Summary of Contents for Salter Labs PRO2 Check

  • Page 1 CAUTION 30 Spur Dr. El Paso, TX 79906 Phone: 1.800.421.0024 Email: customercare@salterlabs.com www.salterlabs.com...
  • Page 2 30 Spur Dr El Paso, TX 79906 Phone: 1.800.421.0024...
  • Page 5 Purchaser will be responsible for all shipping charges to return the unit to Salter Labs. Salter Labs will be responsible for normal shipping charges for the return of the unit to purchaser. ®...
  • Page 6: Table Of Contents

    Full "Two Step" Calibration Procedure VII. Maintenance Battery Replacement Cleaning VIII. Environmental Effects Temperature Barometric Pressure Vibration Relative Humidity Unit Specifications Frequently Asked Questions 11 - 12 Trouble Shooting Check List XII. Warranty Manufactured by: ® - El Paso, Texas Salter Labs...
  • Page 7: Warning Notes

    ® check Ultrasonic Oxygen Indicator WARNING NOTES The PRO ® USAGE: check Ultrasonic Oxygen Indicator is designed for intermittent usage to determine the oxygen concentration produced by an oxygen concentrator. It may also be utilized to do spot checks of gaseous or liquid oxygen.
  • Page 8 "LO BATT" appears in display Battery is low. Change battery. screen when unit is turned on Unit may have been Call Salter Labs' Display shows "err" used to check a Customer Service or a reading other concentrator with a Department at than 100% ±...
  • Page 9 ® check Ultrasonic Oxygen Indicator Trouble Shooting Checklist Suggested Complaint Probable Cause Corrective Action Bad batteries. Display is blank when Replace batteries. Corroded battery unit is turned on. Clean battery terminals. terminals. When unit is turned on Perform "Full Unit requires display flashes Calibration Procedure"...
  • Page 10: Introduction

    Indicator. By relying on a method already used in many oxygen concentrators to trigger "Low O " alarms – ultrasound – Salter Labs has produced a more cost effective product to check the output of oxygen concentrators. The check ®...
  • Page 11: Initial Setup

    ® check Ultrasonic Oxygen Indicator INITIAL SETUP check ® Remove the Ultrasonic Oxygen Indicator from the package. Make sure you have the following items: check ® • indicator unit • 9-volt battery (factory installed) • 2 foot connecting tubing • Carrying case and strap •...
  • Page 12 ® check Ultrasonic Oxygen Indicator FREQUENTLY ASKED QUESTIONS (continued) When I am in the Concentrator Mode and hook the unit up to a U.S.P. oxygen(> 99.0%) source the LCD does not read 99% (± 2), why? When the unit is in its default (concentrator) mode the device's software enables the unit to measure the output of an oxygen concentrator by compensating for the argon that passes through the sieve beds.
  • Page 13 ® check Ultrasonic Oxygen Indicator FREQUENTLY ASKED QUESTIONS ® check Thank you for purchasing the Ultrasonic Oxygen Indicator. It will give you years of trouble-free operation. The following are some of the frequently asked questions about the operation of this device. ®...
  • Page 14: Instructions For Checking O

    ® check Ultrasonic Oxygen Indicator INSTRUCTIONS FOR CHECKING O CONCENTRATORS To turn unit on and off: To turn the unit on or off press and release the “on/off” button on the lower left-hand side of the unit. (EXAMPLE: Concentrator O 20.9) When the unit is turned on the device defaults to the mode in which the software enables the unit to measure the output of an oxygen concentrator compensating for the argon that passes through the...
  • Page 15: Instructions For Spot Checking Liquid Oxygen Vessels Or Oxygen Cylinders

    ® check Ultrasonic Oxygen Indicator INSTRUCTIONS FOR SPOT CHECKING LIQUID OXYGEN VESSELS or OXYGEN CYLINDERS check ® Although the Ultrasonic Oxygen Indicator is designed, as its name implies, to check the various functions of an oxygen concentrator, it can also be used to perform spot checks (± 2%) of gaseous or liquid oxygen.
  • Page 16 ® check Ultrasonic Oxygen Indicator UNIT SPECIFICATIONS General Specifications Physical Data Dimensions: 3.60" W x 5.75" H x 1.29" D (9.14 cm W x 14.6 cm H x 3.28 cm D) Weight: 9 ounces (255.15 Grams) Connector: ¼" anodized aluminum hose barb Case: Solid color ABS plastic with threaded metal bolt channels Battery: The battery capacity is a standard 9-volt "transistor"...
  • Page 17 ® check Ultrasonic Oxygen Indicator ENVIRONMENTAL EFFECTS Temperature Under normal operating conditions 32º F to 105º F (0º C to 41º C), the check ® unit's microprocessor and temperature circuit will compensate for variations in temperature. The temperature during storage and shipping should not reach below -29º...
  • Page 18: Verification & Calibration Instructions

    ® check Ultrasonic Oxygen Indicator VERIFICATION & CALIBRATION INSTRUCTIONS PURE OXYGEN VERIFICATION PROCEDURE To quickly verify that the indicator unit is accurately measuring the concentration of oxygen being produced by a concentrator, perform the following steps: ® check 1. Connect the inlet port on the unit to a USP oxygen source (cylinder or liquid oxygen) via a supply tube.
  • Page 19: Full "Two Step" Calibration Procedure

    2 LPM, then repeat procedure. NOTE: If unit displays "---" during verification or calibration procedures, release the "pure O2/calibrate” button or the unit will shut off automatically. If any problems are encountered during the calibration sequences, contact Salter Labs at 1-800-421-0024 or 1-661-854-3166.
  • Page 20: Maintenance

    ® check Ultrasonic Oxygen Indicator MAINTENANCE Battery Replacement When the LCD displays "LO BATT" in the lower left corner, the battery is low and should be replaced. check ® NOTE: The must be removed from the carrying case to replace battery. Install a new 9-volt battery by sliding the battery door (located on the rear of the unit) away from the case.