Contents Symbols used ....................3 General information ..................3 Warranty provisions ....................3 Technical data ......................4 Original package contents ..................4 Tested accessories ....................4 User group and environment ..................5 Indication and intended use ................5 Contraindication .................... 6 Safety notes and warnings ................
Symbols used Refer to instructions for use ON/OFF (press resp.) Consult instructions for use Zero adjustment Caution Power supply control lamp Manufacturer Battery status indicator Serial number Equipotentiality Catalogue number Type BF applied part NRTL approval mark for USA (US Date of manufacture testing company SGS) Separate collection for WEEE (Waste...
The digital display then shows the mean as well as the systolic and diastolic intracranial pressure (ICP). The ICP monitor output provides both the mean pressure as well as the pulsatile signal. After 10 minutes and once hourly thereafter, the ICP monitor opens the pressure transducer to atmospheric pressure for zero adjustment.
Page 7
Routinely check electrical plugs and cables. Do not use if they are damaged. Do not immerse the cables in liquids and do not bring the electrical connection into contact with liquids. Do not sterilize the cable in an autoclave. HDM29.2 - Front panel...
Application and handling In the following, the application parts are referred to as "probes". This term includes both the ICP probes and the IAP catheter. The individual steps apply equally to both product groups. Setting up the ICP monitor The ICP monitor must be placed on a level surface in order to be used for its intended purpose. The display panel must be freely accessible to the user.
ICP monitor and contact the manufacturer. Connecting to a patient monitor NOTE! Only cables authorized by Spiegelberg may be used for connecting the ICP monitor. 1. Connect the monitor cable to the rear connector "Monitor" (11) and the ABP input of the patient monitor.
Maintenance must be carried out in accordance with the instructions specified by Spiegelberg. These are available from Spiegelberg upon request. Alternatively, the ICP monitor can be sent to Spiegelberg for maintenance purposes. Spiegelberg reserves the right to charge the maintenance costs to the customer.
Disposal The ICP monitor must be disposed of by separate waste disposal for electrical and electronic equipment. If necessary, contact your dealer or the manufacturer. Faults and errors If problems occur with the ICP monitor, error messages are shown in the display (3). The following table briefly explains the error messages, their causes and possible solutions for correction.