2420 FIELD HOSPITAL SURGICAL LIGHT
7. Appendix
o
APPENDIX D
PERIODIC MAINTENANCE ADJUSTMENTS AND PROCEDURES
Adjustments and Procedures
Perform all CHECKS and ADJUSTMENTS in Section 1 of this manual once a year or when the equipment is
returned to active use after a significant period of inactivity. ALWAYS perform these Checks and Adjustments
IN THE ORDER SHOWN
IN THE MANUAL. Random adjustments may defeat other adjustments.
Special Equipment and Tools Required
||
Screwdrivers:
Standard
Phillips
11
Wrenches:
Open End
Allen - 0.050" to 5/16"
Adjustable
Socket
||
Pliers:
Standard
Needle Nose
@
No. 4 External Snap Ring
Wire Cutters and Stripper
Pocket or Skill Knife
Wire Terminal Crimp
Light Meter
White Cardboard (or other white surface for illumination tests)