Color

To ensure proper testing results:

!
Insert tubes into ColorQ with line
and arrow facing forward.
!
"Sun"= Shade meter from
bright sunlight. Press/Hold
button to turn off.
!
"Er7 "=Tube is misaligned or
scratched, or light chamber
is stained. See Users Guide for
more instructions.
!
Brush and rinse all tubes promptly
after use.
!
Allow tubes to dry before
returning to case.
!
Replace reagent caps securely.
!
If the unit should auto-off during
testing rinse and fill one tube to
Blank and return to testing.
!
Use only LaMotte replacement
reagents.
!
bAt = Battery is low.
NOTE: Wash and rinse tubes immediately after use. Tubes used for the
ultra high range test may become stained. Stain will not wash out. Stain
may fade if stained tubes and caps are placed in direct natural light or
sunlight. Designate tubes for use with the ultra high range test. Tubes can
be marked with a permanent marker. Blank and test must be done in the
same tube to eliminate error caused by stained tubes. Replace heavily
stained tubes.
INTERFERENCES: bromine, chlorine dioxide, iodine, oxidized manganese
and chromium, and ozone.
UCL
Q
For software Version 2.00 or higher
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 • 410-778-6394 www .lamotte.com
WARNING! This set contains chemicals that may be harmful
if misused. Read cautions on individual containers carefully .
Not to be used by children except under adult supervision.
09/17 62063
Order Code 2063

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ColorQ and is the answer not in the manual?

Questions and answers

Summary of Contents for LaMotte ColorQ

  • Page 1 09/17 62063 Color Order Code 2063 For software Version 2.00 or higher To ensure proper testing results: Insert tubes into ColorQ with line and arrow facing forward. ”Sun“= Shade meter from bright sunlight. Press/Hold button to turn off. ”Er7 “=Tube is misaligned or scratched, or light chamber is stained.
  • Page 2 4. Press button to turn meter on. When “bLA” appears press button to “Blank” the meter, and go to “FCL” Free Chlorine. To go to “ ” Ultra High Ra nge Total Chlorine press button FOUR times. Remove tube. * Reagent is a potential health hazard. Read SDS at www.lamotte.com.
  • Page 3 1. Add 1 DPD 1 Tablet to same tube. 2. Cap. Shake for TWENTY seconds. Invert slowly 5 times. Tap tube on flat surface 3 times to remove air bubbles. Insert tube into ColorQ. 3. Press button to read “FCL” Free Chlorine.
  • Page 4 7. Allow tube to sit for TWO minutes. Tap tube on surface 3 times to remove air bubbles. 8. Invert 1 time to mix. Insert tube into ColorQ. 9. Press button to read “uCL” Ultra High Range Total Chlorine. 10. Press button to return to blank meter or hold button until OFF is displayed.