Remove the meter from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If noticeable damage is evident, contact your local Hanna Instruments Office. Each meter is supplied with:
Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in the original packaging with the supplied accessories.
The probe is shipped dry. Before using the electrode, remove the protective cap. Then follow the calibration procedure.
Range | 0 to 2000 ppm 0.0 to 50.0 ˚C (32.0 to 122.0 ˚F) |
Resolution | 1 ppm 0.1 ˚C (0.1 ˚F) |
Accuracy (@25°C/77°F) | ±2% F.S. ±0.5°C (±1.0°F) |
TDS Conversion factor (CF) | 0.5 |
Calibration Solutions | HI70032 (1382 ppm) |
Calibration | automatic, single point (1382 ppm) |
Temperature Compensation | automatic, 0.0 to 50.0 ºC (32.0 to 122.0 ºF) |
Battery type | CR2032 Li-Ion 3V (1 pc.) |
Battery life | approximately 250 hours of continuous use |
Environment | 0 to 50 ºC (32 to 122 ºF); RH 100% max |
Weight (without battery) | 68 g (2.4 oz.) |
Press the ON/OFF button to turn the meter on. At start-up, all the LCD segments are displayed for 1 second, then the percent indication of the remaining battery life is displayed for another second. The meter then enters the normal measuring mode.
Note: Keeping the ON/OFF button pressed while turning the meter on will display all LCD segments as long as the button is pressed.
Press the CAL button. "CAL" message is displayed.
Remove the battery cover and press the Setup button located on the side of the battery.
While in measurement mode, remove the battery cover. Press the Setup button located on the side of the battery in the battery compartment. The meter will enter in setup mode. Press the ON/OFF button to move through setup parameters. Change option by pressing CAL button. The default settings are: "Set t" measure unit - °C, "AOFF" - 8 min.
To select the temperature unit when "SET t" is displayed press the CAL button to change between °C or °F.
To select the AUTO OFF when "AOFF" is displayed press the CAL button to change between 8 min, 60 min or --- (disabled).
Press the ON/OFF button.
For better accuracy, frequent calibration of the meter is recommended. In addition, the instrument must be recalibrated:
Press the ON/OFF button to turn the meter ON. Place the probe to the desired solution.
The conductivity value, automatically compensated for temperature, will be shown on the LCD.
The measured temperature will be shown on the secondary LCD.
Note: Before taking any measurement make sure the meter has been calibrated.
Place meter in calibration mode. Press ON/OFF and "Clr" is displayed. The meter will now be at default calibration.
During user calibration, if the reading is out of the accepted range, the meter will display "---- Err".i.e. 2000 ppm blinking means the limit of the measured value is out of range.
If the measured temperature is higher than 50.0 ºC or lower than 0.0 ºC, the 50.0 ºC or 0.0 ºC temperature value will blink on the LCD in measure mode.
To obtain the highest accuracy for measurements it is important to follow these tips:
The meter features a low battery indicator. When the battery is running low (under 10%), the battery indicator will blink on the LCD. When the battery is discharged "dEAd bAtt" will be displayed on the LCD for 2 seconds and the meter will turn off.
To change the CR2032 Li-ion battery, turn the battery cover located on the back of the meter counterclockwise to unlock. Remove cover and replace with new battery + side facing up.
Note: Batteries should only be replaced in a safe area using the battery type specified in this instruction manual. Old batteries should be disposed in accordance with local regulations.
Before using Hanna Instruments products, make sure that they are entirely suitable for your specific application and for the environment in which they are used. Operation of these instruments may cause unacceptable interferences to other electronic equipment, thus requiring the operator to take all necessary steps to correct such interferences. Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance. To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument safety do not use or store the instrument in hazardous environments.
Solutions
Code | Description |
HI70032P | 1382 ppm (mg/L) calibration solution, 20 mL sachets (25 pcs.) |
HI7032M | 1382 ppm (mg/L) calibration solution, 230 mL bottle |
HI7032L | 1382 ppm (mg/L) calibration solution, 500 mL bottle |