Technical Specifications - Sennheiser ADC1 Instruction Manual

All-day clear otc hearing aids
Table of Contents

Advertisement

provide "normal use"; a sum of 6 or greater would
indicate "excellent performance". The equipment
performance measurements, categories and system
classifications are based upon the best information
available but cannot guarantee that all users will be
satisfied. The immunity of this hearing aid is at least
M2/T2.
the performance of individual hearing aids may
vary with individual cell phones. Therefore,
please try the hearing aid with your cell phone or,
if you are purchasing a new phone, be sure to try
it with your hearing aid prior to purchase
Important information in case of battery ingestion
If a battery is swallowed, consult your physician
immediately.
60 |
All-Day Clear

Technical specifications

Measurements
Maximum output limit value (OSPL90)
Maximum output (OSPL90) curve
Full On Gain Value @50 dBSPL input
Frequency Response Curve
.
Output Distortion
100dB SPL Out
70dB SPL Out
Input Distortion @ 100dB SPL In
Self Generated Noise value
Latency Value
Upper and Lower Cut-off Frequencies
for Bandwidth
Value
Reference
Standard
110 dB SPL
ANSI S3.22
2014
see Figure 1
34 dB
ANSI S3.22
2014
see Figure 2
1.0%
ANSI/CTA
2051 2017
4.0%
ANSI/CTA
2051 2017
1.0%
ANSI/CTA
2051 2017
25 dB SPL
ANSI/CTA
2051 2017
6.5 ms
ANSI/CTA
(milli-
2051 2017
seconds)
100 Hz to
ANSI/CTA
5000 Hz
2051 2017
All-Day Clear