LACO Technologies TITAN Test Manual

LACO Technologies TITAN Test Manual

Leak detector
Table of Contents

Advertisement

TitanTest Production and Maintenance Models
From Software Version 1.16
ORIGINAL
- OPERATING INSTRUCTIONS
- MAINTENANCE INSTRUCTIONS
- COMMUNICATION PROTOCOL
LEAK DETECTOR

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TITAN Test and is the answer not in the manual?

Questions and answers

Rory Walsh
June 16, 2025

Titan Helium leak detector (TMW04-BXX11): Inlet flange to hose barb adapter. Do you have a part number for this item? can you provide the type of flange?

Subscribe to Our Youtube Channel

Summary of Contents for LACO Technologies TITAN Test

  • Page 1 ORIGINAL - OPERATING INSTRUCTIONS - MAINTENANCE INSTRUCTIONS - COMMUNICATION PROTOCOL LEAK DETECTOR TitanTest Production and Maintenance Models From Software Version 1.16...
  • Page 2 Salt Lake City, UT 84104 WEB www.lacotech.com info@lacotech.com Copyright© 2017 LACO Technologies, Inc. No part of this manual may be reprinted, translated or duplicated without the expressed written consent of LACO Technologies, Inc. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 Titel.fm] - 1807...
  • Page 3 LACO ORIGINAL T ™ ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 Titel.fm] - 1807...
  • Page 4 LACO ORIGINAL T ™ ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 Titel.fm] - 1807...
  • Page 5: Table Of Contents

    Recurring function symbols ..............24 6.2.4 Structure of the windows and the menus..........24 6.2.5 Measurement display ..................25 6.2.6 Setup window display .................27 6.2.7 Standby......................28 Global settings ....................30 6.3.1 Setting the display ..................30 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 IVZ.fm] - 1807...
  • Page 6 Replace the filter of the oil mist filter (M-Series only) ......72 Maintenance work (P-Series only)............73 8.5.1 The system at a glance (P-Series only)...........74 8.5.2 Removing the cover (P-Series only)............75 8.5.3 Backing pump (P-Series only)..............76 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 IVZ.fm] - 1807...
  • Page 7 12.4 Commands** ....................117 12.5 Enumerations**................... 127 12.6 Group commands** .................. 131 12.7 Error messages** ..................133 ASCII Protocol ............134 13.1 Communication Parameters ..............134 13.2 Command Format..................134 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 IVZ.fm] - 1807...
  • Page 8 Serial communication via RS232 (common)** ........ 143 14.2 LD Protocol specific** ................144 14.3 ASCII Protocol specific** ................144 INDEX ...............145  ** Feature not available in T ™ Maintenance model ITAN viii © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 IVZ.fm] - 1807...
  • Page 9: About This Manual

     Operating Instructions SplitFlow 80  TCM, SMT-07-1027  ATLAS Displaying information 1.4.1 Warnings Imminent threat of danger resulting in death or severe injuries Dangerous situation potentially resulting in death or severe injuries © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 10 LACO ORIGINAL OPERATING INSTRUCTIONS T ™ ITAN Dangerous situation resulting in minor injuries Dangerous situation resulting in damage to property or the environ- ment © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 11: Safety

    (see Chapter 1.3: "Other applicable documents", page 1), especially the informa- tion regarding safety, maintenance and repairs, before starting work. Define the responsibilities, authorizations and supervision of personnel. ► © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 12: Operator Requirements

    Do not use tracer gases with a hydrogen concentration > 5%. ► Only use the device outside potentially explosive areas. ► No smoking. Do not subject the device to open fire and avoid sparking. ► © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 13 This may result in the burn- ing out of the cathode. Do not test gas whose molecules contain halogens such as fluorine or ► chlorine. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 14: Scope Of Delivery, Transport, Storage

    Don´t change orientation from upright and level for transporting. ► To avoid risks use the plug supplied with the device to close the exhaust ► line. Remove the plug to be operational again after the transport. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 15 Storage Always store the device in consideration of ambient conditions (see Chapter 4.3: "Technical Data", page 10). © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 16: Description

    In the following, the basic unit will be referred to merely as “device” as long as the meaning remains clear. Fig. 2: Frontal view Connection flange DN 25 ISO-KF Display © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 17 Connection strip Mains plug a The mains plug for switching the device on and off. Power supply Connection for power supply cable. The power supply is conducted via this cable. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 18: Markings On The Device

    Max. permitted acceleration during operation 1 G (horizontal) Test port DN 25 ISO-KF Cooling air Inlet below, with dust filter Outlet on side Exhaust port M-Series for hose ø8 mm © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 19 20 Hz Display rate 3 Hz Alarm Acoustics / Volume adjustable Setpoint / Pre-warning adjustable Relay exit adjustable Screen displays Leakage rate vs. time, analog / digital Vacuum mode © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 20 Table 3: Factory settings Factory settings Parameters Default value Reset to default val- Included in set of ue when factory set- parameters tings are loaded Contrast Invert display Display automatic © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 21 Max. pressure 0.65 mbar Min. pressure 0.25 mbar Gross leak protection Limit value, gross leak protection 1E-3 mbar l/s Minimum volume Volume ULTRA enabled Switching threshold ULTRA 0.5 mbar © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 22 Venting with Stop Calibration request int. calibration leak see calibration leak Ext. Calibration leak (vacuum) 1E-7 mbar l/s Ext. Calibration leak (sniff) 1E-5 mbar l/s Calibration mode int. auto © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 23: Titan Test

    38,7 GROSS 88,5 77,59 40,5 86,1 52.4 FINE 135,3 103,7 49,8 112,5 65,5 ULTRA 258,9 1168,3 75,2 182,9 109,9 ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 24: Vacuum Diagrams

    LACO ORIGINAL OPERATING INSTRUCTIONS T ™ ITAN Vacuum Diagrams 4.6.1 Vacuum Schematic with Internal Fore Pump 4.6.2 Vacuum Schematic with Dry Scroll Pump © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 25: Installation

    To remove the plug press the release ring in the direction of the device so ► that the locking device releases. Pull the plug out while release ring is pressed. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 26: Installing The External Oil Mist Filter (P-Series Only)

    A variety of accessories exist for the T ™ (see Chapter 9.2: "Accessories ITAN and consumables", page 100). Refer to the Accessory Manual for information on Installation, Operations and Maintenance of these accessories. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 27: Mounting The Accessories

    The connection is used for venting. Remote control The following remote controls are optionally available for the T ™: ITAN  Remote control TR100, wired  Remote control TR101, wireless © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 28: Fore Pump Exhaust Line

    1. Check whether the supply voltage specified on the T ™ rating ITAN plate matches the supply voltage available on site. 2. Connect the device to the electric power supply using the supplied power cable. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 29: Operation

    4. Allow the device to warm up for 30 minutes before making exact mea- surements or executing a calibration.  ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 30: Operating The Device

    Background cleanup (6.4.4) Check internal calibr. leak( (6.4.10.3) Pumping down (6.2.7) M.-mode (Bar-Graph) (6.2.5) M.-mode (Q(t)-Graph) (6.2.5) Warning/Error (7.) ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 31: Display And Buttons

    Starts or stops the measuring operation. The associated LED lights up permanently during measurement mode and flashes during an evacuation. Display The display shows measured values, device states and settings in different menus. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 32: Recurring Function Symbols

    Fig. 7: Window and menu structure Window name Main display area Menu name Main display area The main display area shows the current state of the device: operating mode, measuring situation, background, tracer gas © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 33: Measurement Display

    The current measured value is displayed. Under this the pressure at test port p2 is shown. Bar graph, logarithmic The current measured value is shown in a bar graph. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 34 Current measured value The current measured value is displayed as a logarithm. Measured value displays as function Shows the progress of the measured values as a function of time. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 35: Setup Window Display

     Changed values that are not saved will not be effective.  Press [Return] to go back to the last menu.  Press [Home] to jump to the "Standby" window. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 36: Standby

    Evacuation time & Vent", you cannot vent. If the word "Vented" is displayed, the venting process is complete and you can replace the test object. ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 37 (see Chapter : "Background cleanup", page 45). Select "Standby > Utilities > Background cleanup". ► ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 38: Global Settings

    In order to adjust the contrast automatically to the device temperature, ► select "Automatic". In order to have the background of the display dark and the letters bright, ► select "Invert Display". © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 39 Time axis 16 ... 960 Time axis, time scale in seconds Select " Setup > Global settings > Display settings > Display range" and ► make the desired setting. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 40: Access Control

    TMP SplitFlow80. For more information, please refer to the separate operat- ing instructions. ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 41 PIN can access the device menu. The menu PIN is requested when accessing the menus. If you enter an incor- rect PIN, the message "Wrong PIN" is displayed. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 42: Set Interfaces

    The menu "Interfaces" allows access to the settings for the following inter- faces:  Analog output  Control location  Relay  Serial port  Ext. Pump Setup Select " Setup > Global settings > Interfaces". ► © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 43 STOP and ZERO buttons on the device are deactivated. Select " Setup > Global settings > Interfaces > Control Location". ►  ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 44 Interface protocol for device diagnosis. Baud rate: 19200 RC protocol Remote control protocol. Baud rate: 9600 LD protocol Baud rate: 19200 ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 45: Load/Save Parameters

    The current parameters are saved as the indicated set name. Any parameter set saved as this name is overwritten. Load factory settings The factory settings are loaded. For more information, see Chapter : "Factory settings", page 12. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 46: Volume & Beep

    & vent", page 45)  Set the "Delay Timer"  Set the "Test Timer"  Set the "Background Limit" ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 47 Setup > Auto Test > Auto test parameters". 2. Select "External Trigger" and set the value to "enabled". After the "Background Limit" is reached, the device activates the PLC output (pin 20 on the D-sub plug). © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 48 The device determines the test result on the basis of the maximum leakage rate measuring during the measuring time. Process diagrams The following process diagrams show where in the Auto Test process the set values become effective. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 49 Is Fill Complete Input On/High? (Input #12) To Page 2 Fig. 12: Page 1 Auto Test Process Diagram**  ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 50: Settings For The Measurements

     Set leakage rate factor If a menu PIN is required to access the software menu, this function is inac- cessible until entered. ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 51: Set Filter & Zero

    In order to deactivate the "ZERO" function, press the ZERO button on the device or on the sniffer probe for longer than 3 seconds. ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 52: Set Vacuum Ranges

    Enabled Enabled Disabled Disabled 0.1 = 5 mbar Pressure at which valve V3 opens GROSS Enabled Enabled Disabled Disabled 0.1 = 25 mbar Pressure at which valve V1 opens © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 53: Define Evacuation Time & Vent

    Avoid permanent operation in this pressure range. ►  Changing the default settings may result in a considerable reduction of the performance of the device. Consult with LACO Technologies with any questions.  Note the default settings. To view or change presets, select "...
  • Page 54: Set Gross Leak Protection

    Gross leak protection is activated at the earliest once the alarm delay time elapses (see Chapter 6.4.7: "Set the setpoint and alarm", page 47). To view or change presets, select " Setup > Test settings > Gross leak ► protection". © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 55: Define Pressure Limits For Sniff Mode

    After pressing START, the acoustic signal can be activated as soon as the leakage rate falls below the warning limit or an alarm delay time elapses or the alarm type "Prop. Leakage Rate"/"Pinpoint" or "Sniff" operating mode is set. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 56: Define Calibration Settings

    Calibration leak value Calibration leak value in the selected unit. Depending on the (internal/external) selected calibration mode, this is either an external or the inter- nal calibration leak. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 57: Set Calibration Request

    "Calibration settings" (see Tab. 27: "Calibration settings", page 48). Internal calibration leak There are two variants for calibration with the internal calibration leak (see Chapter 6.4.8: "Define calibration settings", page 48): © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 58 1. In order to start the calibration, select "Calibration" in the "Standby" win- dow. The calibration progresses over multiple steps which are shown in the display. The first step is evacuation. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 59: 6.4.10.2 Calibrate In "Sniff"Operating Mode

    6.4.10.3 Check calibration with internal calibration leak This function is available for the operating modes "Vacuum" and "Auto Test**" with Mass 4. ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 60: Measuring

    3. In order to achieve the most precise possible measuring results, allow the device to warm up for approx. 30 minutes (see Chapter 6.1: "Switch- ing on the device", page 21). © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 61: Measuring In "Sniff" Operating Mode

    If you want to test a weld seam or similar, you must guide the tip along the path at a speed of less than 2.5 cm/s (1 in/s). When you check a spot, hold the sniffer line at it for at least 1 second. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 62: Executing An Auto Test

     Analog/digital with bar graph and large-format numbers or  Graphically as a function of measuring time ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 63: Calling Up Information About The Device

    For each error, the data and time, the error message number, and part of a comment are displayed. One of the error messages is always selected. You can move the selection up or down using buttons next to the arrow sym- © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 64: Switching Off The Device

    (10 atm.cc/sec range). For demanding applications where the leak rate requirements is below 10 atm.cc/sec, we recommend to have the vent valve disconnected. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 65: Warning And Error Messages

    – After the confirmation of this warning a warning triangle remains in the "Standby" menu, which signals the existence of this interruption. The warning triangle disappears as soon as this interruption has been corrected. W105 Global Reset – A global reset has been performed. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 66 – After the confirmation of this warning a warning triangle remains in the "Standby" menu, which signals the existence of this interruption. The warning triangle disappears as soon as this interruption has been corrected. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 67 – A software update has been performed and one or more new parameters determined. The message can be ac- knowledged in this case. The parameter(s) is (are) automatically initialized. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 68: Error Messages

    – Brief increase in pressure in the mass spectrometer. nominal value by over 10%. – MSV is defective. – MC 68 defective ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 69 E54 Foreline pressure p1 too low! – The output voltage of pressure measuring point p1 is too low – Check cable to Pirani sensor – Pirani sensor defective – Sensor electronics I/O board defective © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 70 (>5 mV) – Defective supply voltage pre-amplifier E71 TMP temperature control faulty – Short circuit in the temperature sensor E72 TMP temperature sensor faulty – Temperature sensor broken © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 71: Maintenance, Cleaning And Service

    Maintenance work on the T ™ may only be performed by person- ► ITAN nel who have been authorized by LACO Technologies to execute these tasks.  When handling the gases used, comply with the applicable regulations and safety measures.
  • Page 72: Information Sources

    "with Stop" must be set for "Venting" (see Chapter 6.4.4: "Define evacuation time & vent", page 45). 1. Select " Setup > Global settings > Maintenance & Service > Burn-in". 2. Make sure the test port is blank-flanged. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 73 PIN. The service PIN cannot be changed. After entering the correct service PIN, the Service menu may be accessed. Select " Setup > Global settings > Maintenance & Service > Service". ► © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 74: Maintenance And Service Schedule

    Replace Oil Wick SplitFlow Turbo Pump Maint. Kit T10004 8.6.1.1 Valve Block Clean/Rebuild LD Seals Kit T10040 8.6.2.1 Valves Rebuild Kit T10052 Cooling Fan Filter Clean/Replace Cooling Fan Filter T10023 8.6.3.1 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 75: Maintenance Work (M-Series Only)

    LACO Technologies to execute these tasks. Contamination of the vacuum area For all tasks in this area, wear clean, lint-free gloves. ► Use clean tools. ► © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 76: The System At A Glance (M-Series Only)

    14 MC 68 printed circuit board 6 Calibrated leak valve V5 15 MSV printed circuit board 7 Valve block 16 Speaker 8 Valve V2 17 Electrometer amplifier 9 Valve V3 18 Wiring board © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 77: Removing The Cover (M-Series Only)

    Replenish the oil (M-Series only) Required material: Pump Oil, 1 qt. Ordering no. LVOEZ1QT 1. Switch the device off at the mains plug. 2. Pull out the mains plug. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 78: Change The Oil (M-Series Only)

    13. Replace the oil drain plug. 14. Connect the grounding cable. 15. Replace the cover and secure it with an 8 mm Allen wrench on the left and right side. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 79: Oil Mist Filter (M-Series Only)

    10. Connect the grounding cable. 11. Replace the cover and secure it with an 8 mm Allen wrench on the left and right side. 12. Insert the mains plug again. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 80: Replace The Filter Of The Oil Mist Filter (M-Series Only)

    11. Replace the cover and secure it with an 8 mm Allen wrench on the left and right side. 12. Insert the mains plug again. For further maintenance steps see Chapter 8.6: "Maintenance work fur- ther steps (All Series)", page 78. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 81: Maintenance Work (P-Series Only)

    LACO Technologies to execute these tasks. Contamination of the vacuum area For all tasks in this area, wear clean, lint-free gloves. ► Use clean tools. ► © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 82: The System At A Glance (P-Series Only)

    14 MC 68 printed circuit board 6 Calibrated leak valve V5 15 MSV printed circuit board 7 Valve block 16 Speaker 8 Valve V2 17 Electrometer amplifier 9 Valve V3 18 Wiring board © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 83: Removing The Cover (P-Series Only)

    Fig. 20: How to remove the cover After the maintenance replace the cover and secure it with an 8 mm Allen wrench by turning it in the other direction. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 84: Backing Pump (P-Series Only)

    To change the mist filter on the P-Series, twist the upper mist filter hous- ► ing clock-wise to remove. Then, remove the retaining thumb screw to remove the filter cartridge. ► © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 85 Remove the retaining thumb screw to remove the filter cartridge, for pic- ► tures see the previous chapter. Replace the filter of the oil mist filter. ► © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 86 7. Carefully place the device on its back or left side (as seen from the front). 8. Use the special tool T10071 to unscrew the cover on the back. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 87 Make sure that the new O-ring is inserted in the correct position. An in- correctly mounted O-ring can result in gross leaks and a malfunction of the device. 15. Set up the device again. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 88 Fig. 24: Venting filter Venting connection 3. Loosen the cover on the left and right side using an 8 mm Allen wrench (see Chapter 8.4.2: "Removing the cover (M-Series only)", page 69). © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 89 Damage to the turbo pump from rotating rotor The rotor may no longer rotate. ► After switching it off, do not move the T ™ for at least 4 minutes. ► ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 90 5. In order to force away the ion source from the mass spectrometer, use one of the hexagon socket screws and screw it into the threaded hole in the ion source flange. 6. Pull the ion source out of the mass spectrometer. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 91 8.5 Nm of torque. The elevated groove inside the com- pression ring needs to be fitted on the ion source flange. 10. Slip the ion source connector back on and lock the connector using the cap nut. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 92 250 V, Ø5 x 20 mm). Fig. 28: Replace the mains fuses (2) 5. Lock the fuse holders back into place. 6. Close the cover. 7. Re-connect the mains plug. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 93 Sending in the T ™ for repairs ITAN You can send in your device to LACO Technologies for maintenance or repairs. For further information on this topic, see Chapter 10.2: "Sending in the TitanTest™", page 103. © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 94 T10014 PCB MSV Board T10015 Loud Speaker complete T10016 – Mounting Hardware PCB Wiring Board T10017 Chassis complete T10018 – Rubber Feet D=20; H=12; M6 – Handgrip – Latches © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 95 Fig. 30: T ™ P-Series specific, Chassis Front View * Wet Series Only ITAN Pos. Description Quantity Specifications Order No. P-Series external oil mist filter kit with external oil drain T109514 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 96 – Hose 4 mm x 1 mm L=500 mm – Hose 5 mm x 1,5 mm L=500 mm – Hose 10 mm x 3 mm L=500 mm 10** External Pump Connection Kit T10076 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 97 ™ P-Series specific, Back View *1 and 2 Wet Series Only ITAN Pos. Description Quantity Specifications Order No. 1 and 2* P-Series external oil mist filter kit with external oil drain T109514 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 98 T10071 Valve Block Inlet Plug & Gasket G 1/4 " T10077 Valve Block Inlet Flange & Gasket DN 25 - G1/4" T10078 Cooling Fan Filter Pore Size 500μ T10023 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 99 Fig. 35: TitT ™ M-Series, Housing View ITAN Pos. Description Quantity Specifications Order No. Cover Hood complete T10035 Screen Mat Cover Hood T10036 Display panel T10037 Front Cover without Display T10038 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 100 ™ ITAN Fig. 36: TitT ™ P-Series specific, Housing View ITAN Pos. Description Quantity Specifications Order No. – P-Series external oil mist filter kit with external oil drain T109514 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 101 T10041 Valve V6 2/2 way valve 24 V T10042 Valve V4 2/2 way valve 24 V T10043 Valve V3 2/2 way valve 24 V T10044 Pirani Gauge (each) T10045 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 102 Maintenance Kit for Valves V1 to V6 T10052 Maintenance Kit for Valve V5 (Calibrated Leak) T10067 Maintenance Kit for Valves V1, V2, V3, V6 T10068 Maintenance Kit for Valve V4 T10069 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 103 1 / 1 Ion Source T10059 Ion Source Cable T10060 TMP SplitFlow 80 with TC110 T10061 TMP SplitFlow 80 with TC110 (Refurbished Exchange) T10073 Cable TMP SplitFlow 80 T10062 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 104 Maintenance Kit & Set of O-Rings UNO 006 T10065 Overhaul Kit & Set of O-Rings UNO 006 T10066 Pump Oil 19 Grade, replace with Pump Oil Elite Z Synthetic 1 Qt. LVOEZ1QT © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 105 – Display gasket Replacement battery 200003316 Remote Control RC1000WL refurbished 551-015R Remote Control RC1000C refurbished 551-010R Coiled Cable 20099027 Radio Transmitter 551-020 Wall Power Supply for Remote Control 200003317 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 106 Maintenance Kit V5 (Cal. Leak) T10067 Maintenance Kit (V1, V2, V3, V6) T10068 Maintenance Kit (V4) T10069 Table 43: MAINTENANCE KIT FOR V5 (CAL. LEAK VALVE), P/N: T10067 Item Quantity Gasket Plunger/Spring © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 107 O-ring, 26.7x1.78 Table 46: TMP SPLITFLOW 80 MAINTENANCE KIT, P/N: T10004 Item Quantity Oil Wick SplitFlow 80 T10034 Oil Wick Cap SplitFlow 80 T10070 Oil Wick Cap Removal Tool T10071 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 108 WiFi Adapter for T ™ Communication Module TCM-WIFI ITAN Leak Test Accessories Order number Sniffer probe Sniffer Probe - 10 ft (3 m) LSP-01N-10 Sniffer Probe - 33 ft (10 m) LSP-01N-33 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 109 Internal sintered bronze filter LVF-B-2075-40-NW25 External filter (high flow) LVF-E-NW25-40-10 Accessories Order number Large chamber vent valve assembly T102752 Pass/fail light assembly T103181 Helium fill module assembly T104545 Chambers Order number © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 110 Pumping speeds 12.6 m3 /hr at 50 Hz, 15 m3 /h at 60 Hz Dry16-xxxx TriScroll 600 Two-stage dry (oil-less) scroll pump Pumping speeds 25.2 m3 /hr at 50 Hz, 30 m3 /h at 60 Hz See www.lacotech.com © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 111 10.1 Disposal of the T ™ ITAN The device can be disposed of by the operator or sent to LACO Technologies. The device is made of recyclable materials. You should use this option to avoid waste and to protect the environment.
  • Page 112 LACO ORIGINAL OPERATING INSTRUCTIONS T ™ ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...
  • Page 113 Starts the calibration or for confirmation of "Calibrate Acknowledge" (PIN 19)** PARA 2** Upon activation: "Loading parameter set 2."** Upon deactivation: "Loading parameter set 1."** Not used ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 114 Reference potential of the digital outputs, not galvanically insulated +24 V +24 V e.g. for actuating the digital inputs 0.8 A slow fuse  ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 115 The RS232 interface enables communications between the TitanTest™ and a computer. A terminal can also be connected for test purposes. TitanTest Fig. 44: RS232 Connector ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 116 PIN 5 D+ (galvanically isolated) PIN 6 GND (0 V) PIN 7 D- (galvanically isolated) PIN 8 not used ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 117 A twisted, screened two-wire cable is recommended for the connection! After receiving a command it takes about 5 to 10 ms until the T ™ ITAN sends a reply. © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 118 Connection for sniffer probe LP 503, LP 505, LP 510 RJ-45, 8-pin  Do not confuse the connectors! The connector is similar to the "RS485" connector. Fig. 49: Connector: RJ-45 (8-pin) © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 119  Binary transmission of numerical values – no conversion needed in PC/ PLC software  High efficiency (for example: 4 Byte data bytes for one float value) ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 120 LACO PROTOCOL DESCRIPTIONS T ™ ITAN Disadvantages:  Not human readable  Not useable with simple terminal program © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 121 Calculate CRC for all bytes (except CRC byte) Polynomial: 0x98, Name: DOWCRC, Maxim/Dallas, X +1 Info: CRC calculation see document "CRC_calculation.c" (C souce code) ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 122  Write single elements: Array index in first DATA byte and values in follow- ing DATA bytes  Read all elements: Pseudo array index 255 in first DATA byte © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 123 STANDBY 0011 VENT 0100 EVACUATION 0101 MEASURE 0110 CALIBRATION 0111 DISPLAY CAL 1000 ERROR 1001 WAIT EVACUATION 1010 not used 1011 not used 1100 not used 1101 not used © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 124 ™ ITAN Value Bit 3..0 Device state 1110 not used 1111 not used Value Bit 8..6 Meaning NO RANGE GROSS FINE ULTRA EVACUATION not used not used not used © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 125 FLOAT TMP temperature bearing [deg. C] 146 0092 TMP temperature motor [deg. C] FLOAT TMP temperature motor [deg. C] ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 126 / enumerations 261 0105 PLC input state UINT16 Get PLC input state see table / enumerations 262 0106 PLC output state UINT16 Get PLC output state see table / enumerations © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 127 UINT8 Present warnings Each bit represents a warning see table / enumerations HW-version UINT8[4] R HW-version Index 1: IO board Index 2: display Index 3: backplane Index 4: MC68 © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 128 394 018A Test leak intern [mbar*l/s] FLOAT R/W Test leak intern in mbar*l/s 401 0191 Operation mode UINT8 R/W Operation mode 0 = Vacuum 1 = Sniff 2 = Auto Test © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 129 (7 - g/a) (8 - oz/yr) 6...8 only in Sniff 430 01AE Pressure unit UINT8 R/W Pressure unit 0 = mbar 1 = Pa 2 = atm 3 = Torr © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 130 1 = 470 MOhm 2 = 15 GOhm 3 = 500 GOhm 4 = 13 MOhm (fixed) 5 = 470 MOhm (fixed) 6 = 15 GOhm (fixed) 7 = 500 GOhm (fixed) © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 131 R/W Evacuation time [s] 600 0258 Audio alarm type UINT8 R/W Audio alarm type 0 = Leak rate prop 1 = Trigger alarm 2 = Setpoint 3 = Pinpoint © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 132 5 = 1.6 (FIXED) 6 = 6.4 (FIXED) 7 = 25.6 (FIXED) 652 028C Postamplifier voltage [V] FLOAT Postamplifier voltage [V] 700 02BC AutoTest status UINT8 AutoTest status see table / enumerations © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 133 4, 5 = save 1, 2 2591 0A1F Control location UINT8 R/W Control location 0 = local 1 = RS232 2 = PLC 3 = local / RS232 4 = all © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 134 UINT8 maintenace list index (0...11). Without index you will get the last (new- est) entry. see table / enumerations 2660 0A64 Maintenace activ UINT8 R/W Maintenace activ 0 = disabled 1 = enabled © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 135 Wait "Test leak signal stable" Autotune Measure ULTRA Measure FINE Measure GROSS Wait "Test leak closed" or "Background stable" Background ULTRA Background FINE ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 136 Pin 20 Fill Helium 0x0040 Pin 21 Bypass valve 0x0080 Pin 22 No Leak 0x0100 Relay 1 0x0200 Relay 2 0x0400 ----- 0x0800 ----- 0x1000 ----- 0x2000 ----- 0x4000 ----- 0x8000 ----- © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 137 Warning calibration request 0x20 Warning capillary 0x40 Warning temperature electronic 0x80 Warning temperature preamplifier Valve state (command 449) and  Valve cycle counter (command 1350) Index Meaning VBypass ----- ----- © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 138 Evacuation GROSS State 4 State 5 FINE State 7 ULTRA State 9 State 10 State 11 State 12 Evacuation State 14 Evacuation (MS-Evac) State 16 State 17 State 18 © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 139 Factor GROSS FLOAT Calibrated leak [mbar*l/s] FLOAT Ion current (leak opened) [A] FLOAT Ion current (leak closed) [A] FLOAT  ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 140 UINT8 UINT8 Hour UINT8 Minute UINT8 Second UINT8 Operating hours UINT32 Hours UINT32 Component UINT18 0 = TMP 1 = Fore pump 2 = Ion source 3 = Device © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 141 ERR_CMD_NOT_ALLOWED Command actually not allowed (e.g. calibration during Run-Up) ERR_DATA Data not in range ERR_NO_DATA No data available  ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 142 Timeout between request to and answer from TitanTest: 1500 ms After sending a command the answer must be waited for before sending a new command. Otherwise the receive buffer may be overwritten. © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 143 Minimum audio volume *CONFig:MODE Mode (VAC, SNIFF, AUTO_TEST) *CONFig:PARTIALFlow:EVACuation Configuration of partial flow pump for evacuating (Fore_PUMP, Fore_AND_Partial_flow_PUMP, Partial_flow_PUMP) ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 144 Calibration factor sniff mass 2 *FACtor:SNiff:M3 Calibration factor sniff mass 3 *FACtor:SNiff:M4 Calibration factor sniff mass 4 *FACtor:VACUUM_Fine:M2 Calibration factor vacuum FINE mass 2 *FACtor:VACUUM_Fine:M3 Calibration factor vacuum FINE mass 3 © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 145 *IDN:MC68 Hardware identification MC68 *IDN:SERial Serial-number leak detector *IDN:TCNAME Name of TMP frequency converter *IDN:TURBO Software version TMP controller *IDN:VDversion Hardware identification mother board *IDN:VERsion Software version leak detector © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 146 AutoTest result in g/a *MEASure:RESULTLR:LB/yr AutoTest result lb/yr *MEASure:RESULTLR:MBAR*l/s AutoTest result in mbar*l/s *MEASure:RESULTLR:OZ/yr AutoTest result oz/yr *MEASure:RESULTLR:PA*m3/s AutoTest result Pa*m3/s *MEASure:RESULTLR:PPM AutoTest result in ppm *MEASure:RESULTLR:TORR*l/s AutoTest result in Torr*l/s © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 147 "TL_OPEN_ULTRA", "TL_OPEN_FINE", "TL_OPEN_GROSS", "WAIT_CLOSE", "TL_CLOSE_ULTRA", "TL_CLOSE_FINE", "TL_CLOSE_GROSS", "WAIT_RESULT" *STATus:CALHist Calibration history 1 to 12 (date, time, type of calibration, mode, calibration factors) *STATus:CALMode Kind of calibration (AUTO, MANUAL, EXTERNAL) © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 148 2641 R *STATus:ZERO Zero (ON, OFF) *STOp Stop (switch from Measure to Standby) *VENt Vent inlet port *ZERO Switch zero on *ZERO:OFF Switch zero off *ZERO:ON Switch zero on © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 149 OK (CR) start measurement *conf:trig1? (CR) 1.0E-9 (CR) retrieve trigger 1 *conf:trig1 2.0E-9 (CR) OK (CR) set trigger 1  ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 150 2 illegal command word 3 illegal control by RS232 not enabled argument faulty no data available error buffer overflow command invalid query not allowed only query allowed not yet implemented © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 151 Parity, Stop bits) IO1000 and PC/PLC match) Wrong protocol selected in Select correct protocol in the leak detector the leak detector ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 152 ESC, ^C correctly ging in the RS232 cable or ^X during operation) ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - TitanTest itra01en1-07 SB.fm - 1807...
  • Page 153 Information – Calibration history 56 Protocol descriptions 111 – Device 55 – Error list 55 – Settings 55 Relay – Vacuum diagram 55 – Interface 110 Input/Output interface 105 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 SIX.fm] - 1807...
  • Page 154 Venting – Evacuation and venting settings 45 – Line 20 – Menu in Standby window 28 – triggering manually 28 Warning messages 57 Warning symbols 4 ZERO button 23 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 SIX.fm] - 1807...
  • Page 155 LACO ORIGINAL OPERATING INSTRUCTIONS T ™ ITAN ** Feature not available in T ™ Maintenance model ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 SIX.fm] - 1807...
  • Page 156 LACO ORIGINAL OPERATING INSTRUCTIONS T ™ ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 SIX.fm] - 1807...
  • Page 157 LACO ORIGINAL OPERATING INSTRUCTIONS T ™ ITAN © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 last.fm] - 1807...
  • Page 158 © 2018 LACO TECHNOLOGIES, INC. - [TitanTest itna01en1-07 BA.fm] - 1807...