Industrial physics sheen DV2300 Operating Instructions Manual

Krebs viscometer

Advertisement

Quick Links

Krebs Viscometer
DV2300
Operating Instructions
(V3.0 072023)

Advertisement

Table of Contents
loading

Summary of Contents for Industrial physics sheen DV2300

  • Page 1 Krebs Viscometer DV2300 Operating Instructions (V3.0 072023)
  • Page 2 Manufacturer | COPYRIGHT © 2023 - Sheen www.industrialphysics.com Sheen is a brand of Industrial Physics Inc.
  • Page 3 START - NOTE NOTE Please check the startup screen of the Krebs viscometer to check if this manual, written for version 1.0.8.0 is applicable to you.
  • Page 4 INDEX GENERAL 1.1 Importance of operating manual 1.2 User-responsibility 1.3 Responsibility of personnel 1.4 Dangers 1.5 Designated purpose 1.6 Copyright 1.7 Manufacturer’s/Supplier’s address SAFETY INSTRUCTIONS 2.1 Meaning of symbols 2.2 Availability of safety information 2.3 Training of personnel 2.4 Power supply 2.5 Special attention 2.6 Care, maintenance and repairs 2.7 Modifications to the equipment...
  • Page 5: Table Of Contents

    SCOPE OF SUPPLY Technical Data DV2300 Sheen Krebs viscometer ASSEMBLY MENU EXPLANATION 7.1 General menu structure 7.2 Main menu 7.3 Actual measurement 7.4 Clean high - low 7.5 Setup menus 7.5.1 Measurement type 7.5.2 Units 7.5.3 Calibration 7.5.4 Display FREQUENTLY ASKED QUESTIONS CARE AND MAINTENANCE Inspection and Maintenance 9.2 Customer Service...
  • Page 6 The design is setup for use with a Sheen spindle and is carefully calibrated to conform to ASTM D562. Other applications constitute improper use. Industrial Physics will not be held liable for damage resulting from improper use.
  • Page 7 1.6 Copyright The copyright of this operating manual remains with Industrial Physics. Its instructions and guidelines may not be duplicated, circulated or otherwise passed on to others, neither fully, nor partly, without prior permission. Infringement of these restrictions may lead to legal action to be taken.
  • Page 8 2.3 Training of personnel • Anyone who operates the instrument should be trained properly. • It has to be clear who has which responsibility regarding commissioning, maintenance and repairs, installation and operation. • Anyone who hasn’t finished training should be supervised by an experienced person while working with the instrument.
  • Page 9 2.7 Modifications to the equipment • Any modifications, additions or alterations to the instrument may solely be made with permission from Industrial Physics. • All measures involving modifications require written confirmation of approval from Industrial Physics. • Instruments which are not in fault-free condition must immediately be switched off.
  • Page 10: Scope Of Supply

    4 SCOPE OF SUPPLY • DV2300 - Sheen Digital Krebs viscometer • One 0,5l can • Handle • 110 - 250VAC power adapter • Standard Krebs spindle DV2305 Note: Calibration verification oils have to be ordered separately. Standard Krebs spindle DV2305...
  • Page 11: Technical Data Dv2300 Sheen Krebs Viscometer

    5 TECHNICAL DATA DV2300 SHEEN KREBS VISCOMETER Operating voltage 100-240 VAC / 50-60 Hz from a grounded wall outlet Operating temperature +5 °C - +40 °C / +41 °F – 104 °F Net weight 8,5 kg / 18,7 lbs Speed 200 rpm Range 52 –...
  • Page 12: Assembly

    6 ASSEMBLY 1. Install the handle. 2. Install the power supply. 3. Install the Krebs spindle . - The spindle is keyed and magnetic for easy installation and removal after taking a measurement. - The base of the instrument can be screwed to the tabletop for extra stability. ON/OFF Power input Stop the chuck from rotating.
  • Page 13: Menu Explanation

    7 MENU EXPLANATION This section describes in detail all the menu options the instrument offers. 7.1 General menu structure Below is a high level overview of the Krebs viscometer menu structure. The menu is navigated using the buttons as indicated.
  • Page 14: Main Menu

    7.2 Main menu Manual MAIN MENU After switching on the unit, this is the main menu. The machine will revert to the settings used the Lower Viscometer previous time. and press START START CLEAN SETUP 7.3 Actual measurement Place a can with the fluid to be measured on the base of the instrument making sure it is well centered and magnetically secured.
  • Page 15 In this example Manual measurement was selected previously and it’s the operator that has to watch the duration of the measurement and press the OK button again to stop the spindle from turning. (see section 7.5.1 for measurement type selection) Follow your local instructions for the time to Manual MEASURE...
  • Page 16: Clean High - Low

    7.4 Clean High - Low In the main menu, use the control buttons to highlight CLEAN and press the OK button. CLEAN is just the spindle turning in an appropriate solvent, without measurements being made. Next highlight Low speed or High speed and push the OK button.
  • Page 17: Setup Menus

    7.5 Setup menus The menus available for setting up the instrument’s behavior are explained in this chapter. In the main menu, use the control buttons to high- light SETUP and press the OK button. 7.5.1 Measurement type SETUP Measurement type The Measurement type selection will allow the user Units to select either Manual, Max hold or Timed measure-...
  • Page 18 MEASUREMENT TYPE Max hold will show the selected unit with a large maximum and smaller current value shown. [ ] Manual [ √] Max hold Press the OK button to select and then navigate to [ ] Timed [0010]s BACK and press the OK button to return to the main window.
  • Page 19: Units

    Here is the now fixed, timed measurement and continuous current measurement in the unit selected. To perform a new measurement press the OK button. 7.5.2 Units SETUP Measurement type Units In the main SETUP menu, navigate to Units and Calibration press the OK button.
  • Page 20: Calibration

    A separate calibration manual will be made available on request. Your Krebs viscometer can be recalibrated at the Industrial Physics factory and selected service centers and distributors. Using calibration oils with certificates, calibrated thermometers with 0,01 °C reading capability and 0,1 °C accuracy at 25 °C, in a well controlled environment, following ASTM D562 will allow a user to...
  • Page 21: Frequently Asked Questions

    0.1 °C calibrated thermometers, a warm bath and in-depth viscosity knowledge. Q: Which oils are appropriate for instrument verification ? A: The Industrial Physics oils suitable for Krebs viscometer verification are: DV2205, DV2206,DV2207, DV2208, DV2209, DV2210. Also see: https://industrialphysics.com/product/viscositycalibration-oils/...
  • Page 22: Care And Maintenance

    Do not allow any paint to dry on the spindle. 9.2 Customer Service When requesting service always include Model No. and Serial No. from device tag. Customer service is provided on request by: Industrial physics Molenbaan 19 2908 LL Capelle aan den IJssel The Netherlands info@industrialphysics.com...
  • Page 23: Disclaimer

    The information given in these instructions may be incomplete. Anyone using the product for a purpose other than that described in this document does so at his/her own risk. Sheen is a brand of Industrial Physics Inc.
  • Page 24 www.industrialphysics.com...