Page 1
DT201D Auto Range Digital Clamp Meter OPERATOR’S INSTRUCTION MANUAL DT201D TRUE RMS ⁓ 1000A ⁓ 600A EN 61010-1 ⁓ CAT III 600V CATII 1000V ⁓ Duty TEMP Range Select 1.8.8.8 1000v 1000v WARNING READ AND UNDERSTAND THIS MANUAL BEFORE USING THE INSTRUMENT.
Page 2
Resistance, Audible Continuity, Diode and Temperature measurements. It is a 3 5/6 digits, 6000 counts auto ranging digital clamp multimeter. DT201D digital clamp multimeter has been designed according to EN61010-1 oncoming electronic measuring instruments with an over voltage category (CAT III 600V CAT II 1000V) and Pollution degree 2.
Page 3
is danger of electric shock. • Use the proper terminals, function, and range for your measurements. • Do not use or store the meter in an environment of high temperature, humidity, explosive, inflammable and strong magnetic field. The performance of the meter may deteriorate after dampened.
2. GENERAL CHARACTERISTICS Display LCD, 6000 counts updates 2/sec LCD size 30 x 49 mm (1.2 x 1.9 in) Polarity Indication “-” displayed automatically Over-range Indication “OL” displayed Low Battery Indication “ ” displayed True-RMS ACV ACA measure 0°C to 40°C (32ºF to 104ºF), Operation Temperature less than 80%RH -10°C to 50°C (14ºF to 122ºF),...
Conforms to European Union directive Double insulated 4. PANEL DESCRIPTION DT201D TRUE RMS ⁓ 1000A 600A ⁓ EN 61010-1 ⁓ CAT III 600V CATII 1000V ⁓ Duty TEMP Range Select 1.8.8.8 1000v 1000v - 5 -...
Page 6
a. Transformer Jaws Pick up the AC Current flowing through the conductor. b. Trigger Press the level to open the transformer jaws when the finger press on the level is released the jaws will close again. c. Hz/Duty Button Push this button to select Hz or DUTY CYCLE measurement, at Hz positing.
Page 7
and “H” symbol will appear till pushed again. Back light function, push this button for more than 3 seconds, for back light, after 15 seconds, back light will turn off automatically. 1. Duty Cycle is selected 2. Fahrenheit temperature test is selected 3. Celsius degree test is selected 4. Continuity test is selected 5. Diode test is selected...
14. AC 15. Low Battery and replaced immediately 5. SPECIFICATIONS Accuracy is guarantied for 1 year 23°C±5°C less than 80%RH 5-1. DC Voltage (Auto ranging) Input Impedance: 10MΩ Range Resolution Accuracy 600mV 0.1mV ±(0.8% of rgd + 5dgts) ±(0.8% of rgd + 3dgts) 10mV 600V 100mV...
Page 9
5-3. AC Voltage (Auto ranging) Range Resolution Accuracy 600mV 0.1mV ±(1.2% of rgd + 8dgts) ±(1.2% of rgd + 6dgts) 10mV 600V 100mV ±(1.2% of rgd + 8dgts) 750V Input Impedance: 10MΩ Input Impedance: 10MΩ Overload Protection: 600V AC RMS CAT III, 1000V AC RMS CAT II Max.
Page 10
600KΩ 100Ω ±(1.5% of rgd + 5dgts) 6KMΩ 1KΩ 60MΩ 10KΩ Open Voltage: about 0.25V Overload Protection: 250V DC/AC RMS 5-6. Diode and Continuity Range Introduction Remark Open circuit The approximate forward voltage: voltage drop will be displayed about 1.5V The built-in buzzer will sound Open circuit if the resistance is less than...
Page 11
5-9. Duty Cycle Range Resolution Accuracy 0.01%~99.9% 0.01% ±1% For value between 10% and 90% duty cycle at 50Hz. 6. OPERATION INSTRUCTION 6-1. Measuring Voltage 1. Connect the BLACK test lead to the “COM” jack and the RED to the “V” jack. 2.
Page 12
2. Press the trigger to open the transformer jaws and clamp one conductor only it is impossible to make measurements when two or three conductors are clamped at the same time. 3. Display reading is flowing the conductor AC current. 6-3.
Page 13
6-4. Continuity Test 1. Connect the BLACK test lead to the “COM” jack and the RED to the “ ” jack (Note: The polarity of the red test lead is positive “+”). 2. Set the function switch to “ ” range 3.
2. Make sure the polarity of the thermocouple is correct; put the cold end (free end) of the thermocouple sensor into the terminal (black to COM jack and red to “ ” jack). 3. Set the working end (testing end) on or inside the object under test.
Note: Do not apply more than 250V RMS to the input. Indication is possible a voltage higher than 100V rms, but reading maybe out of specification. At the same time using the clamp hand to test AC current, press key, the meter shows the frequency of the measured current.
Page 16
9. ACCESSORIES Owners manual: 1 piece Test leads: 1 pair K type thermocouple: 1 piece Battery(9V 6F22): 1 piece DISPOSAL Dear Customer, If you as some point intend to dispose of this product, then please keep in mind that many of its components consist of valuable materials, which...