2420 FIELD HOSPITAL SURGICAL LIGHT
6. Repair
NOTE:
FILE TOP AND BOTTOM CORNERS
ON THIS SIDE TO PROVIDE COVER
CLEARANCE.
OUTLET BOX.
CONTROLLER SHAFT
-
e
43734
CONTROLLER SPACERS
VENT HOLES
CIRCUIT SWITCH, DPST. 7.0 AMP
60908 SHAFT ADAPTER
Figure 23. 120V INTENSITY CONTROL
WARN
the light or Intensity Control to avoid
hazardous electrical contact.
PLUG
sai
тт.
1.
Remove
the
stopscrew
and
back
off friction
ΟΠΩΝ
plug setscrew.
RETAINING
RING
2.
Refer to Figure 26. Disconnect and remove wir-
ee
ing from extension arm.
m
3.
Remove the bearing post retaining ring to lower
ao
the support arm from the extension arm.
EXTENSION
ος
ος NASSER
4.
Wipe bearing post clean and apply athin coat of
POHON B
grease.
5.
Reassemble
the
arms
in
reverse
order
of
BEARING
4
WASHER
disassembly.
BEARING
6.
Use retaining ring pliers to insert the retaining
ring into the groove in the bearing post.
POST
retaining ring is seated and secure in the post
SM
Y
aroove.
.
PLUG
O
A
A loose or unseated retaining ring could
BUTTON
CAUTION]
cause the stopscrew to loosen and allow
the support arm to disengage.
Figure 24. REMOVING
SUPPORT ARM