Sleep Fragmentation Chamber
User's Manual
3700 Sagamore Pkwy N
Lafayette, IN 47904 USA
Tel: (765) 423-1505
info@lafayetteinstrument.com
www.lafayetteinstrument.com
Fax: (765) 423-4111 . www.lafayetteinstrument.com . info@lafayetteinstrument.com
Rel. 8.10.23
© 2023 by Lafayette Instrument Company, Inc. All Rights Reserved-

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MAN733 and is the answer not in the manual?

Questions and answers

Summary of Contents for Lafayette Instrument MAN733

  • Page 1 Model 80391 User’s Manual Sleep Fragmentation Chamber User’s Manual 3700 Sagamore Pkwy N Lafayette, IN 47904 USA Tel: (765) 423-1505 info@lafayetteinstrument.com www.lafayetteinstrument.com Fax: (765) 423-4111 . www.lafayetteinstrument.com . info@lafayetteinstrument.com Rel. 8.10.23 © 2023 by Lafayette Instrument Company, Inc. All Rights Reserved-...
  • Page 2 • External Initiate/Event Output: 2.5mm connection, 3 conductor • Base: Ground • Ring: Event output, Open collector, Pull up if needed (30 VDC max) • Tip: Ext Initiate, active low switch closure (30 VDC max) Contact Lafayette Instrument for more information • Dimensions: •...
  • Page 3 User’s Manual Operating Instructions The chamber, including cover and sweeper arm, may be removed from the drive mechanism for cleaning and animal husbandry. 1. Orient the chamber on the base so the sweep arm carriage is facing the drive mechanism. 2.
  • Page 4 A momentary pulse of 20ms will initiate a single sweep of the sweeper arm. A mating connector or cable is not provided. Contact Lafayette Instrument Co. for more information. Event Mark: A 50ms active low signal is generated at the start of each sweep. Connection may be made through the same 2.5 mm jack used for an external initiate.
  • Page 5 User’s Manual Polycarbonate Cleaning Instructions The life of polycarbonate can be prolonged by following these Usage/Cleaning Instructions. Care Contact between animal waste and cages should be kept to a minimum by using a sufficient quantity of an absorbent bedding material and replacing it with fresh material as often as possible.
  • Page 6 Model 80391 Sleep Fragmentation Chamber 2. Use as short of a cycle as possible. Exposure of 15 minutes at 15-17psi and 250 F is recommended. 3. Do not stack cages in heights exceeding 30 inches. 4. Effective steam sterilization depends upon proper temperature controls and an adequate steam supply.
  • Page 7 User’s Manual NOTE: Small adjustments in the tension screws result in large changes in sweep arm force. Typically the tension screws should never need to be tightened beyond finger tight. Do not increase the sweep arm force beyond the drive motor capability, which is indicated by severe vibration and buzzing sound.
  • Page 8 Polygraph system to any other country without an export license. If the ultimate call Lafayette Instrument to receive a RGA number. Your RGA number will be destination is not one of the above listed countries, contact us for the required good for 30 days.