8 Maintenance, servicing, repairs, and disposal ................. 18 8.1 Repairs to the product ..................... 18 8.2 Disposing of the product ....................18 9 Accessories and spare parts ....................19 9.1 Accessories ........................19 10 Subsidiaries ..........................20 Instructions for use • IMAGE1 PILOT • HLI417_EN_V2.0_11-2021_IFU_CE-MDR...
ð Interim result of an action Step 2 ð Result of a completed action Actions in safety notes or in the case of a single step: Step 1 Lists Numbered list – Unnumbered list, 1st level Instructions for use • IMAGE1 PILOT • HLI417_EN_V2.0_11-2021_IFU_CE-MDR...
Designates a possible imminent risk. If this is not avoided, it could lead to minor injuries. NOTICE NOTICE Designates a possibly harmful situation. If this is not avoided, the products could be damaged. Instructions for use • IMAGE1 PILOT • HLI417_EN_V2.0_11-2021_IFU_CE-MDR...
2 Normal use 2.1 Intended use The IMAGE1 PILOT is an input device and is used for operating the IMAGE1 camera control unit and connected devices / camera heads. The IMAGE1 PILOT has no direct contact with the human body.
Observe the instruction manuals and interface specifications of the devices and components used in combination. Comply with national and local regulations. Only make changes to the product if these changes are approved by KARL STORZ. 3.5 Failure of devices The product may fail during use.
The product can be combined with camera heads, articulated stands, and rotation sockets, e.g.: – VITOM 3D (TH200) – Articulated stand (28272 HB) – Rotation socket (28172 HR) 4.3 Technical data Description Value Dimensions (L x H x W) 123 x 61 x 92 mm Weight 355 g Instructions for use • IMAGE1 PILOT • HLI417_EN_V2.0_11-2021_IFU_CE-MDR...
EU directives. A code number after the CE mark indicates the responsible notified body. The EU legislation relevant to the product can be found in the EU Decla- ration of Conformity, which can be requested from KARL STORZ. 4.5 Ambient conditions Storage and transport conditions Temperature -10°C ... +60°C (14°F ... 140°F)
Page 10
Product description Operating conditions Temperature 0°C ... 40°C (32°F ... 104°F) Relative humidity, non- 20 – 95% condensing Air pressure 700 – 1,060 hPa Instructions for use • IMAGE1 PILOT • HLI417_EN_V2.0_11-2021_IFU_CE-MDR...
Keep packaging for further transport. 5.2 Assembling the product The product is attached to the operating table using a KARL STORZ articulated stand. Two IMAGE1 PILOT products can be connected at the same time, enabling the product to be operated in both sterile and unsterile environments.
Connect the USB cable of the control unit to a USB socket on the IMAGE1 S CONNECT or IMAGE1 S CONNECT II. ð The message “IMAGE1 PILOT connected” appears on the display. ð The LED indicating the function status lights up. 5.3 Attaching a sterile drape The product cannot be sterilized and must be fitted with a sterile drape (041150-20) in a sterile environment for the duration of the procedure.
Page 13
Ensure that the elastic insert on the tip of the sterile drape is placed directly over the 3D wheel. NOTICE! Do not pull the drape on too tightly as this may inadvertently operate the 3D wheel. Instructions for use • IMAGE1 PILOT • HLI417_EN_V2.0_11-2021_IFU_CE-MDR...
Push the 3D wheel to navigate through the IMAGE1 menu and operate a camera-specific function. Press the 3D wheel to confirm or select a function. Pull the 3D wheel to exit a menu, cancel an action, or operate a camera-specific function. Instructions for use • IMAGE1 PILOT • HLI417_EN_V2.0_11-2021_IFU_CE-MDR...
The four function keys can be programmed and functions can be assigned to them. Depending on which camera system is connected, certain functions can be performed quickly, or rapid access can be gained to menus. Instructions for use • IMAGE1 PILOT • HLI417_EN_V2.0_11-2021_IFU_CE-MDR...
Page 16
Application Select the button for the IMAGE1 PILOT in the camera’s live menu. Select the required function for the function key from the drop-down menu. Instructions for use • IMAGE1 PILOT • HLI417_EN_V2.0_11-2021_IFU_CE-MDR...
Disassembly 7 Disassembly 7.1 Disassembling the product Remove the sterile drape. Open the KSLOCK quick release coupling (female) on the articulated stand by turning it in the “OPEN” direction. Remove the product. Instructions for use • IMAGE1 PILOT • HLI417_EN_V2.0_11-2021_IFU_CE-MDR...
8 Maintenance, servicing, repairs, and disposal 8.1 Repairs to the product Repair work may only be performed by KARL STORZ or by a company authorized by KARL STORZ. Please contact your local KARL STORZ subsidiary or authorized dealer (see the list of subsidiaries).
Accessories and spare parts 9 Accessories and spare parts 9.1 Accessories Item Order no. Drape, elastic tip 041150-20 Instructions for use • IMAGE1 PILOT • HLI417_EN_V2.0_11-2021_IFU_CE-MDR...