Hanna Instruments Checker HI 721 Quick Start Manual

Hanna Instruments Checker HI 721 Quick Start Manual

Iron high range

Advertisement

Quick Links

Dear Customer,
Thank you for choosing a Hanna Instruments Product.
HI 721
Iron High Range
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occured during shipment, please notify your Dealer.
Each HI 721 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• Six powder reagents for Iron High Range
• 1 x 1.5V AAA Battery
• Instruction Manual
Range
Resolution
Precision
Typical EMC Dev.
Light Source
Light Detector
Method
Environment
Battery Type
Auto-Shut off
Dimensions
Weight
For more details about spare parts and accessories
see "Accessories".
Technical specifications:
0.00 to 5.00 ppm
0.01 ppm
±0.04 ppm ±2% of reading @ 25°C
±0.01 ppm
Light Emitting Diode @ 525 nm
Silicon Photocell
Adaptation of the EPA Phenantroline method 315B,
for natural and treated waters. The reaction between
iron and reagent causes an orange tint in the sample.
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
1 x 1.5V AAA
After 2 minutes of non-use and 10 seconds after reading.
81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5")
64 g (2.25 oz.).
Functional description:
1. Cuvette cap.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted cuvettes: The sample and the zero cuvette are inverted.
Under range: A blinking "0.00" indicates that the sample absorbs
less light than the zero reference. Check the procedure and make sure
you use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
an over range condition. The concentration of the sample is beyond
the programmed range: dilute the sample and re-run the test.
Battery low: The battery must be replaced soon.
Dead battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, the meter will lock up.
Change the battery and restart the meter.

Advertisement

loading
Need help?

Need help?

Do you have a question about the Checker HI 721 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hanna Instruments Checker HI 721

  • Page 1 Functional description: Dear Customer, Thank you for choosing a Hanna Instruments Product. HI 721 1. Cuvette cap. Iron High Range Preliminary examination: 2. Cuvette with cap. Please examine this product carefully. Make sure that the instrument is not damaged. If 3.
  • Page 2 “bAd” then ”bAt” each for 1 second and turns off. • Remove the cuvette, open it and add the Hanna Instruments reserves the right to modify the design, construction and To restart the instrument, the battery must be replaced with a fresh one.