.0/NBL-W-NS

  • Page 1 94dB (31.5Hz - 8kHz). Black Power Ground Frequency range: 31.5Hz to 8kHz. (Green) Time response: Fast response: T = 200ms. A+/TX Yellow The time response simulates the time response of the B-/RX Blue human ear. http://www.niubol.com...
  • Page 2 1 、 Preset the 16-bit register as hexadecimal FFFF Return 00 10 00 7C (i.e. all 1s). Call this register as CRC register; [2] Read device address 2, put the first 8-bit data and the low bit of the Send: 00 20 CRC (4 bytes) http://www.niubol.com...
  • Page 3 Please check whether the package is in good Email:sales@niubol.com condition, and check whether the product model is Website: http://www.niubol.com consistent with the selected model; Address: Room 102, District D, Houhu Industrial 2 、 Do not carry electricity wiring, wiring is Park, Yuelu District, Changsha City, Hunan completed to check for errors before power;...