Appendix 3 Login Procedure

When using the login function
App-4
When using the Setting/Measurement server or the Maintenance/Test server via the Ethernet interface (/C7
option), you must log into the recorder from the PC. If you complete the procedure successfully up to login
complete in the following figure, the commands in chapter 4 become functional.
No
Within the number of
E1 421
connections
Disconnect
No
User name
E1 422
Disconnect
Within the number of
1
simultaneous uses?
E1 420
Disconnect
Login complete
Start
End
User entry
E1 403
Response from the SBR-EW (message omitted)
For a description of the response format, see section 5.1.
Code (for a description of codes and messages, see page App-6)
Connections cannot exceed the maximum number of connections (see section 2.1).
††
If you attempt to log in using a wrong password four consecutive times, the communication is dropped (the
number of retries for login is three).
†††
If you attempt to log in causing the number of simultaneous uses at the administrator or user level to be
exceeded (see section 2.1) four consecutive times, the communication is dropped (even if the password is
correct).
Connect
E1 400
Within 2
minutes?
Yes
Entered
1
"quit"?
E1 401
Within 2
minutes?
Password
Wait 5 s
No
Verification
Within 3
match?
retries?
No
†††
E0
Judgment
Transition
Increment retry count
E1 403
E1 404
No
1
††
Within 3
retries?
No
1
††
IMSBR08-E2