Download Print this page

Advertisement

Quick Links

TORQUE TESTER
MINI K
OPERATOR'S HANDBOOK
KOLVER S.r.l - Via Corner, 19/21 - 36016 THIENE - ITALY - tel. +39 0445 371068 fax +39 0445 371069
kolver@kolver.it - www.kolver.it
Vers. 140323 rev

Advertisement

loading
Need help?

Need help?

Do you have a question about the MINI K and is the answer not in the manual?

Questions and answers

Summary of Contents for Kolver MINI K

  • Page 1 TORQUE TESTER MINI K OPERATOR’S HANDBOOK KOLVER S.r.l - Via Corner, 19/21 - 36016 THIENE - ITALY - tel. +39 0445 371068 fax +39 0445 371069 kolver@kolver.it - www.kolver.it Vers. 140323 rev...
  • Page 2 MINI KE/100/S 021405/100/S ± 20 cNm MINI KE + KTE100 ▪ Built-in transducer to measure on joint simulator (MINI K.../S) ▪ External transducer to measure on joint simulator (MINI KEF…) ▪ External rotary transducer to measure the real torque being applied from the tool (MINI KE/…/S) ▪...
  • Page 3 3. DESCRIPTION 1. Mounting holes 2. Display 4 digit / 8 lines 3. “ON/OFF” key : press for 3 seconds to switch tester on or off 4. “CLEAR” key : press to reset the displayed value 5. “UNIT” key : press to select the unit of torque measurements 6.
  • Page 4 The Joint Simulator (JS) consists of a screw compressing a series of washers enclosed in a case with hex female head for proper fit to any Kolver torque tester. The head of the joint simulator can be supplied with different attachment depending on the model used which allows the screwdriver to be connected to the joint simulator.
  • Page 5: Selecting The Unit

    NB. We recommend to grease the JS each 1000 cycles. 6. STARTING AND OPERATING THE TESTER 1. Immobilize the tester when checking torque values over 1 Nm. This is critical for the safety the operator as well as for the accuracy of torque measurements during operation. 2.
  • Page 6 Push at the same time CLEAR+UNIT for 5 sec. “Prg” Push “UNIT” per scorrere la lista. After 3 sec. see -TRACK: Track Mode active. “SEnS” -OFF : Automatic reset off. : Automatic reset on . Push UNIT Push “ON/OFF” to save or “CLEAR” to Push ON/OFF exit “rST”...
  • Page 7 Push UNIT MANUAL Mode: Push on “UNIT” and “CLEAR” to modify Push ON/OFF ON/OFF Push UNIT the FATC. Remove the weight and push “ON/OFF” to save. “DATE” Push ON/OFF Set date and hour: Push UNIT Push “UNIT” to set hour.min (24hours, day.month.year.
  • Page 8 Example of MINI KE + KT series external cell + joint simulator Vers. 140323 rev...
  • Page 9 Example of MINI KE + KTE series external cell + joint simulator Vers. 140323 rev...
  • Page 10 KT SERIES TRANSDUCERS Code Model Torque range Attachment 023001 0,05-1 Female hex 13mm 023005 0,3-5 Female hex 13mm 023020 KT20 0,5-20 Female hex 13mm Female Sq 3/8” 023050 KT50 2-50 Female Sq 1/2” 023100 KT100 5-100 KT1 – KT5 – KT20 Vers.
  • Page 11 KT50 Vers. 140323 rev...
  • Page 12 KT100 Vers. 140323 rev...
  • Page 13 Vers. 140323 rev...
  • Page 14 ROTARY TRANSDUCERS Code Model Torque range Attachment Maschio-femmina esagonale 1 / 4 “ 022405 KTE5 0,5-5 Maschio-femmina esagonale 1 / 4 “ 022425 KTE25 2-25 Maschio-femmina quadro 3 / 8 “ 022450 KTE50 5-50 Maschio-femmina quadro 1 / 2 “ 022411 KTE100 10-100...
  • Page 15 KTE50 MODEL KTE10 MODEL Vers. 140323 rev...
  • Page 16: Maintenance

    7. No one, whether an agent, servant or employee of KOLVER, is authorized to add to or modify the terms of this limited guarantee in any way. However it’s possible to extend the warranty with an extra cost.
  • Page 17: Exploded Views And Part Lists

    EXPLODED VIEWS AND PART LISTS Minik1/S DESCRIPTION CODE Plastic support (4 pcs) 800016 Internal transducer 1Nm (miniK1) 240505 Washer M4 (7 pcs) 241015 Screw M4 x 20 241014 Flat washer M3 (4 pcs) 800042 Nut 6,3 mm (4 pcs) 241003 Metal housing miniK../S 240001/BCU Membrane miniK...
  • Page 18 minik5/S – minik20/S DESCRIPTION CODE Plastic support (4 pcs) 800016 Internal transducer 5Nm (miniK5/s) 240503 Internal transducer 20Nm (miniK20/s) 240504 Washer M3 (5 pcs) 800041 Flat washer M3 (4 pcs) 800042 Nut 6,3 mm (4 pcs) 241003 Metal housing miniK../S 240001/BCU Membrane miniK 241008...
  • Page 19 Case 241000 Power supply 12V 241009/N Minike/xx/S DESCRIPTION CODE Plastic support (4 pcs) 800016 Washer M3 (3 pcs) 800041 Flat washer M3 (4 pcs) 800042 Nut 6,3 mm (4 pcs) 241003 Metal housing miniK../S 240001/BCU Membrane miniK 241008 Connector M 5 pin 231666 Battery 9V not rechargeable 241010...
  • Page 20 Screw M3 x 22 241012 Screw M4 x 8 (3 pcs) 241011 Case 241000 Power supply 12V 241009/N M6 (code 240600) Pos. Description Code Seiger 240601 Joint shaft 240602 Washer spring (8) 240603 Joint housing 240604 Screw M3x5 872443/ZN Vers. 140323 rev...
  • Page 21 JOINT SIMULATORS Threaded attachment (only on request) Code Model Upper attachment Lower attachment Torque tester 240620 Hex13/M1,6 Female thread M1,6 MiniK1 240621 Hex13/M2 Female thread M2 Hex Female 13mm MINIKEF1 240622 Hex13/M3 Female thread M3 Vers. 140323 rev...
  • Page 22 M4 joint simulator with bearings and washers Vers. 140323 rev...
  • Page 23 Code Model Upper attachment Lower attachment Torque tester MINIK1 e K1 (on request) Hex13 – 1/4“ M4 240640 Male 1/4“ Hex Female 13mm MINI KEF1 (standard) M6 and M8 joint simulator with washer Vers. 140323 rev...
  • Page 24 Code Model Upper attachment Lower Standard equipment attachment MiniK5 Hex13 – 1/4“ Female hex. Male 1/4 “ 240600 MINIKEF5 13mm MiniK20 Hex13 – 1/4“ Female hex. Male 1/4 “ 240800 MINIKEF20 13mm Vers. 140323 rev...
  • Page 25 M6 joint simulator Cod. 240600 M8 joint simulator Cod. 240800 - Standard S-E version Vers. 140323 rev...
  • Page 26 Simulatore di giunto M12 con cuscinetti e molle a tazza Code Model Upper attachment Lower attachment Standard equipment MINIKEF50 3/8” M12 Female 3/8” Male 3/8” 240901 Vers. 140323 rev...
  • Page 27 MINIKEF100 240902 1/2” M12 Female 1/2” Male 1/2" Vers. 140323 rev...
  • Page 28 M12 3/8” Joint simulator Cod 240901 M12 1/2” Joint simulator Cod 240902 Vers. 140323 rev...
  • Page 29: Declaration Of Conformity

    DECLARATION OF CONFORMITY KOLVER S.r.l. VIA MARCO CORNER, 19/21 36016 THIENE (VI) ITALIA Declare that the new tool here described: Torque tester: MINI K1/S 021402/S MINI K5/S 021403/S MINI K20/S 021404/S MINI KEF1 021405/F1 MINI KEF5 021405/F5 MINI KEF20 021405/F20...
  • Page 30: Installation

    2023 KOLVER TORQUE ANALYZER Torque Analyzer software allows for communication between our mini k series torque testers and your PC. Torque Analyzer software allows the capture of torque measurements directly from the torque tester including track mode, graphic display readings, and the real-time calculation of the machine capacity represented by Cm and Cmk dat0061.
  • Page 31 1: Kolver_Torque_Analyzer_ver1_0_1 software main screen The home page is made of three sections: 1- “Control Panel”: to display results, graphics, tables, setting and info of the tester. 2- “Report”: to display values (max, min, cm, cmk, in total, in real time...etc.) 3- “Interface”: connection, modification and export of data.
  • Page 32: Display And Setup

    The mode must be set on the minik/s: on the main screen flag or not the Track option. However, Torque Analyzer has been designed to automatically align the receipt of the first measure by detecting the capture mode set in mini k. Picture 3: To flag Track mode.
  • Page 33 Picture 4: “TABLE” section; 10 values. C. STATISTICS In this section, it is possible to display statistical data based on their detection, classification, synthesis and representation. (See picture 5). Other features: a. PROPRIETIES: it allows to choose how to visualize the curve to plot and any graphics shake if needed.
  • Page 34 D. SETTINGS: Setting section (reports, graphics, date, torque tester info). Picture 6: “SETUP”Section 1- REPORT - Path: path to search the image file to enter in the head of the report. We suggest to upload images not bigger than 240x240 pixels. - Dimension X: Value of adaptation along the X axis of the uploaded image.
  • Page 35 It’s also possible to save the set up of the data tapping Save and keep the same setting for the next sessions. E. TORQUE TESTER: in this section it is allowed to enter all the data of the mini k. Vers. 140323 rev...
  • Page 36 Picture 7 : “TESTER” section. F. BACKUP: Data are automatically saved in a backup file. Available a section of the software to restore the data with filters on day, month, year and delete date. Vers. 140323 rev...
  • Page 37 5. RESULTS: MODIFICATION AND EXPORT Picture 9 : Connection, data capture, export. A- MODIFICATION It’s possible to modify or delete one or more captured data (see picture 7). Removal: in the “TABLE” section of the Control Panel, select one or more rows to be deleted.
  • Page 38 Vers. 140323 rev...
  • Page 39: System Requirements

    -Nominal Torque (Cn): average torque value -Max torque: Cn+Tolerance (Cn)% -Min torque: Cn-Tolerance (Cn)% - Sensor value: torque value given from the mini k. -Current values: values referred to the last one acquired. 7. LANGUAGE Four languages available: English, French, Spanish and Italian.
  • Page 40 Vers. 140323 rev...