Verification On Coaxial Output - EM TEST EFT 500N5 Series Manual For Operation

Table of Contents

Advertisement

EM TEST
6.4.8.

Verification on coaxial output

Verification Setup with 1000 load
Procedure of verification
-
Switch on the EFT 500Nx
Disconnect the line power from the EUT supply input at the rear of the EFT 500Nx
-
Set the input impedance of the oscilloscope to 50 
-
Select Quick Start mode and set the following parameters under Change for further operation.
-
Parameter values for setting
V
= 250 / 500 / 1000 / 2000 / 4000 / 6500V*
f
= 5kHz
td
= 15ms
tr
= 300ms
cpl
= /
+/-
= pos and neg
T
= Endl.
Perform the verification by the following steps
-
1. 1000 loaded condition (KW1000)
For verification connect the oscilloscope via the 1000:1 high impedance attenuator KW1000 to the coaxial
output of the EFT 500Nx and measure as specified. Set the oscilloscope to the 50  input impedance mode.
2. Measure the rise time tr, pulse width td and the peak voltage.
setting
250V
500V
1000V
2000V
4000V
*6500V
t
10% - 90%
r
t
50% - 50%
d
* Only EFT 500N8
Manual for Operation
required Vs on KW1000
 0.24 kV 20%
 0.48 kV
20%
 0.96 kV
20%
 1.9 kV
20%
 3.8 kV
20%
 6.175 kV
20%
30%
5ns
50ns
-15 + 100ns
V 5.23
description
(test voltage settings)
(frequency )
(burst duration)
(repetition rate, time between 2 bursts)
(to measure at the coaxial output port)
(polarity selection)
(test duration )
tolerances
 192V to
 384V to
 768V to
1152V
 1520V to
2280V
 3040V to
4560V
 4940V to
7410V
3.5ns to 6.5 ns
35ns to 150 ns
EFT 500Nx Series
288V
576V
26 / 36