Control system for automatic cleaning of ph electrodes (33 pages)
Summary of Contents for Endress+Hauser Smart System for Surface Water SSP100B
Page 1
Products Solutions Services BA02044S/04/EN/01.20 71484905 2020-06-01 Operating Instructions Smart System for Surface Water SSP100B Package with smart sensors, for measuring the water quality of rivers or lakes...
Page 2
Revision history Smart System for Surface Water SSP100B Revision history Product version Operating Instructions Changes Comments 1.00.XX BA02044S/04/EN/01.20 Initial version Replaces BA01929S Endress+Hauser...
About this document Smart System for Surface Water SSP100B About this document Document function These instructions provide all the information required to use the system: from the product description, installation and use to system integration, operation, diagnostics and troubleshooting to software updates and disposal.
Smart System for Surface Water SSP100B About this document 1.2.3 Electrical symbols Symbol Meaning Direct current Alternating current Direct current and alternating current Ground connection A grounded terminal which, as far as the operator is concerned, is grounded via a grounding system.
Measuring cables CYK10 • Technical Information TI00432C/07/EN • Operating Instructions BA00432C/07/EN For additional documentation, please refer to the Endress+Hauser Operations App or go to www.endress.com/device-viewer. Registered trademarks Modbus is the registered trademark of Modicon, Incorporated. RUT240 is a product of Teltonika Ltd., 08105 Vilnius/Lithuania.
Page 7
Smart System for Surface Water SSP100B About this document RevPi Core 3 is a product of Kunbus GmbH, 73770 Denkendorf/Germany. UNO PS is a product of Phoenix CONTACT GmbH & Co. KG, 32825 Blomberg/Germany. All other brand and product names are trademarks or registered trademarks of the companies and organizations in question.
The personnel for installation, commissioning, diagnostics and maintenance must meet the following requirements: ‣ Trained, qualified specialists: must have a relevant qualification for this specific role and task and have been trained by Endress+Hauser. Experts at the Endress+Hauser service organization. ‣ Personnel must be authorized by the plant owner/operator.
2.4.1 Modifications to the system Unauthorized modifications to the system are not permitted and can lead to unforeseeable dangers: ‣ If, despite this, modifications are required, consult with Endress+Hauser. 2.4.2 Repair To ensure continued operational safety and reliability: ‣ Repairs must be performed only by certified Endress+Hauser specialists.
Smart System for Surface Water SSP100B Product description Function The Smart System for Surface Water SSP100B monitors surface water. The package includes all the necessary sensors and a transmitter for measured data processing and the Modbus Edge Device SGC400. The package also contains immersion assemblies, mounting plates and the connecting cable for Modbus TCP connection.
• Keep all the accompanying documents. The system may not be put into operation if the contents are found to be damaged beforehand. Contact your Endress+Hauser Sales Center if this happens. Return the system to Endress+Hauser in the original packaging where possible.
Incoming acceptance and product identification Smart System for Surface Water SSP100B Storage and transport • The components are packed in such a way that they are fully protected against shock when in storage and during transportation. • The permitted storage temperature is 0 to 40 °C (32 to 104 °F).
Smart System for Surface Water SSP100B Installation Installation Installation conditions 5.1.1 IP degree of protection NOTICE IP degree of protection not observed Possible damage to the device. ‣ Take the IP protection of Liquiline CM444 and the Modbus Edge Device SGC400 into consideration for the mounting location.
Installation Smart System for Surface Water SSP100B ≥ 340 (13.4) A0027798 4 Mounting the mounting plate A0012686 Wall 3 Necessary installation clearance in mm (in) Boreholes (x 4) Mounting plate Screws 6 mm (not included in the delivery)
Smart System for Surface Water SSP100B Installation A0025890 A0025891 7 Disassembly – Steps 1 and 2 8 Disassembly – Step 3 1. Push back the catch. 2. Push up the device to remove it from the holder. 3. Remove the device towards the front.
Installation Smart System for Surface Water SSP100B Assembling the immersion assembly 1. CPF81D and CLS50D: screw the sensor adapter onto the connecting bracket of the immersion tube. 2. Push the weight into the immersion tube. 3. Mount the retaining bracket on the multifunctional clamp ring.
Smart System for Surface Water SSP100B Electrical connection Electrical connection Safety WARNING Transmitter and Modbus Edge Device are live Incorrect connection may result in injury or death! ‣ Only certified electricians may perform the electrical connection. ‣ The certified electrician must read and understand these Operating Instructions and must follow all the instructions.
Electrical connection Smart System for Surface Water SSP100B Cable entries and terminals A0018025 12 Underside of housing with labeled cable entries and terminals Labeling on underside of housing Description for Liquiline CM444 Description for Modbus Device Edge 1 to 8...
Smart System for Surface Water SSP100B Electrical connection 6.4.1 Cable mounting rail A0025171 13 Cable mounting rail and associated function Cable mounting rail Threaded bolt as protective ground connection, central grounding point Additional threaded bolts for ground connections Cable clamps for fixing and grounding the sensor cables 6.4.2...
Electrical connection Smart System for Surface Water SSP100B A0025812 16 Protective ground or grounding connection NOTICE Protective ground or ground cable with wire ferrule or open cable lug The cable can become loose. Loss of the protective function! ‣...
Smart System for Surface Water SSP100B Electrical connection 2. Connect the measuring cable of conductivity sensor CLS50D to connection port 2. 3. Connect the measuring cable of pH sensor CPF81D to connection port 3. Electrical connection of Modbus Edge Device 6.5.1...
Page 22
Electrical connection Smart System for Surface Water SSP100B Mounting the LTE antenna 1. Select a suitable mounting location for the antenna head. 2. Disconnect the antenna connection cables on the LTE modem in the Modbus Edge Device. 3. Pull the antenna connection cables through the cable glands on the Modbus Edge Device.
Smart System for Surface Water SSP100B Electrical connection 6.5.3 Connecting the supply voltage for the Modbus Edge Device, 100 to 240 V AC Output DC 24V 30W DC OK Input AC 100V-240V A0037821 20 Connecting supply voltage 100 to 240 V AC...
Electrical connection Smart System for Surface Water SSP100B 6.5.4 Connecting the supply voltage for the Modbus Edge Device, 24 V DC 0V 24V A0042707 21 Connecting supply voltage 24 V DC Terminal blocks Cable entry Power supply cable 1. Route the power supply cable into the housing through the cable gland at connection port G.
Smart System for Surface Water SSP100B Electrical connection A0025985 22 Connecting the Ethernet cable to the transmitter ‣ Connect the Ethernet cable to the M12 socket of connection port I on the transmitter. Post-connection check Are the system, devices and cables undamaged (visual check)? ...
The device can now be operated in your chosen language. 7.1.3 Configuring the Liquiline CM444 The Liquiline CM444 transmitter is preconfigured on delivery from Endress+Hauser. Take the following steps if you want other settings or want to connect an additional sensor.
Page 27
Smart System for Surface Water SSP100B Commissioning Modifying the Modbus assignment Each sensor is assigned one analog output for the primary value and one analog output for the secondary value. The sensors are assigned to the analog outputs according to the connection on the transmitter.
Commissioning Smart System for Surface Water SSP100B 9. Check the settings in the System information menu. Menu path: DIAG > System information The Liquiline CM444 transmitter can be found by the Modbus Edge Device and read out data. Commissioning the Modbus Edge Device No settings need to be made on the Modbus Edge Device.
Diagnostics and troubleshooting Smart System for Surface Water SSP100B Diagnostics and troubleshooting Smart System SSP For an overview of the alarms currently pending, see the "History" view in the Smart System App. Liquiline CM444 For detailed information on transmitter diagnostics and troubleshooting, see the Operating Instructions for Liquiline CM444 .
For more information, see the next section. 10.1 Endress+Hauser services Endress+Hauser offers a wide variety of services for maintenance such as recalibration, maintenance service, system tests or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services.
11.1 General information Please note the following: • Repairs may only be performed by Endress+Hauser staff or by individuals authorized and trained by Endress+Hauser. • Observe local and national laws and regulations. • Standard components can be replaced with identical components.
Smart System for Surface Water SSP100B Technical data Technical data For detailed information on the technical data, see the Technical Information for the particular product → 6. Endress+Hauser...