PAR PHYSIO-PORT UP Operator's Manual

Ambulatory blood pressure system

Advertisement

Quick Links

PHYSIO-PORT UP
Ambulatory Blood Pressure System
Firmware Version 3.0
Operator's Manual
A9951 - ENG
Revision C

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PHYSIO-PORT UP and is the answer not in the manual?

Questions and answers

Summary of Contents for PAR PHYSIO-PORT UP

  • Page 1 PHYSIO-PORT UP Ambulatory Blood Pressure System Firmware Version 3.0 Operator's Manual A9951 - ENG Revision C...
  • Page 2 Note The information in this manual only applies to PHYSIO-PORT UP, firmware version 3.0. It does not apply to earlier firmware versions. Due to continuing product innovation, specifications in this manual are subject to change without notice © 2023 PAR Medizintechnik GmbH & Co. KG. All rights reserved.
  • Page 3: Table Of Contents

    A9951 - ENG Revision B 2022-03-28 Changing of the manufacturer Address Updated for MDR requirements of the PHYSIO-PORT A9951 - ENG Revision C 2023-02-09 UP device. Update of the order information in Chapter 10, Change of the battery charger A9951 Revision C PHYSIO-PORT UP...
  • Page 4 Revision History General Information  performance and correct operation and ensures The product PHYSIO-PORT UP bears the CE patient and operator safety. Please note that marking CE (notified body MEDCERT 0482 information pertinent to several chapters is GmbH) indicating its compliance with the given only once.
  • Page 5 General Information PAR Medizintechnik GmbH & Co. KG Rigistr. 11 12277 Berlin Germany Tel. +49 30 235 07 00 The country of manufacture is indicated on the device label. A9951 Revision C PHYSIO-PORT UP...
  • Page 6: Application, Safety Information

    The parts of the equipment described in this manual, including all accessories that come in The PHYSIO-PORT UP device is intended to be used contact with the patient during the intended use, in combination with a suitable blood pressure cuff for...
  • Page 7 (Each 10 cm inflation measurement method, which may be due to difference result in a pressure deviation of 8.0 mmHg.) motion artifacts, for example, PHYSIO-PORT UP will automatically switch to the deflation measurement When the patient is sitting, lying, or standing during method and complete the blood pressure measurement.
  • Page 8 PORT UP device to another port of the PC can lead to interference during data transfer) disconnection of items update or upgrade of equipment  PHYSIO-PORT UP may be connected to a PC with PhysioPortWin software. While connected to any of these devices, PHYSIO- PORT UP must be disconnected from the patient.
  • Page 9 Caution Risk to Persons — Equipment damage, risk to persons—  Before cleaning, PHYSIO-PORT UP must be  Before connecting the battery charger to the disconnected from other equipment (f. e. PC). power line, check that the voltage ratings on ...
  • Page 10 : push to start and stop a measurement and to confirm entries (Rechargeable) batteries Lid covering battery compartment Fig. 2-1 Controls and indicators of PHYSIO-PORT UP Nameplate (example) Port for connection to PC (RS232) Port for connection to PC (USB)
  • Page 11 Further relevant symbols used on the battery charger Temperature limits Protection class II equipment Humidity limits For indoor use only Air pressure limits UL marking USB port, connection to PC Serial port, connection to PC A9951 Revision C PHYSIO-PORT UP...
  • Page 12: Controls And Indicators

     Use only the original charger to recharge the when the cuff is closed. NiMH batteries.  Do not attempt to recharge alkaline batteries.  If PHYSIO-PORT UP will not be used for one month or more, remove the (rechargeable) batteries Latex-free blood pressure cuff. from the device.
  • Page 13  Open the battery compartment on the back of the BP monitor is put into service for the first PHYSIO-PORT UP as shown in Fig. 3-1. time or when you change from NiMH to alkaline batteries and vice versa. Charging NiMH Batteries Caution Equipment damage, patient hazard —...
  • Page 14 To switch OFF: Remove the batteries or press for 3 seconds. Performance Check When turned on, PHYSIO-PORT UP runs a self-test that Fig. 3-2 Charging unit and power supply and USB includes all symbols and segments on the LCD (Fig. 3-4). cable Then it checks the batteries and indicates the remaining capacity.
  • Page 15 Setup Before using PHYSIO-PORT UP on a patient Selecting the Measurement Method  clear the memory Briefly switch PHYSIO-PORT UP off and on again and wait for the time to be displayed. check date and time and adjust if required ...
  • Page 16  Selecting a Measurement Protocol If the indicated year is correct, confirm it with  Briefly switch PHYSIO-PORT UP off and on again or correct it with , then confirm with  and wait for the time to be displayed.
  • Page 17: Application

     The pressurization of the cuff can temporarily cause loss of function of simultaneously used monitoring equipment on the same limb.  By watching the limb, it is necessary to check that operation of the PHYSIO-PORT UP does not result prolonged impairment patient blood circulation.
  • Page 18  the cuff and the PHYSIO-PORT UP are used measurement, tighten the cuff a little and push inside the ambient conditions for operation and again (see also chapter „Error Codes). inside the measuring range (s. chapter „Technical If the trial measurement has been completed Specifications“...
  • Page 19 (see page 36).  to place the PHYSIO-PORT UP with the wearable pouch on the night stand while in bed,  how to switch the device manually from the day to Absolute Contraindications: the night phase (see chapter „Toggling Between...
  • Page 20 In the three measurement protocols, the day phase lasts BP Measurement from 7 a.m. to 10 p.m. and the night phase from 10 These are the buttons on PHYSIO-PORT UP used p.m. to 7 a.m. On the display, the two phases are during an ambulatory blood pressure measurement:...
  • Page 21: Data Output

    2001589-011 if the serial port of PHYSIO-PORT UP is used (a, Fig. 5-1)  Turn on PHYSIO-PORT UP and wait for the time to be displayed. For more information about data output, please refer to the Operator’s Manual of PhysioPortWin.
  • Page 22: Error Codes

    At the same time the device 40 mmHg and no diastolic pressure could be switches to a deflation rate of 4 mmHg/s. identified (PHYSIO-PORT UP does not When it detects more than 13 oscillations later measure diastolic pressures below 40 mmHg).
  • Page 23: Software Installation

    4. Click Finish to complete the first part of the USB installation is then started as follows: driver installation procedure.  Open the Windows Explorer. 5. Turn on PHYSIO-PORT UP and connect it to the  Select the USB stick drive. PC, using the USB connection cable. Windows will ...
  • Page 24 The selected port must be also be set in PhysioPortWin. 6. Select OK where needed and/or close all windows to save the settings. Disconnect the USB cable and restart Windows. A9951 Revision C PHYSIO-PORT UP...
  • Page 25: Cleaning, Maintenance, Disposal

    Shock Hazard — cleaning agent that contains a disinfectant (do not machine-wash). Ensure that no liquid penetrates Disconnect PHYSIO-PORT UP from the PC or into the cuff bladder or the pressure tubing. printer before cleaning. The cuffs can be disinfected with isopropyl ...
  • Page 26 For safety, the device requires regular maintenance. To that must be between 25 and 100. If the displayed ensure functional and operational safety of PHYSIO- value is outside this range, PHYSIO-PORT UP PORT UP, Technical Safety Inspections should be must be returned for repair.
  • Page 27: Technical Specifications

    - USB (1.1 or 2.0) - IP20: PHYSIO-PORT UP - RS-232 (9.600 Bd / 8N1) - IP02: wearable pouch of the PHYSIO-PORT UP Battery - IP22: PHYSIO-PORT UP in wearable pouch - 2 AA size rechargeable NiMH batteries, 1.2 V, Expected Service Life >...
  • Page 28: 10 Order Information

    Order Information 10 Order Information EMC Compliant Cables and Accessories PHYSIO-PORT UP PC System with A20931 Battery charger with USB connection Warning cable, and user manual The use of accessories, transducers and cables A20041 4 NiMH batteries other than those specified may result in increased...
  • Page 29: Appendix - Electromagnetic Compatibility (Emc)

    Guidance and Manufacturer’s Declaration—Electromagnetic Emissions PHYSIO-PORT UP is intended for use in the electromagnetic environment specified below. It is the responsibility of the customer or user to ensure that PHYSIO-PORT UP is used in such an environment. Emissions Test Compliance...
  • Page 30 Guidance and Manufacturer’s Declaration—Electromagnetic Immunity PHYSIO-PORT UP is intended for use in the electromagnetic environment specified below. It is the responsibility of the customer or user to ensure that PHYSIO-PORT UP is used in such an environment. EN/IEC 60601 Electromagnetic...
  • Page 31 RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which PHYSIO-PORT UP is used exceeds the applicable RF compliance level above, PHYSIO-PORT UP should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the equipment.
  • Page 32 Recommended separation distances between portable and mobile RF communications equipment and PHYSIO-PORT UP PHYSIO-PORT UP is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of PHYSIO-PORT UP can help prevent electromagnetic interference...
  • Page 33 Place PHYSIO-PORT UP with the wearable pouch on your nightstand while you are sleeping. You are allowed to change the day phase and the night phase manually, if you go to bed before 10 pm or get up before 7 am. To change the phases press the button once.
  • Page 34 Functional Description 8 Switching off 15 Switching on 15 Symbols used on the battery charger 12 Symbols used on the display 12 General Information 5 Symbols used on the equipment 12 Symbols used on the packaging 12 A9951 Revision C PHYSIO-PORT UP...
  • Page 35 Technical Inspections of the Measuring System 27 USB Driver installation 24 Technical Safety Inspections 27 Time set 17 Toggling Between Day and Night Phase 21 Warning 5 Trial Measurement 19 Weight 28 A9951 Revision C PHYSIO-PORT UP...
  • Page 36 Index PAR Medizintechnik GmbH & Co. KG Rigistr. 11 12277 Berlin Germany Tel: +49 30 2350700 www.par-berlin.com Printed in Germany A9951 Revision C PHYSIO-PORT UP...