Cleaning, Disinfection And Sterilization Precautions - Hitachi ASU-1014 Instruction Manual

Electronic convex sector scanner
Table of Contents

Advertisement

Scan for the minimum length of time necessary for the diagnosis and at the lowest suitable output.
Overuse can adversely affect the internal tissues of the patient.
For details about the acoustic output, please refer to the documentation supplied with the ultrasound
diagnostic instrument.
Regularly perform maintenance inspection and safety tests of the ultrasound diagnostic instrument
and scanner.
If you use equipment for a long period of time, it can reduce the performance, or cause smoke or
fire. If anything unusual occurs, immediately stop using it and contact one of our offices and/or
distributor's offices listed on the back cover.
Use, move and transport the scanner under the environmental conditions specified in this manual.
Otherwise, it may be damaged.
See section 2-5 "Environmental conditions" and section 7-4 "Environmental conditions during
transportation".
1-2-2. Cleaning, disinfection and sterilization
Warning
Wear protective gloves and other protective gear during cleaning, disinfection and sterilization.
Handling of the scanner with your bare hands before disinfection can result in an infection.
After soaking in cleaning agents, thoroughly wash the scanner with running water.
Residual cleaning agents can cause an adverse reaction on the bodies of the operator or patient.
After soaking in a disinfectant, throughly wash the scanner with deionized water. Leavings of the
disinfectant can cause an adverse reaction on the bodies of the operator or patient.
Perform aeration completely after gas disinfection.
Residual gas can cause an adverse reaction on the bodies of the operator or patient.
Do not clean, disinfect or sterilize using procedures other than those specified in this manual.
Infection could result due to incomplete cleaning, disinfection and sterilization. It can also result in
damage to the scanner or reduced performance. The scanner cannot withstand autoclave sterilization
or boiling and other types of sterilization at temperatures exceeding 50°C (122°F).
For details on the usage conditions of chemicals procedures, refer to the documentation supplied with
the respective chemical equipment.
Infection could result due to incomplete disinfection. This could also cause deterioration of the
scanner.
Caution
Do not immerse the probe in any liquids beyond the range of IPX7 shown in section 2-2 "Specifications".
Use when liquid has gotten inside the connector can result in a risk of electric shock to the operator
or patient. If liquid gets inside the connector, immediately stop use and contact one of our offices and/
or distributor's offices listed on the back cover.
precautions
-3-
MN1-5933 Rev.5

Advertisement

Table of Contents
loading

Table of Contents