EM TEST
6.4.8.
Verification on coaxial output
Verification Setup with 1000 load
Procedure of verification
-
Switch on the EFT 500Nx
Disconnect the line power from the EUT supply input at the rear of the EFT 500Nx
-
Set the input impedance of the oscilloscope to 50
-
Select Quick Start mode and set the following parameters under Change for further operation.
-
Parameter values for setting
V
= 250 / 500 / 1000 / 2000 / 4000 / 6500V*
f
= 5kHz
td
= 15ms
tr
= 300ms
cpl
= /
+/-
= pos and neg
T
= Endl.
Perform the verification by the following steps
-
1. 1000 loaded condition (KW1000)
For verification connect the oscilloscope via the 1000:1 high impedance attenuator KW1000 to the coaxial
output of the EFT 500Nx and measure as specified. Set the oscilloscope to the 50 input impedance mode.
2. Measure the rise time tr, pulse width td and the peak voltage.
setting
250V
500V
1000V
2000V
4000V
*6500V
t
10% - 90%
r
t
50% - 50%
d
* Only EFT 500N8
Manual for Operation
required Vs on KW1000
0.24 kV 20%
0.48 kV
20%
0.96 kV
20%
1.9 kV
20%
3.8 kV
20%
6.175 kV
20%
30%
5ns
50ns
-15 + 100ns
V 5.23
description
(test voltage settings)
(frequency )
(burst duration)
(repetition rate, time between 2 bursts)
(to measure at the coaxial output port)
(polarity selection)
(test duration )
tolerances
192V to
384V to
768V to
1152V
1520V to
2280V
3040V to
4560V
4940V to
7410V
3.5ns to 6.5 ns
35ns to 150 ns
EFT 500Nx Series
288V
576V
26 / 36