Model 1203-15A LOW PROFILE LOAD CELL Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Page 2
Piezotronics, user servicing or repair is critical test. not recommended and, if attempted, may void the factory warranty. Routine PCB Piezotronics maintains an ISO- maintenance, such as the cleaning of 9001 certified metrology laboratory and electrical connectors, housings, and offers calibration services, which are...
Page 3
50% of the general contact numbers are: replacement cost returned item(s). PCB will provide a price PCB Piezotronics, Inc. quotation replacement 3425 Walden Ave. recommendation for any item whose Depew, NY14043 USA repair costs would exceed 50% of...
GENERAL PURPOSE LOW PROFILE LOAD CELL OPERATION MANUAL 3.1 Dimensions 1.0 INTRODUCTION The following figure and tables give the general outline General purpose low profile load cells manufactured by PCB dimensions of the general purpose low profile load cells with Load &...
GENERAL PURPOSE LOW PROFILE LOAD CELL OPERATION MANUAL 3.2 Standard Components NOTE: The use of the tension base greatly increases the performance of the low profile load cell. The following figure describes the standard components of the general purpose low profile load cells. 4.0 MECHANICAL INSTALLATION Load Cell The following mechanical installation instructions should be...
GENERAL PURPOSE LOW PROFILE LOAD CELL OPERATION MANUAL 5.2 Cable & Grounding Considerations Proper grounding and shielding is required to prevent electrical noise in strain gage load cell measuring systems. The cable must be shielded twisted pairs with a drain wire. Cable shields must be grounded only at one end, for example, on the instrument or control system ground.
GENERAL PURPOSE LOW PROFILE LOAD CELL OPERATION MANUAL 6.1.1 Measured Output 7.0 MOMENT COMPENSATION The applied load starting at zero is measured in five When an eccentric axial load is increments to full scale. Output (mV/V) is measured at each applied to a load cell that is not Primary increment.
GENERAL PURPOSE LOW PROFILE LOAD CELL OPERATION MANUAL are not allowed to migrate into devices that are not Check bridge balance. hermetically sealed. Such devices should only be wiped with a damp cloth, and never be submerged or have liquids poured 10.
GENERAL PURPOSE LOW PROFILE LOAD CELL OPERATION MANUAL Standard calibration certificates list five force points ascending and one point descending in both tension and compression. Additional data points are available at extra cost upon request. Certificate information includes tabulated measurement variable data zero balance, bridge input/output resistance, computer nonlinearity and hysteresis, static error band (SEB) calculations and entries abilities and traceability statements.
Page 15
Model Number Revision: B LOW PROFILE LOAD CELL 1203-15A ECN #: 43809 Performance ENGLISH OPTIONAL VERSIONS Measurement Range 10 klbf 44 kN [6][7] Optional versions have identical specifications and accessories as listed for the standard model Sensitivity(± 10 %)(RO) 4.0 mV/V 4.0 mV/V...
Page 16
PCB Load & Torque Inc. claims proprietary rights in REVISIONS the information disclosed hereon. Neither it nor any reproduction thereof will be disclosed to others without DESCRIPTION the written consent of PCB Load and Torque Inc. CONNECTOR "B" HUB DIAMETER WAS 1.25 [31.8] - 10.05.16, PTE 46027 22.5°...