® Safety information R&S RT‑Z1M 1 Safety information The product documentation helps you use the product safely and efficiently. Follow the instructions provided here and in the following chapters. Intended use The R&S RT‑Z1M 1 MOhm adapter is designed for measurements on circuits that are only indirectly connected to the mains or not connected at all.
Page 6
® Safety information R&S RT‑Z1M ● Do not apply voltages higher than the maximum rated voltage, see the imprint on the adapter. ● When using the adapter in combination with a passive probe that can measure vol- tages higher than the maximum rated input voltage of the adapter (e.g. R&S RT- ZP10): Observe the maximum rated voltage of the probe.
® Product description R&S RT‑Z1M Unpacking the instrument 2 Product description 2.1 Key features and key characteristics The R&S RT‑Z1M 1 MOhm adapter is an accessory for oscilloscopes that have only inputs with 50 Ω input impedance. It allows you to connect measurement equipment requiring a 1 MΩ...
® Product description R&S RT‑Z1M Description of the adapter ● User manual of the R&S RT‑Z1M ● R&S RT-ZP10 10:1 passive probe ● User manual of the R&S RT-ZP10 ● Accessories for R&S RT-ZP10 (delivery contents is listed in the user manual of the probe) ●...
® Product description R&S RT‑Z1M Description of the adapter 1 = Adapter box 2 = R&S Probe interface to connect an active probe to the adapter, and a contact ring to connect a passive probe 3 = R&S Probe interface to connect the adapter to an oscilloscope 4 = Release knob 5 = R&S RT-ZP10 passive probe 2.3.1 Adapter box...
® Product description R&S RT‑Z1M Accessories 2.3.3 Dimensions of adapter and probe The R&S RT‑Z1M 1 MOhm adapter and the R&S RT-ZP10 passive probe have the fol- lowing dimensions (all values in mm): ‑ Z1M adapter with R&S RT-ZP10 Figure 2-3: Dimensions of the R&S RT 2.4 Accessories 2.4.1 Supplied accessories Table 2-2...
® Putting into operation R&S RT‑Z1M Connecting the adapter to the oscilloscope 3 Putting into operation Read and observe the instructions in Chapter 1, "Safety information", on page 5 before using the adapter. For your safety, follow the safety instructions delivered with your Rohde & Schwarz oscilloscope and probe, and the instructions given in the manuals of your measure- ment equipment.
® Putting into operation R&S RT‑Z1M Disconnecting the adapter from the DUT ► To disconnect the adapter: a) Press and hold the release button (3). b) Pull the adapter away from the oscilloscope. 3.2 Identification of adapter and probe When the R&S RT‑Z1M is connected to the oscilloscope, the oscilloscope recognizes the adapter and reads out the specific parameters from the data memory of the adapter.
® Putting into operation R&S RT‑Z1M Using the adapter 3.5 Using the adapter All settings of the R&S RT‑Z1M are adjusted on the oscilloscope. These settings are: ● Offset ● Input coupling ● Lowpass filter (bandwidth) ● Self-alignment The gain or attenuation of the adapter is automatically set by the oscilloscope depend- ing on the vertical scale.
® Putting into operation R&S RT‑Z1M Using the adapter 3.5.3 Analog bandwidth The R&S RT‑Z1M features two lowpass filters of 200 MHz and 20 MHz. The filters reduce bandwidth and thus noise during the measurement. The bandwidth filter is part of the vertical settings of the channel to which the adapter is connected.
® Typical characteristics R&S RT‑Z1M Step response 4 Typical characteristics The R&S RT‑Z1M 1 MOhm adapter has several gain or attenuation modes. The mode of the adapter, and thus the gain or attenuation, depends on the vertical scale and is set automatically by the oscilloscope to achieve optimized system behavior.
Page 16
® Typical characteristics R&S RT‑Z1M Step response ‑ Z1M (all 7 modes) Figure 4-2: Step response of the R&S RT The rise time of a system is inversely proportional to the bandwidth. The following approximation applies: rise The minimum measurable rise time of the whole setup also depends on the bandwidth of the probe connected to the R&S RT‑Z1M, and the bandwidth of the oscilloscope.
® Maintenance and service R&S RT‑Z1M Contacting customer support 5 Maintenance and service Like all Rohde & Schwarz products, Rohde & Schwarz probes and adapters are of high quality and require only minimum service and repair. However, if service or calibration is needed, contact your Rohde &...
® Maintenance and service R&S RT‑Z1M Disposal 5.3 Returning for servicing Use the original packaging to return your R&S RT‑Z1M to your Rohde & Schwarzser- vice center. A list of all service centers is available on: www.services.rohde-schwarz.com If you cannot use the original packaging, consider the following: 1.
Page 19
® Maintenance and service R&S RT‑Z1M Disposal Figure 5-2: Labeling in line with EU directive WEEE Rohde & Schwarz has developed a disposal concept for the eco-friendly disposal or recycling of waste material. As a manufacturer, Rohde & Schwarz completely fulfills its obligation to take back and dispose of electrical and electronic waste.
Page 20
® Functional check R&S RT‑Z1M 6 Functional check The functional check confirms the basic operation of the R&S RT‑Z1M 1 MOhm adapter and the R&S RT-ZP10 passive probe using simple measurement equipment. The functional check is not suitable to verify compliance with the specifications of the R&S RT‑Z1M 1 MOhm adapter, because the test results are influenced by the used oscilloscope.