Cable Termination - ABB AnalyzeIT 7600 Series Instruction Manual

Ph/orp electrode system
Table of Contents

Advertisement

...4 ELECTRICAL INSTALLATION
4.3 Cable Termination – Fig. 4.3
If the provided standard length cable requires shortening, the
recommended preparation is as shown in Fig. 4.3.
Strip back the overall cable sheath to measurements as shown
and fit suitable sleeves and solder tags to the leads, for
connection to the transmitter.
3.2 mm i.d Heat Shrink Sleeving
Screen
Co-axial Cable Preparation
6mm i.d. Heat
Shrink Tubing
Black
Grey
Blue
Green
30mm
30mm
60mm
Fig. 4.3 Cable Termination
14
Trim back black conductive
layer to here
Wire, 16/0.2, PVC covered,
soldered to screen
Measuring
Electrode
(Inner Core)
Screen
Fly Lead
Ref
Temp.
Comp.
Solution
Earth
White
Red
35mm
40mm