Download Print this page

Omron NE-C303K-KDUK Instruction Manual

Compressor nebulizer

Advertisement

Quick Links

Compressor Nebulizer
Nami Cat (NE-C303K-KDUK)
Nami Cat (NE-C303L-KDUK)
Nami Cat (NE-C303J-KDUK)
Instruction Manual
IM-NE-C303K-KDUK-01-09/2020
3A4362 rev.00
Issue Date: 2020-10-05
Intended use
Medical Purpose:
This product is intended to be used for inhaling medication for respiratory disorders.
Intended User:
• Legally certified medical experts, such as doctor, nurse and therapist, or
healthcare personnel or patient under the guidance of qualified medical
experts.
• The user should also be capable of understanding the general operation
of Nami Cat nebulizer and the content of this instruction manual.
Intended Patients:
Persons who suffer from respiratory disorders and require inhalation of
medications.
Environment:
This product is intended for use in a medical facility, such as hospital, clinic
and doctor's office, and in general household.
Durable period:
Durable periods are as follows, provided the product is used to nebulize
3 times a day for 10 minutes each time at room temperature (23°C).
Durable period may vary depending on usage environment.
Compressor (Main unit):
5 years
Air Tube - Nebulizer Kit - Mouthpiece -
Child Mask - Adult Mask - Adapter:
1 year
Air Filter:
70 applications
Frequent usage of the product may shorten the durable period.
Know your unit
Compressor (Main unit)
Rear View
Air Tube
Power Switch
Connector
Air Filter Cover
Including air filter
Nebulizer Kit
Holder
Ventilation Slots
Ventilation Slots
Nebulizer Kit
Adapter
Inhalation Top
Mouthpiece
Vaporiser Head
EN
Adult Mask (PVC)
Medication Tank
Child Mask (PVC)
Air Tube Connector
Air Tube
Elastic bands
(PVC, 100 cm)
(2 pcs)
Instruction manual
How the nebulizer kit works
The medication that is pumped up through
the medication channel is mixed with
Aerosol
compressed air which is generated by a
compressor pump.
The compressed air mixed with medication
Medication
is turned into fine particles and sprayed
when in contact with the baffle.
Compressed Air
Important safety instructions
Read the instructions carefully before use.
Warning:
1.
Use the device only as inhalator for therapeutic purposes. Any other
use is improper and may be dangerous. The manufacturer is not liable
for improper uses.
2.
Do not use in anaesthetic or ventilator breathing circuits.
3.
Always disconnect the power plug after use.
4.
Do not to cover the ventilation slots during use. The compressor could
get hot and there is a possible risk of a burn if touched.
5.
The compressor and power plug are not waterproof. Do not spill water
or other liquids on these parts. If liquid does spill on these parts,
immediately unplug the power plug and wipe off the liquid.
6.
Do not attempt to repair the device. See Troubleshooting section.
7.
Keep the device out of reach of unsupervised infants and children. The
device may contain small pieces that can be swallowed.
8.
Provide close supervision when this device is used by, on, or near
children or invalids.
9.
For type, dose, and regime of medication follow the instructions of your
doctor or respiratory therapist.
10. This product should not be used on patients, who are unconscious or
are not breathing spontaneously.
11. After finishing the treatment remember to disconnect the air tube from
the nebulizer kit and from the compressor.
12. In view of their length, the power plug and air tube could constitute a
strangulation hazard.
13. Be sure to use the compressor at a place where the power plug is
easily accessible during treatment.
14. If you experience any allergic reactions or other difficulties during use,
stop using the device immediately and consult your doctor.
15. While using this device, make sure that no mobile phone or any other
electrical devices that emit electromagnetic fields is within 30 cm. This
may result in degradation of performance of the device.
16. Do not use oils, decoctions and tinctures from herbs for inhalation in
the compressor nebulizer.
How to use
1. Make sure that the power switch is in the off (
Power Cord
2. Plug the power plug into a power outlet.
Note: Do not place the device in a location where it is difficult to
disconnect the power cord.
3. Remove the inhalation top from the medication tank.
1) Rotate the inhalation top anti-clockwise.
2) Lift the inhalation top out of the medication tank.
Power Plug
4. Add the correct amount of prescribed medication
into the medication tank.
5. Verify the presence of the vaporiser head inside the
medication tank.
6. Put the inhalation top back onto the medication tank.
1) Lower the inhalation top on the medication tank.
2) Rotate the inhalation top clockwise.
7. Attach the adapter, and the Mask or
mouthpiece to the nebulizer kit tightly.
8. Attach the air tube. While twisting the air
tube plug slightly, push it firmly into the air
tube connector.
9. Hold the nebulizer kit as indicated on the right.
Follow the instructions of your doctor or
respiratory therapist.
Caution:
Do not tilt the nebulizer kit at an angle of greater than 30 degrees in all directions.
Baffle
Medication may flow into the mouth or it may result in ineffective nebulization.
Aerosol
10. Turn the power switch to the on (
) position. As the compressor starts,
nebulization begins and aerosol is generated. Gently inhale the
medication. Exhale through the nebulizer kit.
Medication
11. When treatment is completed, turn the power off and unplug the
compressor from the power outlet.
Cleaning and daily disinfecting
Wash hands thoroughly before cleaning and disinfecting the nebulizer parts.
Always disinfect the device parts before first use, and if the device has not
been used for a long time, and also after the last treatment session of a day.
After each use remove any remaining medication and clean the nebulizer
parts. This will prevent inefficient nebulization and reduce risk of infection.
Please refer to the instructions below
Cleaning the Nebulizer Kit, Masks, Mouthpiece and Adapter:
Fully disassemble the nebulizer kit. Wash the parts in warm water and mild, neutral
detergent. Rinse them thoroughly with clean hot tap water, gently tap to remove
excess water and allow to air dry in a clean place. It is advisable to replace the
Nebulizer Kit after some 100 to 120 treatments or after about 20 boiling cycles.
Cleaning the Compressor:
Firstly, make sure that the power plug is unplugged from the power outlet.
Wipe clean with a soft cloth moistened with water or mild, neutral detergent.
Removing condensation from the Air Tube:
If moisture remains in the air tube, follow the procedure described below. Make sure
that the air tube is connected to the compressor. Disconnect the air tube from the
nebulizer kit. Turn on the compressor and dry the air tube by passing air through it.
Disinfecting the Nebulizer Kit, Masks, Mouthpiece and Adapter:
Always disinfect the parts before using the product for the first time and
after the product has not been used for an extended period of time as
well as after the last treatment of the day. If the parts are heavily stained,
replace them with new ones. The Nebulizer Kit, Masks, Mouthpiece and
Adapter can be disinfected by using a chemical disinfectant such as ethanol,
sodium hypochlorite (Milton), Quaternary ammonium (Osvan), Chlorhexidine
(Hibitane) and Amphoteric Surfactant (Tego), following product instructions
and rinsing off with clean warm water. Leave to air-dry.
Note: Never clean with benzene, thinner or a flammable chemical. The
(non-latex) elastic bands cannot be disinfected. Remove them when
disinfecting the masks.
Boiling:
The Nebulizer Kit, Mouthpiece and Adapter can also be disinfected by boiling
between 15 to 20 minutes in plenty of water. After boiling, carefully remove
the parts, shake off excess water and allow to air-dry in a clean environment.
Note: Do not boil the Mask and the Air Tube.
Replacing the Air Filter:
If the air filter has changed colour, or has been used for
more than 70 applications, replace it with a new one. To
) position.
purchase the filters – see Other Optional/Replacement
Parts paragraph. Remove the air filter cover using a flat
screwdriver as shown in figure; remove the filter and fit
the new filter as shown.
Put the air filter cover back in place.
Note: Do not wash or clean the air filter. If the air filter becomes
wet, replace it. Damp air filters can cause blockages.
Technical data
Product Category:
Aerosoltherapy Equipment
Product Description:
Compressor Nebulizer
Model (code):
Nami Cat (NE-C303K-KDUK/
NE-C303L-KDUK/NE-C303J-KDUK)
Rating:
230V ~50Hz
Power Consumption:
150VA
Operating Mode:
Continuous Use
Operating Temperature/
+5°C to +40°C / 15% to 85%RH /
Humidity/Air Pressure:
700 to 1060 hPa
Storage and Transport
-20°C to +60°C / 5% to 95%RH /
Temperature/Humidity/
500 to 1060 hPa
Air Pressure:
Classifications:
Type BF (Applied part): Mouthpiece and Masks
IP Classification:
IP21 (Ingress Protection)
Weight of Compressor:
Approx. 1.2 kg (compressor only)
Dimensions of Compressor: Approx. 180 (W) x 135 (H) x 148 (D) mm
Contents:
Compressor, Nebulizer Kit, Air Tube (PVC,
Right angle
100 cm), Mouthpiece, Adapter, Adult Mask
(PVC), Child Mask (PVC), Instruction Manual
Appropriate Medication
2 ml minimum - 12 ml maximum
Quantities:
Residual volume of
Approx. 0.7 ml
medication:
Sound: Noise level
Approx. 58 dB
(at 1 m distance)
Particle Size (MMAD):
Nebulization Rate
Approx. 0.35 ml/min (NaCl 0.9%)
(by weight loss):
Aerosol Output
Approx. 0.21 ml
(2 ml, 1% NaF):
Aerosol Output Rate
Approx. 0.06 ml/min
(2 ml, 1% NaF):
MMAD = Mass Median Aerodynamic Diameter
Results of Cascade impactor measurements for Particle size
Cumulative % particle mass of sodium fluoride undersize
100.00
80.00
60.00
40.00
20.00
0.00
0.10
____ Mean
........ Individual
1.00
10.00
100.00

Advertisement

loading

Summary of Contents for Omron NE-C303K-KDUK

  • Page 1 Product Description: Compressor Nebulizer Nebulizer Kit Adapter 6. Put the inhalation top back onto the medication tank. Model (code): Nami Cat (NE-C303K-KDUK/ 1) Lower the inhalation top on the medication tank. NE-C303L-KDUK/NE-C303J-KDUK) Inhalation Top 2) Rotate the inhalation top clockwise. Mouthpiece...
  • Page 2 It is designed to give you the highest level of comfort, provided it is properly operated and maintained as described in the instruction manual. This product is guaranteed by OMRON for a period of 3 years after the date of purchase. The proper construction, workmanship and materials of this product is guaranteed by OMRON.

This manual is also suitable for:

Nami catNe-c303l-kdukNe-c303j-kduk