RETURN OF GOODS TO MANUFACTURER If equipment needs to be returned to Ludlum Measurements, Inc. for repair or calibration, please send to the address below. All shipments should include documentation containing return shipping address, customer name, telephone number, description of service requested, and all other necessary information.
Model 42-49 & 42-49B Neutron Detector Technical Manual Table of Contents Introduction Unpacking and Repacking Connecting to an Instrument Specifications Safety Considerations Environmental Conditions for Normal Use Cleaning Instructions and Precautions Calibration Calibration Operating Voltage Plateau Gamma Rejection Check Conversion Chart...
He→ H + 0.764 MeV. The Ludlum Model 42-49 and 42-49B utilize this reaction in the form of helium-3( He), which fills the gas proportional tube of the detector The Model 42-49 and 42-49B are designed to be used with portable counting instruments and has a top bracket that allows for convenient mounting of a portable instrument.
Page 6
Shown above are the two different boron moderators attached to the detector, which fit into a rem ball. The one on the left is a 42-49, and the one on the right is a 42-49B. Ludlum Measurements, Inc.
Every returned instrument must be accompanied by an Instrument Return Form, which can be downloaded from the Ludlum website at www.ludlums.com. Find the form by clicking on the “Support” tab and selecting “Repair and Calibration” from the drop-down menu.
Model 42-49 & 42-49B Neutron Detector Technical Manual Section 4 Section Safety Considerations Environmental Conditions for Normal Use Indoor or outdoor use (in a dry environment) No maximum altitude Temperature range of -20 to 50 °C (-4 to 122 °F); may be certified for operation at -40 to 65.6 °C (-40 to 150 °F)
Model 42-49 & 42-49B Neutron Detector Technical Manual Section 5 Section Calibration Calibration To calibrate the detector, the following items are needed: Counting instrument: Ludlum Model 2200 Scaler/Ratemeter AmBe neutron source or equivalent Cs gamma source for gamma rejection check Operating Voltage Plateau Connect the Model 42-49 to a Model 2200.
Model 42-49 & 42-49B Neutron Detector Technical Manual Section 5 For example, an assumed operating voltage is 1100 volts, based upon the flattest part of the plateau. The count rate at that voltage is 2380 counts per minute (cpm), and the neutron field dose-equivalent rate is 20 rem/hr.
Model 42-49 & 42-49B Neutron Detector Technical Manual Section 5 Conversion Chart Expose the detector to an AmBe neutron source at a dose- equivalent rate of 400 mrem/hr. Take a one-minute count and record the value, including the range/scale setting of counting instrument.
Model 42-49 & 42-49B Neutron Detector Technical Manual Section 6 Section Parts List Reference Description Part Number Model 42-49 UNIT Completely Assembled Neutron Detector Model 42-49 Neutron Detector 47-4057 2 EA. SCREW-6-32 X 3/8 BRASS KNURL 17-8659 Model 42-49 SPHERE ASSY 47-4058 4 EA.
Model 42-49 & 42-49B Neutron Detector Technical Manual Section 7 Section Drawings and Diagrams Model 42-49 Overall, Drawing 14 x 584 Model 42-49 Probe Assembly, Drawing 14 x 587 Model 42-49B Probe Assembly, Drawing 14 x 620 Ludlum Measurements, Inc. Page 7-1 October 2021...