58
MeasurementTime
which the sample is taken in
(page 60). This parameter must be greater than 17 and
less than 1,179,630.
When measuring a periodic source such as a monitor,
sample at an integer number of periods. Setting the proper
integration time for a monitor at a specific frequency is
essential to obtaining accurate, repeatable measurements.
Each value set in
microsecond. Use this equation to find the integration time:
Multiply the integration time by the number of cycles
you want to measure (31 cycles are recommended). For
example, with a 75 Hz monitor, use this calculation:
Thus, the integration time is:
413,333 microseconds = .413 seconds
ReadTimeout
data when reading from the calibrator port.
You can still use Version 1 (one) of the structure, which
does not include
(one), the
and you must set
MeasurementTime
period of
one cycle
integration time for one cycle
------------------- - =
6
00010
00
1
6
----- -
10
31
=
413333
75
is the number of milliseconds to wait for
ReadTimeout
ReadTimeout
Version
is the number of microseconds over
DCalibratorMeasure
. If you do use Version 1
value defaults to 2000 (2 seconds),
to 1 (one).
DCalibratorInit
is equal to one (1)