7.1 Run Live Assessment ............33 2.1.2 Sensor light status color guide .........6 7.1.1 Starting a Run Live Assessment ......33 2.1.3 Charging ViMove2 sensors........6 2.1.4 Turning ViMove2 sensors on ........6 7.1.2 Live Assessment screen .........34 2.2 Sensor Adhesives ..............7 7.1.3 Capturing running data ...........34 7.1.4 Run metrics overview ..........35...
1. Regulatory Information 1.1 Symbols and labeling information The following table lists the labels attached to various components of ViMove2. Symbol Explanation Symbol Explanation Follow instructions for use Use by date Class II equipment Do not use if packaging is damaged...
• Always ask the subject about skin allergies they may have prior to ViMove2 use. Gauge the severity of the allergy and 1.3 Warnings assess whether a reduced session on ViMove2 and / or using a protective film on the skin prior to application, would be None noted.
Furthermore, a ISO 60601-1-6 verify and validate usability, as it relates to strong password should be chosen for the user’s ViMove2 account. basic safety and essential performance of The following are characteristics of a strong password: medical electrical equipment.
ViMove2 is a wireless electronic device used by healthcare professionals to accurately measure, record and analyse movement. The software guides the client through a series of movements and postures. ViMove2 then measures the movement, streams the data via Bluetooth Low Energy (BTLE) to a device, and generating personalized assessments.
For the fastest charging time it is recommended to charge your If the sensors are not in assessment or monitoring mode, they will device using the provided power plug. ViMove2 sensors take 1 turn off automatically after 10 minutes. hour to fully charge using a power plug.
3.4 Software installation ViMove2 can be download from the Apple App Store on your compatible tablet device. To download the software, open the App Store, search for “ViMove2” and follow the prompts to download and install the software. Once the installation has finished, a “ViMove2” app icon will appear on your home screen.
4.1 Logging into your clinic To login to ViMove2 for the first time, you will receive an email from dorsaVi containing a temporary password to gain access to your clinic. Follow the prompts in this email to login to your clinic and set a new password for your account. For subsequent logins, open the ViMove2 app, enter your login details (email address and password) and press “Login”.
All client information and assessment data recorded by ViMove2 is securely stored in dorsaVi’s central servers. To store this information or share it with other users in the clinic, data must be “synced” between the ViMove2 app and the servers. To check if your device is synced refer to the “sync status”...
Some of the detailed assessment data is kept in the cloud and is only downloaded when you need to see it. If you are planning to use your device offline, this tab will start an automatic data offload. This will take some minutes, depending on the volume of data. Once complete, users will be able to view that data offline. ViMove2 User Manual | A.474.J...
You can also give this user administrative rights to the app by tapping ‘make account manager’ To delete the user tap ‘Remove User’. Note that all assessments performed by that user will remain in the app and will NOT be deleted. ViMove2 User Manual | A.474.J...
Prior to starting an assessment please ensure your device meets the following requirements: • iOS device is reasonably charged (above 50% is ideal). • ViMove2 sensors are reasonably charged (above 50% is ideal). • Bluetooth on your iOS device is turned on. 4.8 Starting an assessment...
5. Client Mode Client Mode is a tool within the ViMove2 app which can be used to gather information about your clients and promote your business. It is opened by tapping the ‘client mode’ tab on the home screen. Note, Client Mode is designed for your clients to use directly while waiting for an appointment.
5.2 Client Sign In Clients can tap on ‘sign in’ to register themselves in the ViMove2 App. This includes entering contact details and answering some standardized questions about the reason for their visit. This information will sync to populate the client’s history, saving the clinician time during the actual appointment.
6. ViMove2 – Low Back The ViMove2 – Low Back includes 4 session types: Quick Assessment Live Assessment Monitoring Live Training Quick and Live assessments are performed in the clinic according to a protocol. Monitoring is used to assess unplanned activities over a longer period of time.
Page 17
ViMove2 sensors. on the client. If this is correct, select Select “I attached both sensors” in the “Yes” in the ViMove2 app. If the colors ViMove2 app. are incorrect select “No, swap colors”. ViMove2 User Manual | A.474.J...
Stop capturing data for the test and move onto the next test. Recalibrate the sensors if the avatar’s posture does not look correct. Start capturing data for the test. Move on to the next test. ViMove2 User Manual | A.474.J...
Recalibrate the sensors if the avatar’s posture does not look correct. a real-time visualisation of the client’s posture for each movement. Start capturing data for the test. Stop capturing data for the test and move onto the next test. ViMove2 User Manual | A.474.J...
6.1.6 Quick Assessment test protocol In a ViMove2 Low Back Quick Assessment, the following movements can be tested. For all tests, use the “Start” and “Stop” buttons located at the bottom right hand corner of the screen to start and stop recording the movement data.
Page 21
6.1.8 Share a Low Back Quick Assessment report To share a Quick Assessment report select the “Share” icon located in the top right hand corner of the ViMove2 app while viewing a report and select your report type. After selecting a report type you will be able to view the client’s assessment data and notes in ViMove2. Select the mail button located in the bottom right hand corner to email the report as a PDF document and select the arrow icon to expand the report on the screen.
Page 22
2. Push the button on the side of the 3. Peel the Sensor Adhesives off the client you would like to test and select ViMove2 sensors to turn them on. From adhesive roll and apply to the back of the “New Test”. Following the on-screen the sensor selection screen, tap on the ViMove2 sensors.
Page 23
End the current assessment without saving the session data. Record pain scores for movements that caused the client pain. The pain score must be recorded while ViMove2 is capturing movement The side panel displays all tests that are to be performed in the Low data for the score to be tied to the selected test.
Page 24
Record pain scores for movements that caused the client pain. The anterior and posterior tilt movement in real-time as they perform pain score must be recorded while ViMove2 is capturing movement pelvic tilt tests. data for the score to be tied to the selected test.
Page 25
Record pain scores for movements that caused the client pain. The anterior and posterior tilt movement in real-time as they perform pain score must be recorded while ViMove2 is capturing movement the sitting pelvic tilt test. data for the score to be tied to the selected test.
6.2.6 Low Back Live Assessment test protocol In a ViMove2 Low Back Real Time Assessment, the following movements can be tested. Any pain experienced during the session can be recorded on a 0-10 scale by selecting a number on the right hand corner of the screen during the movement which caused the pain (0 being not painful at all, and 10 being extremely painful).
Page 27
The detailed view screen provides an in-depth look at the individual movements performed by the client so assessors can pinpoint areas of concern and re-watch selected movements using the ViMove2 avatar. View a general overview of the assessment report and suggested This area displays the movement data obtained throughout the exercises based on the client’s movements throughout the test.
Page 28
6.2.8 Compare Low Back Real Time Assessment data ViMove2 allows assessors to compare up to three Low Back Live Assessments with one another to observe changes in the client’s movement data. To access this feature, navigate to a client’s Low Back Live Assessment results and select the “Compare” button located at the top right hand corner of the screen.
ViMove2 sensors. press down firmly on the sensors. on the client. If this is correct, select “Yes” in the ViMove2 app. If the colors Select “I attached both sensors” in the are incorrect select “No, swap colors”.
6.3.2 Offloading Monitoring session data At the conclusion of the Monitoring session, the client will need to return to the clinic. Upon their return, remove the ViMove2 sensors from the client, navigate to the client’s profile on ViMove2 and select “Download Monitoring data”.
6.3.5 Share a Low Back Monitoring session report To share a Monitoring Assessment report select the “Share” icon located in the top right hand corner of the ViMove2 app while viewing a report. After selecting this option you will be able to view the client’s Monitoring data and notes in ViMove2. Using the mail button located in the bottom right hand corner to mail the report as a PDF document, and select the arrow icon to expand the report.
ViMove2 sensors. client. If this is correct, select “Yes” 8. Select “I attached both sensors” in the in the ViMove2 app. If the colors are ViMove2 app. incorrect select “No, swap colors”. ViMove2 User Manual | A.474.J...
End the live training session, noting that session data is not saved. Recalibrate the sensors if the avatar’s posture does not look correct. Select the movement traces to be displayed. Stop and start the data display. ViMove2 User Manual | A.474.J...
7.1.1 Starting a Run Live Assessment 1. From the client list screen, select the 2. Push the button on the side of the ViMove2 3. Peel the Sensor Adhesives off the client you would like to test and select sensors to turn them on.
Each test should be performed in the order it appears on the screen. For each Run test, the client must run for at least 10 seconds and must maintain a minimum speed of 6km/h to be recorded by ViMove2.
7.1.4 Run metrics overview ViMove2 Run captures a wide array of metrics during a client’s run. Below is an overview of each metric captured by ViMove2 Run. ASI - Absolute Symmetry Index ASI measures the percentage of asymmetry between average GRF values of the left and right legs. ASI is typically calculated as follows: ASI = 100 * ((LEFT GRF VALUE –...
7.1.6 Share a Run Live Assessment report To share a Live Assessment report select the “Share” icon located in the top right hand corner of the ViMove2 app while viewing a report. After selecting this option you will be able to view the client’s Live Assessment data and notes in ViMove2. Use the mail button located in the bottom right hand corner to mail the report as a PDF document, and select the arrow icon to expand the report.
Page 38
Tap ‘Connect to Sensors’ Attach sensors following steps 1-4 on screen Ask the client to stand still, then tap ‘calibrate standing position’ Use the slide bar to select the duration of the session ViMove2 User Manual | A.474.J...
Page 39
Once the Monitoring session has started you can home return to the client overview screen and continue to use ViMove2 whilst the monitoring session is in progress. Once session is completed you can download the data. Firstly, make sure the sensors are turned on, then tap ‘Download Monitoring Data’.
Swipe screen up to segment and compare metrics. There is an option to automatically split data into even segments by either: distance or time. Select 2 or more metrics to compare by tapping on metric labels. Tap again to deselect. ViMove2 User Manual | A.474.J...
Page 41
A Detailed Report is recommended if you have used the Run Analysis feature and want to create a report which is customized to the metrics you have selected. After Trimming ViMove2 User Manual | A.474.J...
8.1.1 Starting a Knee Live Assessment 1. From the client list screen, select the 2. Push the button on the side of the ViMove2 3. Peel the Sensor Adhesives off the client you would like to test and select sensors to turn them on.
Selecting a repetition performed during the assessment will display this button. Pressing this button will delete the repetition from the Reference values for the most recent repetition recorded. assessment. Client’s knee test data for the most recent repetition recorded. ViMove2 User Manual | A.474.J...
8.1.3 Knee Live Assessment test protocol In ViMove2’s Knee Live Assessment the following movements can be tested. For all tests, use the “Start Capture” and “Stop Capture” buttons located at the bottom right hand corner of the screen to record the following movements. To move on to the next test select the “Next”...
To share a Live Assessment report select the “Share” icon located in the top right hand corner of the results screens. After selecting this option you will be able to view the client’s Live Assessment data and notes in ViMove2. Use the mail button located in the bottom right hand corner to mail the report as a PDF document, and select the arrow icon to expand the report.
9. iOS Native Sharing ViMove2 allows assessors to print or export an assessment report using Apple’s native sharing functionality. Printing only works with AirPrint compatible printers on the same wireless network as the iPad. 9.1 Printing Select the share icon.
10. Data Export ViMove2 allows assessors to export an assessment’s captured data as a csv file. 10.1 Configuring an email account To export data via email, your iPad will need to have an email account configured. To configure an email account, select the icon.
When available, this represents the average body orientation the subject was in for the captured time period. Minor Orientation When available, this represents the orientation sub-type the subject was in for the captured time period. ViMove2 User Manual | A.474.J...
Page 49
Load Speed (Left/Right) The load speed in degrees per second (where applicable) Land Speed (Left/Right) The land speed in degrees per second (where applicable) *Note that columns which contain “Stream” and “PointIndex” should be ignored ViMove2 User Manual | A.474.J...
11. ViMove2 Exercises The dorsaVi Exercise Library allows you to create electronic exercise programs for your clients. These can be emailed as a printable pdf or shared with your client on their own myViMove app. Your app is pre-populated with a number of exercise templates ready for use.
Page 51
- ‘share’ with others by emailing pdf - ‘expand’ to view details If you choose to ‘save as template’, you will be asked to give the exercise template a new name. Tap ‘save’ to save in exercise template list. ViMove2 User Manual | A.474.J...
This screen displays the saved exercise templates. They can be modified or deleted from this screen. Create your own template by accessing the exercise library. Follow the same steps as you would for creating a program for a client. ViMove2 User Manual | A.474.J...
Android - tested on Samsung Galaxy S9 & LG G6 phones. 12.3 Inviting your clients to myViMove When a client profile is created in ViMove2, you will be able to send an email to the client with details about how to download and log into the myViMove app.
9 - 12 km/h 13 - 15 km/h >16 km/h ACCURACY GRF Accuracy 130N 140N 150N * Refer to the precautions section on page 4. **For subjects with body masses between 75kg and 100kg (165lb and 220lb) ViMove2 User Manual | A.474.J...
For long-term storage, maintain sensors at 20% to 50% charge. 15. Normative Data Normative values provided in the ViMove2 assessment reports are based on an analysis of lumbo-pelvic range of motion in 50 adults with no history of chronic lower back pain (LBP).
16. Electromagnetic Emissions and Immunity ViMove2 complies with the requirements of IEC 60601-1-2. The recommended tables for electromagnetic emissions per sub-clause 5.2.2.1(c) and electromagnetic immunity per sub-clause 5.2.2.1(f) & 5.2.2.2 are provided below: ViMove2 User Manual | A.474.J...
Page 57
Laird, Robert A., Peter Kent, and Jennifer L. Keating. “How consistent are lordosis, range of movement and lumbo-pelvic rhythm in people with and without back pain?.” BMC Musculoskeletal Disorders 17.1 (2016): 403. ViMove2 User Manual | A.474.J...
(2) Cet appareil doit accepter Toute interférence, y compris les interférences qui peuvent causer un mauvais fonctionnement du dispositive Any changes or modifications not expressively approved by dorsaVi could void the user’s authority to operate this equipment. ViMove2 User Manual | A.474.J...
Australia / APAC: www.dorsavi.com/au/en/vimovestore Europe: www.dorsavi.com/uk/en/vimovestore 19. Troubleshooting Unsure about a part of our software or need extra help to use ViMove2? Visit: http://support.dorsavi.com/?b_id=16080 to view troubleshooting guides and FAQs related to ViMove2 and myViMove 20. dorsaVi Contact Details United States Email info@dorsavi.com...
Need help?
Do you have a question about the ViMove2 and is the answer not in the manual?
Questions and answers