CALIBRATION
Temperature/CO
1. Press the MENU Key from the Top Display to display the menu, select Tools, and press the ENTER
Key.
1
T e m p
2
37.0
3
4
5
6
7
2. Select Temp/CO2/O2 Calibration from the Select Tools Display, press the MENU Key to display the
menu, select OK, and press the ENTER Key.
S e l e c t
T e m p / C O 2 / O 2
A l a r m
L C D / D A Q
U V
K e y
D a t e
3. The Temp/CO2/O2 Calibration Display will appear.
Example
Enter the value shown below if the displayed chamber temperature is 37.0
o
is 36.5
C.
●Temperature span: Enter 365.
1
T e m p / C O 2 / O 2
3
T e m p
4
C O 2
5
C O 2
O 2
6
O 2
/O
2
o
3 7 . 0
C
C O 2
S t a t u s
O K
T o o l s
C a l i b r a t i o n
S e t t i n g
S e t t i n g
S e t t i n g
L o c k
P W
S e t t i n g
T i m e
C a l i b r a t i o n
S p a n
3 6 . 5
Z e r o
S p a n
0 . 0 %
Z e r o
S p a n
0 . 0 %