Page 1
SurePrint Slide Printer Instructions for Use V3.0 - English Version - Apr. 2023...
Page 2
Product name: Slide Printer Product model: S200/S200Pro/S100 Date of manufacture: See the label Name of manufacturer: Dakewe (Shenzhen) Medical Equipment Co., Ltd. Address of manufacturer: Floor 5, Building B, No.2 Luhui Road, Jinsha Community, Kengzi Street, Pingshan District, Shenzhen, China Telephone: 0755-26413421 Service provider: Dakewe (Shenzhen) Medical Equipment Co., Ltd.
Page 3
Foreword Dear users, Thank you for choosing our products! In order to give you an overall understanding of our products and facilitate your use, we have specially provided this user manual of the instrument, which includes the structural features, technical features, operation instructions, simple troubleshooting, maintenance, etc.
Page 4
This document is protected by copyright law. All copyrights for this document are reserved by Dakewe (Shenzhen) Medical Equipment Co., Ltd. No text and illustrations (or any part thereof) of this document shall be reproduced by printing, photocopying, microphotography, webcam or other means, including any electronic system and media, without the explicit written permission of Dakewe (Shenzhen) Medical Equipment Co., Ltd.
Contents I. Information ..............................1 (1) Symbols and Meanings ..........................1 (2) Intended Use of Equipment ........................3 (3) Personnel Qualifications ..........................3 II. Safety ................................4 (1) Safety Precautions ............................4 (2) Warnings ..............................4 (3) Safety Functions ............................6 III. Initial Installation ............................8 (1) Device Overview .............................
Page 6
VI. Cleaning and Maintenance ........................63 (1) Routine Cleaning of the Whole Instrument ...................63 (2) Remove glass debris ..........................63 (3) Replace filter ............................65 (4) Other Cleaning and Maintenance ......................67 VII. Abnormalities and Troubleshooting ......................68 (1) Fault Alarm Types ..........................68 (2) Faults Handling ............................68 (3) List of Common Abnormalities ......................
If the use of a restricted substance exceeds the maximum allowable limit, this symbol is used. The item indicator number in the illustrations. TILTXINKE: Keep upright. SHOCKXINKE: Handle with care. Forklift caution sign: notice- lifting gear lifts from this side. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 8
The recycling and disposal shall meet the requirements of local laws and regulations. The CE marking is the manufacturer’s declaration that the medical product meets the requirements of the applicable EC directives and regulations Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
I. Information (2) Intended Use of Equipment The Dakewe Slide Printer is a professional printer specially designed for histology slides, which is used to print information on the slides: barcode, QR code, numbers, characters, multiple languages, graphics. The equipment can only be operated according to the instructions in this manual. Any other operation is considered inappropriate! ...
It must be used as specified and only when all safety functions are in proper operating state. Faults that compromise safety must be corrected immediately. (2) Warnings The safety devices installed by the manufacturer on this equipment constitute only the basis for Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 11
3. Warning - Operation The Dakewe slide printer should only be operated by trained laboratory personnel in accordance with the intended use and instruction manual. In case of emergency, turn off the main power and unplug the power switch.
1.Indicator light and Hopper Lights The Dakewe Slide Printer is equipped with a status indicator light and hopper lights. The status indicator is located above the touch screen. Users can know the status of the instrument in real time based on the color change of the status indicator.
Page 13
II. Safety 2. Alarm The Dakewe Slide Printer is equipped with an alarm function, and the alarm is given from the laser inside the instrument. When the temperature of the laser inside the instrument exceeds 40°C, the laser alarm system of the equipment will send out a continuous short tone of “Beep...”.
III. Initial Installation (1) Device Overview The Dakewe Slide Printer is a professional printer specially designed for pathological tissue slides. It can print English letters, numbers, Chinese characters, one-dimensional codes (barcodes), QR codes, pictures (bitmap formats supported by the current system: .DXF, .PLT, .SVG) on the painted surface of the glass slide.
(3) Remove the foam in the box, and take the instrument out. (4) To reassemble the crate: press the fasteners into the slots on the box based on the length direction. Figure3- 7 Fasteners Figure3- 8Fastener Press-fit Point Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 17
100 slides in each . The Dakewe Slide Printer S200 and S200Pro will consume the slides in the input hopper A during printing. When the input hopper A is not loaded with slides or the slides are exhausted, the software will automatically detect the slides availability in input hopper B;...
Page 18
A (Figure 3-21) is low on the left and the right is high, and the input hopper B (Figure 3-18) is high on the left and low on the right. Batch output hopper has equally high columns. Three hoppers cannot be used exchangeably. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
2 External power cable connector 2. Turning on the instrument The following six steps are required to start the Dakewe Slide Printer: (1) Connect the power switch of the instrument to a grounded power outlet. (2) Turn on the main power switch on the back of the instrument and keep the switch at the “丨” end.
Page 20
3. Turning off the Instrument The following three steps are required to completely turn off the Dakewe Slide Printer: (1) Long press the ON/OFF button in the front of the instrument for 3 seconds so that the software will pop up the account logout prompt to enter the shutdown process.
(1) Description of Print Method 1. Print mode The Dakewe Slide Printer S200 has three options for print mode: on-demand mode, batch mode, and auto mode, which can be changed on the homepage or [Menu] - [Print Settings] - [Others] interface. The S200Pro has the batch mode only, and the S100 has the on-demand mode only.
Page 22
Figure 4- 4 S200Pro Batch output hopper 2. Print channels The Dakewe Slide Printer S200 and S200Pro have three options for print channel: Both Hoppers, Hopper B, and Hopper A. If you want to change the print channel, you can configure the “Hopper” to the homepage and change it on the homepage, or you can enter the [Menu] - [Print Settings] - [Others] interface to change it.
Page 23
Figure 4- 5 Printing effect of pathology IDs 4. Print Order The Dakewe Slide Printer has three print orders: By Main ID Reversive, By Main ID Sequential, and By Sub ID No. Sequential, which can be changed on the [Menu] - [Print Settings] - [Others] interface.
Figure 4- 7Serial Ports of Scanner Step 3: Enter the [Menu] - [Print Settings] - [Others] interface (Figure 4-8), check “Scanning Data Parser”, and select the corresponding scanning parser rule type to enable the scanning print function.(Figure 4-9) Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 25
8 fields concatenated in sequence (Figure 4-11). [Note]: If the field to be parsed contains the "Prefix", "Suffix", "Custom1", "Custom2", "Custom3", "Custom4" and "Custom5", and "Pathology ID Concatenate" is applied, the concatenated content displayed Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 26
(see 4.2.3 Config Template-Type Correlation); Step 3: Create the corresponding scan parsing rule. Figure 4- 10 Config Template-Type Correlation Figure 4- 11Create a new scan parsing rule - Type Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 27
Step 4: Select a template to determine the template that is used to print the marked ID of this type; Step 5: Click “Add”, click “Apply”; Step 6: Return to the homepage, select type, add ID, and print. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 28
IV. Instructions before Use Example 1: When printing an ID on S200 using the input hopper A for the "DAKEWE(Chinese)" template or input hopper B for the "DAKEWE(English)" template, the configuration is as follows: After the configuration is completed, select the type on the homepage. If the user select “A-Chinese” and add ID “2200001-A01”, this ID will be printed using Hopper A slides and “DAKEWE (Chinese)”...
Page 29
Pathology ID example (Figure 4-17 [Menu] - [Print Settings] -[Pathology ID Setting] interface). Figure 4- 16 Create a new scan parsing rule interface Figure 4- 17 Pathology ID example Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
It is located at the top of the homepage and mainly displays the following information: Table4- 3 Status bar components Components Position/Description logo DAKEWE logo (displayed on the left) It is located in the middle of the status bar. There are eight print Status statuses: Ready, Printing, Stopping, Resetting, Abnormal, Sleep, Processing, and Disconnected To the right of date and time.
Page 31
Figure 4- 19 Homepage Config Step 2: Select the item you want to add to/delete from the homepage (Figure 4-20); Figure 4- 20 “Add”/“Delete” on the Homepage Step 3: Click “Add”/”Delete”; Step 4: Click “Apply”. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 32
Print Jobs List (Figure 4-22): It displays the count of IDs printed and to be printed on that day and the ID pending list. The user can check the pending ID or check “All” to perform the operations of “Delete” or “Prior” for ID. Figure 4- 22 Homepage - Status Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 33
Figure 4- 23 Rename 7. Virtual keyboard The Dakewe Slide Printer can use the virtual keyboard that comes with the software to input information, or use an external mouse and keyboard. The user can enter Chinese characters, English letters, numbers, and special characters using the virtual keyboard or external keyboard; and can also use the shortcut key combination of “Ctrl + Enter”...
Caps lock button: Press this button to switch between uppercase and lowercase. (4) System Settings The System Settings interface (Figure 4-27) consists of multiple system and prompt-related settings, which can be fully viewed by swiping down the scroll bar (Figure 4-28). Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 35
Timeout of time you set To modify the current date format displayed in the Date Format Status Bar of homepage Figure 4- 27 System Settings 1 Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Sub ID: The Sub ID can only be printed and displayed in the composition example when the user clicks “Sub ID”. Date: After the user clicks “Date”, three drop-down boxes will appear at the bottom of the ID Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 37
“Main ID Concat.” is used to set the main ID in the example on the ID Setting interface (Figure 4-31). The main ID is composed of four fields concatenated in sequence. Each field can be set as “ID”, “Year”, “Custom”, or “Empty”. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 38
Figure 4- 32 Sub ID concatenation The user can modify the format of the Sub ID No. to display 1 to 5 digits in the drop-down box. For Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 39
The Others interface includes three parts: ID Generation Rules, Scanning Data Parser and Printing Rules. ID Generation Rules “ID Generation Rules” is mainly used to set the display of the ID, including the following settings (Figure 4-34): Print order, see 4.1.4 Print Order. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 40
Hide the 1st sub ID No.: If the 1st sub ID No. of the main ID is hidden, the 2nd sub ID No. will display (Figure 4- 38). C2023-12345-A Fr Single print C2023-12345-A01 C2023-12345-A02 Fr Batch printing Figure 4- 38 Hide the 1st sub ID No. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 41
After scanning, the Pathology ID will not be added to the list of print jobs immediately; instead, the Edit scanned data scanned content will be parsed using the current scan parsing rule and displayed to the corresponding input Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 42
Start printing immediately after the instrument is reset instrument is reset Start printing immediately after adding a job Print after adding a job manually Print immediately after Start printing immediately after scanning a scanning a code code Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 43
"Edit" button to perform real-time visual editing of the print template (Figure 4-44). This manual will explain the functions in detail. Here are the common buttons (Table 4-8): Figure 4- 43 Print Settings - Others Figure 4- 44 Template Edit Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 44
The template preview area has a two-dimensional orientation map displaying the real time preview of the template chosen in the print parameters. Objects preview The text input box above the template preview area is the object preview editor. For example, the Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 45
Select the “title” object in the object list on the left; Enter "DAKEWE" in the content and click "Save" (Figure 4-45). Print parameter settings To accommodate different materials,...
Page 46
Double-click the object name in the list to modify the object name. Delete the object Check the objects to be deleted in the list and click the " " button to delete them (Figure 4-48). Figure 4- 48 Delete the object Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 47
Spiral inwards: The pattern is cycled inward from the outer outline of the object to offset the fill. Castle line: Similar to a back-and-forth fill, but with connecting lines between the ends of the fill lines. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 48
" button to flip all the objects along the vertical axis and the result is shown synchronously in the template preview; Step 3: Click the " " button to flip all the objects along the horizontal axis and the result is shown Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
“Save” button to take effect after setting. Account levels The Dakewe Slide Printer has 3 levels of permission: Operator, Admin, and Root. The Admin can only be created by the Root, and the operator can only be created by the Admin.
Page 50
Step 1: Click the inverted triangle icon in the status bar and select “Switch User” from the drop-down list to pop up the dialog box of “Switch User” (Figure 4-53); Step 2: Select the corresponding user name and enter the password, and click “Confirm” to log in (Figure 4-54); Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 51
Click "Username" under the Status Bar and select "Log out" in the drop-down box so that the account will be logged out and switched to the Visitor (User) role and the software homepage will be restored. Figure 4- 55 Add user interface Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 52
For the user to be deleted, click the “Delete” button in the operation bar so that a prompt box for confirming the deletion will appear. Click “Confirm” to confirm the deletion, and click “Cancel” to cancel the deletion. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
The exported data is stored on a USB drive in Excel format. The file name is in the format of yyyy-mm-dd_ExportData_serial number, for example, “2022-08-31_exportdata_1”. Figure 4- 58 Data management interface Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Please check the appointed path in the USB drive!” will appear (as shown in Figure 4-59). If the detection is successful, a message saying that the detection is complete and manual confirmation is required will appear (as shown in Figure 4-60). After the confirmation, the installation starts; Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
“Main Firmware”, select the update file for the main firmware, and wait until the update is complete. (9) Help The interface is not yet unavailable, and no operations can be performed. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Set the device IP address in the properties window of Internet Protocol Version 4 (TCP/Ipv4): Step 1: Click the network icon ① in the taskbar of the device, and select "Network and Internet Settings" ② (Figure 5-2); Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 57
Step 5: In the "Internet Protocol Version 4 (TCP/Ipv4) Properties" window (Figure 5-6), select "Use the following IP address"; Step 6: Enter the IP address data in the corresponding input box, such as: IP address: 192.168.113.19 Subnet mask: 255.255.255.0 Default gateway: 192.168.113.1 Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 58
Note: AX88772C_Win... is the name of the USB Ethernet Adapter Driver, and the name will be displayed only when it is used. Before setting "Ethernet Properties", it should be confirmed that the network is the same network that the device is connected to . Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 59
After entering the IP address data, click the "OK" button. The IP address data can be changed according to the actual need, but it must be ensured that the IP address of the device and the computer are in the same network segment. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
The homepage of the PC Software mainly contains seven parts (Figure 5-12): Top Status Bar, Side Menu Bar, Print Settings, Pathology ID Settings, Hopper Status, Template Preview, and Print Jobs List. Figure 5- 12 Homepage Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 61
Table 5- 1 Top status bar components Components Position/Description Logo DAKEWE logo (displayed on the far left) After clicking, the side menu bar only displays the Side menu bar minimize icon interface icon, but does not display the interface name It is located at the middle of the status bar.
Rotate preview pixmap: the slide dynamic display style in the template preview on the home page. The template preview can be rotated by 0°, 90°, 180° and 270°. (6) Interface Settings It is used to set Hospital Name, Connect to the Instrument Software, and Connect to the Pathology Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Interface settings for the device IP; After setting, click the "Test" button. If "Connection succeeded!" is displayed (Figure 5-15), the device IP is set successfully; If "Connection failed!" is displayed, please reset it; Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 64
Step 3: Enable the parse function of Pathology Information System: Click and configure the parse rules of Pathology System. Step 4: Enter the New parse rules interface, Detailed operation steps: Select “Parser” Select “New” to create a new parse rule file (Figure 5-18); Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 65
File name: It refers to the name of the parse rule file, which can be set by the user according to the using habit; Parse file type: There are two options: text and table, which support the .txt and .xlsx formats, Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 66
Figure 5-20; fill in the index count according to the correlation, and click “Apply” to add. Figure 5- 20 New parse rules-Correlation and segmentation Figure 5- 21 Correlation of .txt format-data Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 67
Type resolution: You can configure the feed box and print template for the printed pathology number by analyzing the type. See 4.2.2 Parse the Filed “Type”. Figure 5- 23 Correlation of .xlsx format-data Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
On this interface, the user can update the software and view the current version information (Figure 5-25). Note: During online update, the user needs to update the PC software first and then the instrument software. Figure 5- 25 PC software - Update Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
The areas to be cleaned include the print platform, input/output hoppers, input/output hopper holders, input/output hopper base, input/output hopper slide and base plate, and on-demand output. Clean glass shards and debris in these areas as well as the surrounding areas. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 70
Step 3: Move the transfer push rod rightward to the right side of the print platform (Figure 6-10); Step 4: Use tweezers to remove glass pieces shards, and then use brushes to remove the debris. Figure 6- 9 Maintenance door Figure 6- 10 Transferring push rod Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
(with the ribbon facing outwards and the folds facing upwards); Step 3: Cover the front cover (hide the ribbon), assemble the disassembled screw again, and close the maintenance door. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 72
Step 3: Then replace with a new filter (with the ribbon facing upwards, and the filter placed as close to the front as possible), and install the top cover back (hide the ribbon properly). Figure 6- 15 Back filter Figure 6- 16 Filter - back Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Clean the detection surface of the laser distance sensor with a dust-free cloth (once a year). 3. Screw Rod Clean off the oil on the screw rod with a dust-free cloth, and reapply the lubricating grease (once a year). Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Display on interface (2) Faults Handling 1. In case of a fault of the Dakewe Slide Printer, take the following steps to deal with it: Read error messages; Take action according to the instructions in the message prompt;...
Turn on the instrument within a short Fail to startup the Turn on the instrument 1 min after turning time after turning off the main power instrument off the main power supply supply Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Cause analysis The detection function of the on-demand output is not turned on; The no-slide parameter threshold of the on-demand output is not set or unreasonably set. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 77
(3) Pull out the input hopper on one side (Figure 7-5), observe the stuck slide from the angle of this input, remove the slide stuck in the instrument with a tweezer (Figure 7-6). Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 78
Note: Move the transfer push rod to the right when printing the slide on the right, and move to the left when printing the slide on the left; (2) Move the slide to the print position (Figure 7-9); Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 79
(2) Check whether the slide edge is broken; if so, take out the slide with broken edges (Figure 7-12); Figure 7- 12Slide with broken edges (3) Check whether the slides are regularly shaped; if there are irregularly shaped slides, take out the them (Figure 7-13); Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 80
Temporary resolution: Pause the use of the on-demand mode and auto mode; if they must be used, remove the slide as soon as it is present in the on-demand output to avoid problems; Fundamental resolution: Contact the engineer to troubleshoot the detection sensor of the on-demand output. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 81
Resolution Enter the half width comma in the preset option box as the separator (Figure 7-20). Figure 7- 19 Improperly using full width comma as Figure 7- 20 Preset of English comma separator Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 82
8. Instrument software disconnection Abnormality Description The software connection icon at the top of the instrument interface is gray, and all buttons are gray; no operation can be performed. Figure 7- 22 Instrument disconnection Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 83
Enter “Interface Settings” interface of the PC software, and click “Test” on the right of “Instrument IP”; if “Failed to Connect!” pops up, reconfigure the IP address; see 8.2 IP Address Settings. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 84
(2) Replace the filter element; see 6.3 Filter Element Replacement. Figure 7- 25 Abnormal display of the print effect Figure 7- 26 Non-paralleled print effect 11. Offset print result Abnormality Description The print is not parallel to the slide edge (Figure 7-26). Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 85
Turn on the instrument within a short time after turning off the main power supply. Resolution Turn on the main power switch on the back of the instrument 1 min after turning off the main power supply. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Save changed settings Fill Slides All hopper are unlocked, and slides can be added and taken out. Print The instrument prints according to the print jobs list. Add print jobs according to the set printing. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Appendix 3: Warranty (I) Warranty Dakewe guarantees that a comprehensive quality control procedure based on Dakewe internal test standard is used for the contract products delivered, and that the product is intact and meets the requirements of all technical specifications.
According to 21CFR 1040.10, laser products are classified into 5 levels.The Dakewe slide printer embedded in Class 4 (wavelength :355nm) and Class 2 (wavelength: 655nm) laser parts, but the product by taking protective measures, completely closed laser radiation, its backfire is Class 1.
B. In the test, the function or performance is degraded or lost temporarily, but can restore automatically. C. In the test, the function or performance is degraded or lost temporarily, and cannot restore until reference or system operation by the operator Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 92
Do not use this device in proximity to sources of strong electromagnetic radiation (e.g. unshielded intentional RF sources), as these can interfere with the proper operation. Dakewe (Shenzhen) Medical Equipment Co., Ltd. www.dakewemedical.com...
Page 94
Manufacturer: Dakewe (Shenzhen) Medical Equipment Co., Ltd. Service provider: Dakewe (Shenzhen) Medical Equipment Co., Ltd. Address: Floor 5, Building B, No.2 Luhui Road, Jinsha Community, Kengzi Street, Pingshan District, Shenzhen, China Address of manufacturer: Floor 5, Building B, No.2 Luhui Road, Jinsha Community, Kengzi Street,...