Upper arm blood pressure monitor User manual (Apply to:Q06、Q06B) Edit:2020-11-22 Version: A/0...
Page 2
Web:www.wellkang.ltd.uk & www.CE-marking.eu Email: AuthRep@CE-marking.eu Version Information This manual may upgrade due to software upgrading. User will not be notified further. ■ Version: A/0 ■ Issue date: Nov. 2020 Software version:V1 Shenzhen Hingmed Medical Instruments Co., Ltd. owns and reserves the copyright.
3.3 Display .............................8 3.4Mark content and meaning ......................8 4.Blood pressure measurement ........................9 4.1prepration ............................9 4.2 Taking a measurement(Apply to Q06B & Q06) ................9 4.3Taking a measurement by the App (Apply to Q06B) .................10 6.maintenance ..............................11 7.Disposal ................................12 8、Component replacement list ........................
1.Introduction 1.1product feature Thank you for purchasing the HINGMED upper arm blood pressure monitor.This monitor can be used by multiple users,This blood Pressure monitor use the oscillometric method of blood pressure measurement.It has some functions that measure the blood pressure and display the result.When measuring the blood pressure,the user need to wear the cuff on the...
Model Q06B /Q06 Measuring method Oscillometric method Range From 0 to 289mmHg Pulse range 40-200bpm Resolution ratio 1mmHg Accuracy ±3mmHg (pulse rate ±3bpm or ±3% whichever is grater) Power DC 3.7V Li-ion Cuff size 22cm-41cm or 22cm-36cm Size L:116*W:58*H:17.8 5 ℃ -40 ℃ ;...
Page 6
*please use the adapter which output DC 5V 2A to charge the device When the device start inflating may have a little uncomfortable. Do not leave the device unattended with infants or person who can express their consent. Do not apply strong shocks and vibrations to or drop the device and arm cuff. Do not take measurements after bathing,drinking alcohol,smoking,exercising or eating.
Page 7
* In case of equipment failure, the error code will be displayed in the format of "EC XX", see the section on Common Failures for details. * State that the blood pressure measured should be explained by a professional. The product is not intended for newborns. *The specifications and labels of this product meet the requirements of EN1060-1 and EN1060-3 *Frequent inflating can lead to blood clots and discomfort...
*Three hundreds of discharge cycles after which a rechargeable battery needs to be replaced * Please put the device keep away the children and pet The list of device that can potentially cause interference problem,eg. Microwave oven; 3. Product structure and operation principle 3.1The component of product Compment Count...
(2) Please press the "start/stop" button to wake up the device and the time is displayed . (3) Wear the cuff ,refer to the right picture. 4.2 Taking a measurement(Apply to Q06B & Q06)
(1)press the "start/stop" button,the cuff start to inflate automatically,increasing number on the display ,remain still and do not move your arm and until the measurement process is complete. (2) When the measurement is complete,your cuff completely defeat, your blood pressure and pulse rate are displayed and broadcast.
The measurement timeout, in adult mode, Wait 2-3minutes EC09 exceeded 120 seconds Take another measurement Push the “start/stop” button to Stop Push the “start/stop”button to take another EC 11 measuring measurement EC26 Battery is worn Please recharge If there is other error code ,please contact customer service. 6.maintenance ************************************************************ Note: * Sphygmomanometers and accessories need not be sterilized, but should be kept clean.
The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life. It is generally recommend to have the device inspected every two years to ensure correct functioning and accuracy.please consult your local authorised distributor or dealer. Replace the battery -Remove the setting screw as shown on the right - Use the new battery to replace the old battery...
Hingmed descriptions, calibration instructions, or other information that will assist service personnel to repair those parts of ME equipment that are designated by Hingmed as repairable by after sale SERVICE PERSONNEL .
10 EMC information This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment. Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
Page 15
Guidance and manufacture’s declaration – electromagnetic immunity the device is intended for use in the electromagnetic environment specified below. The customer or the user ofdevice should assure that it is used in such an environment. Electromagnetic environment - Immunity test IEC 60601 test level Compliance level guidance...
Page 16
environment. NOTE is the a.c. mains voltage prior to application of the test level. Guidance and manufacture’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment. Immunity IEC 60601 Compliance...
Page 17
marked with the following symbol: NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted...
Page 18
150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz 0.12 0.12 0.23 0.01 0.38 0.38 0.73 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.