Product implementation standards
EN 61010-1 :2010/A1:2019
EN 61010-2-011:2017
EN 61326-1 :2013
USER MANUAL
Freez

Advertisement

Table of Contents
loading

Summary of Contents for Vacc-Safe VS-25L308

  • Page 1 Product implementation standards EN 61010-1 :2010/A1:2019 EN 61010-2-011:2017 EN 61326-1 :2013 USER MANUAL Freez...
  • Page 2: Table Of Contents

     The temperature in the biomedical freezer is uniform and controllable, the 4. Product Installation .........07 temperature range is adjustable to -25℃ (VS-25L308)/-40℃ (VS-40L308), and the temperature control is accurate  Standard USB port and traceable temperature data; query and save temperature 5.
  • Page 3: Safety Precautions

    Otherwise, it Dear Vacc-Safe users: We appreciate your trust. Thank you for using Vacc-Safe biomedical freezer. Please Please use the special power supply indicated on the nameplate of the carefully read the following information to better understand the Operating Instructions and use this product, so as to prevent personal injury and damage to items.
  • Page 4 Never store corrosive substances such as acids or alkalis in the biomedical CAUTION freezer. Otherwise, these substances may cause damage to internal components or electrical parts of the biomedical freezer. Keep the biomedical freezer clear of obstruction around it to ensure smooth Never store flammable, explosive, or volatile substances in the biomedical ventilation.
  • Page 5: Precautions For Use

    Precautions for Use Product Installation Ì When the biomedical freezer is running, the left and right rear sides of the biomedical Installation Environment freezer may be heated, which is not a fault. In order to prevent condensation around Ì Ambient temperature: 10 °C~32 °C; ideal temperature: 18 °C~25 °C; air the biomedical freezer housing, the condensate pipe is installed.
  • Page 6 Ì Operation after Power-off Since the ambient temperature has a large impact on the CAUTION equipment, the equipment may not operate normally if the above The set value is memorized by the equipment. After recovery from a power failure, the environmental requirements cannot be met.
  • Page 7: Product Components And Overview

    Product Components and Overview Schematic Diagram of Appearance Control panel Control panel Grip handle Exterior door handle Door Door Caster...
  • Page 8: Operating Instructions

    Operating Instructions Display Panel Control panel Operating Instructions Exterior door handle Key Symbol Function Press and hold the FN key for 3s under unlocking state to Door enter the maintenance mode; SET key: In the unlocking state, press and hold the SET key for 3s to display the current control temperature set value St: in the setting state, press and hold the key for 3s to save and exit;...
  • Page 9 Management of Menu Parameter Setting a. In the normal operation state, after unlocking, press SET key + key for more than Display Symbol Meaning 3s at the same time, and the LCD display screen will display the parameter code "PAS"; Refrigeration symbol flashing: refrigeration on delay;...
  • Page 10: Alarm Description

    Maintenance Manual Data Printing (Reserved) Manual: Under normal operation state, after unlocking, press and hold the down key for 3s to enter the data print interface; WARNING Select dt1~dt7 by pressing the up and down keys, and press and hold the SET key for 3s to confirm the printing.
  • Page 11: Troubleshooting And Repair Service

    Troubleshooting and Repair Service Defrosting Inside Biomedical Freezer Frosting may cause a gap between the biomedical freezer and the door sealing strip, Any product may fail. Please observe the operation of the equipment in time during resulting in poor refrigeration. The inner door shall be defrosted with an ice shovel use.
  • Page 12: Packing List

    Name Name Biomedical Freezer Biomedical Freezer Model No. VS-25L106 VS-40L106 Model VS-40L308 VS-25L308 Climate type Climate type Electrical shock protection class Class I Class I Electrical shock protection class Class I Class I Power supply 220~240V~/50Hz 220~240V~/50Hz Power supply 220-240V~/50Hz...
  • Page 13: Electrical Schematic Diagram

    Electrical Schematic Diagram VS-25308/ VS-25L106 Power plug USB board Operating capacitor Compressor Door switch Controller Ambient temperature sensor Alarm terminal Biomedical block freezer temperature sensor VS-40308/ VS-40L106 Power plug USB board Operating capacitor Compressor Door switch Controller Condenser fan Ambient temperature sensor Alarm terminal...

This manual is also suitable for:

Vs-40l308Vs-25l106Vs-40l106