Measurement of Noise/Drift
For measuring noise and drift, refer to the instruction
manual for CDS (EZChrom or Empower).
Noise and drift need to be measured at wavelengths
250 nm.
Specifications
Noise 0.5 10
Conditions: Measuring conditions using standard
Drift
a) 0.4 10
b) 0.4 10
whichever is greater.
Conditions: Measuring conditions using standard
5 - 3
-5
AU or less
flow cell:
poured methanol at 0.5 ml/min,
COARSE slit, bandwidth 4 nm,
smoothed chromatogram 250 10 nm,
reference 360 nm 50 nm, sampling
period 400 ms, response 2 s, ambient
temperature variation within 2 C (20 to
30C), and after 1 hr warm-up.
-3
AU/hr or less
-3
AU/hr/C or less
flow cell:
poured methanol at 0.5 ml/min,
COARSE slit, bandwidth 4 nm,
smoothed chromatogram 250 10 nm,
reference 360 nm 50 nm, sampling
period 1600 ms, response 2 s, and after
1 hr warm-up.