PCB Piezotronics 422E03 Installation And Operating Manual

In-line charge converter
Hide thumbs Also See for 422E03:

Advertisement

Quick Links

Model 422E03
In-Line Charge Converter
Installation and Operating Manual
For assistance with the operation of this product,
contact PCB Piezotronics, Inc.
Toll-free: 800-828-8840
24-hour SensorLine: 716-684-0001
Fax: 716-684-0987
E-mail: info@pcb.com
Web: www.pcb.com

Advertisement

Table of Contents
loading

Summary of Contents for PCB Piezotronics 422E03

  • Page 1 Model 422E03 In-Line Charge Converter Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
  • Page 2 Returning Equipment – Following the equipment to PCB Piezotronics for these procedures will insure that your repair. User servicing or repair is not returned materials are handled in the recommended and, if attempted, may most expedient manner.
  • Page 3 PCB Piezotronics, Inc. Warranty – All equipment and repair 3425 Walden Ave. services provided by PCB Piezotronics, Depew, NY 14043 USA Inc. are covered by a limited warranty Toll-free: (800) 828-8840 against defective...
  • Page 4 Total Customer Satisfaction Guarantee The Model 422E In-Line Charge Converter Operating Guide with Enclosed Warranty Information 3425 Walden Avenue, Depew, New York 14043-2495 Phone 716 684-0001 716 684-0987 E-Mail electronics@pcb.com Toll Free Line 1-800-828-8840 24 Hour Sensor Line 716 684-0001 MANUAL NUMBER: 18882 MANUAL REVISION: A ECN: 24741...
  • Page 5: Table Of Contents

    Total Customer Satisfaction Guarantee table of contents Introduction........................page 3 Description ........................page 3 Installation ........................page 3 Operation.......................... page 3 Calculation of Output Scaling ................... page 4 TEDS Functionality......................page 4 Special Considerations..................... page 5 422E Series Models for Use with High Temperature Sensors.......... page 5 422E Series Models for Use in Radioactive Environments..........
  • Page 6: Introduction

    Total Customer Satisfaction Guarantee introduction The 422E series in-line charge converters are designed to convert the high impedance of a charge mode piezoelectric transducer to a low-impedance voltage by means of an ICP ® signal conditioner. These units may be used with either quartz or ceramic charge-mode piezoelectric sensors. description The 422E series in-line charge amplifiers operate from an ICP ®...
  • Page 7: Calculation Of Output Scaling

    Total Customer Satisfaction Guarantee calculation of output scaling The 422E contains a high-gain, low-noise amplifier connected in a charge amplifier configuration. The charge output of the transducer is transferred to the feedback capacitor of the amplifier to develop a voltage which may be calculated using the following equation. = 422E Scaling (mV/pC) x Transducer Sensitivity (pC/Engineering Unit) where: = Output of 422E in mV/Engineering Unit...
  • Page 8: Special Considerations

    For example, the level for a 422E02/E12 would be 10,000/10 pF = 1,000 pF, and the level for a 422E03/E13 would be 10,000 pF. If PCB 003 type low noise cable is used, these values correspond to 10 meters and 100 meters of length respectively.
  • Page 9: 422E Series Models For Use In Radioactive Environments

    2 – ESD sensitivity This equipment is designed with user safety in mind; however, the protection provided by the equipment may be impaired if the equipment is used in a manner not specified by PCB Piezotronics, Inc. caution 1 – ESD sensitivity Cables can kill your equipment.
  • Page 10: Warranty, Service & Return Procedure

    Total Customer Satisfaction Guarantee caution 2 – ESD sensitivity ESD considerations should be made prior to performing any internal adjustments on the equipment. Any piece of electronic equipment is vulnerable to ESD when opened for adjustments. Internal adjustments should therefore be done ONLY at an ESD-safe work area. Many products have ESD protection, but the level of protection may be exceeded by extremely high voltage.
  • Page 11: Customer Service

    Total Customer Satisfaction Guarantee Customers outside the U.S. should consult their local PCB distributor for information on returning equipment. For exceptions, please contact the International Sales department at PCB to request shipping instructions and an RMA. For assistance, please call (716) 684-0003, or fax us at (716) 684- 3823.