Interface; Rs-485/Rs-232 Commands - Instrutech VGC-301 User Manual

Vacuum gauge controller
Hide thumbs Also See for VGC-301:
Table of Contents

Advertisement

Instruction Manual

Interface

RS485/RS232 Command Protocol Summary. (All values must be set in Torr)
Note: This protocol was designed to be 100% compatible with the Helix / Granville-Phillips Mini-Convectron
COMMAND
BRIEF DESCRIPTION
READ
Read the current pressure in
Torr
SET ADDR
Set the communications
OFFSET
(RS485) address offset
(upper nibble) *
SET SPAN
Set the span or atmosphere
calibration point
SET ZERO
Set the zero or vacuum
calibration point
SET TRIP
Set the 'turns on below'
POINT #1
pressure point for relay #1
and set the 'turns off above'
pressure point for relay #1.
SET TRIP
Set the 'turns on below'
POINT #2
pressure point for relay #2
and set the 'turns off above'
pressure point for relay #2.
READ TRIP
Read the 'turns on below'
POINT #1
pressure point for relay #1
and read the 'turns off above'
pressure point for relay #1.
READ TRIP
Read the 'turns on below'
POINT #2
pressure point for relay #2
and read the 'turns off above'
pressure point for relay #2.
READ SW
Read the revision number of
VERSION
the firmware.
SET
Force unit to return ALL
FACTORY
settings back to the way the
DEFAULTS
factory programmed them
before shipment.*
SET BAUD
Set the communications baud
RATE
rate for RS485 and RS232.*
SET NO
Set the communications to
PARITY
NO parity, 8 bits for the
RS485 and RS232.*
SET ODD
Set the communications to
PARITY
ODD parity, 7 bits for the
RS485 and RS232.*
SET EVEN
Set the communications to
PARITY
EVEN parity, 7 bits for the
RS485/ RS232.*
RESET
Reset the device as if power
was cycled. (required to
complete some of the
commands.)
* Commands marked with an asterisk will not take effect until after RESET command is sent or power is cycled.
InstruTech, Inc
InstruTech VGC-301 Convection Vacuum Gauge Controller
COMMAND SYNTAX
#xxRD<CR>
(eg: #01RD<CR>)
#xxSAxx<CR>
(eg: #01SA04<CR>)
#xxTSy.yyEzyy<CR>
(eg: #01TS7.60E+02)
#xxTZ0<CR>
(eg: #01TZ0<CR>)
#xxSLzy.yyEzyy<CR>
(eg: #01SL+4.00E+02<CR>)
(eg: #01SL-5.00E+02<CR>)
#xxSHzy.yyEzyy<CR>
(eg: #01SH+4.00E+02<CR>)
(eg: #01SH+5.00E+02<CR>)
#xxRLz<CR>
(eg: #01RL+<CR>)
(eg: #01RL-<CR>)
#xxRHz<CR>
(eg: #01RH+<CR>)
(eg: #01RH-<CR>)
#xxVER<CR>
(eg: #01VER<CR>)
#xxFAC<CR>
(eg: #01FAC<CR>)
#xxSByyyyy<CR>
(eg: #01SB19200<CR>)
#xxSPN<CR>
(eg: #01SPN<CR>)
#xxSPO<CR>
(eg: #01SPO<CR>)
#xxSPE<CR>
(eg: #01SPE<CR>)
#xxRST<CR>
(eg: #01RST<CR>)
®
.
RESPONSE
*xx_y.yyEzyy<CR>
(eg: *01_7.60E+02<CR>)
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
*xx_y.yyEzyy<CR>
(eg: *01_7.60E+02<CR>)
*xx_y.yyEzyy<CR>
(eg: *01_7.60E+02<CR>)
*xx_mmnnv-vv
(eg: *0105041-00)
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
No response
page 11

Advertisement

Table of Contents
loading

Table of Contents