Instructions for use

4.1 Introduction

The MediaClave is a device used for the production of culture media containing agar or culture nutrient
fluids. Additionally, the MediaClave can be used as an autoclave or a water bath.
This section of the manual is divided into three main parts:
MEDIACLAVE program operation
AUTOCLAVE program operation
WATERBATH program operation
ATTENTION:
For Limit values consult section 7.5.

4.2 Safety precautions and specifications

4.2.1

MEDIACLAVE and AUTOCLAVE sterilisation

WARNING:
There must be sufficient inner heating medium of distilled or deionized water in the vessel
at all times. The level must be approximately 7 cm below the upper edge of the vessel.
ATTENTION:
Prior to each run, remove the rubber lid seal, wet the seal well and insert it again. The ap-
propriate temperature sensor must be installed for sterilisation (see section 3.2)
ATTENTION:
The safety cover and vessel lid must remain closed throughout the sterilisation cycle. Fail-
ure to close the safety cover and vessel lid will result in an alarmed error message.
ATTENTION:
During MediaClave sterilisation, the cuvette must be installed and filled with product (see
sections 3.2.1). The magnetic stirrer bar must be inserted.
ATTENTION:
stalled in the vessel (see section 3.2.2).
4.2.2

WATERBATH

WARNING:
The inner circulation medium must be distilled or deionized water.
ATTENTION:
The safety cover and vessel lid must remain open throughout a water bath operation. Fail-
ure to open the safety cover and vessel lid will result in an alarmed error message.
INTEGRA Biosciences
During autoclave sterilisation, the cuvette must be removed and the steel base plate in-
Page 25