Chapter 1 Performance Tests
Test 3: Channel 1 Frequency Accuracy (Agilent 53181A Only)
4
Record the actual reading in the Performance Test Record (Test 3, Line 1).
5
Change the signal generator frequency to 100 MHz.
Verify the Counter reads 100 MHz ±70 mHz.
6
Record the actual reading in the Performance Test Record (Test 3, Line 2).
7
Change the signal generator frequency to 200 MHz.
Verify the Counter reads 200 MHz ±130 mHz.
8
Record the actual reading in the Performance Test Record (Test 3, Line 3).
9
Change the signal generator frequency to 225 MHz.
Verify the Counter reads 225 MHz ±150 mHz.
10
Record the actual reading in the Performance Test Record (Test 3, Line 4).
11
Disconnect the test setup.
Assembly-Level Service Guide
1-57