Channel Configurations; Compatible Sensors; Accuracy; Thermocouple Measurement Accuracy - Omega OMB-AI-EXP32 User Manual

Analog input expansion module for omb-daq-2416 series
Table of Contents

Advertisement

OMB-AI-EXP32 User's Guide
Parameter
Open thermocouple detect
CJC sensor accuracy

Channel configurations

Channel
Category
Thermocouple
CxH/CxL
The OMB-AI-EXP32 GND and DGND pins are isolated from earth
ground. You can connect thermocouple sensors to voltages
referenced to earth ground as long as isolation between the
GND/DGND pins and earth ground is maintained.
Unused input channels can be left floating or tied to GND (pins 17,
18, 49, 50).
CxH/CxL
Voltage
When connecting differential voltage inputs to a floating voltage
source, the user must provide a DC return path from each voltage
input to ground. To do this, connect a resistor from each input to a
GND pin (pins 17, 18, 49, 50). A value of approximately 100 kΩ
can be used for most applications.
CxH/CxL
Voltage

Compatible sensors

Parameter
Conditions
Thermocouple
J: -210 °C to 1200 °C
K: -270 °C to 1372 °C
R: -50 °C to 1768 °C
S: -50 °C to 1768 °C
T: -270 °C to 400 °C
N: -270 °C to 1300 °C
E: -270 °C to 1000 °C
B: 0 °C to 1820 °C

Accuracy

Thermocouple measurement accuracy

The OMB-AI-EXP32 is a multiplexer-based channel expansion module that increases the number of
thermocouple input channels of a OMB-DAQ-2416 series host module. For thermocouple accuracy
information, refer to the electrical specification of the particular OMB-DAQ-2416 host module that is
interfacing with the OMB-AI-EXP32.
There is a CJC sensor for each terminal block of the module. The thermocouple measurement accuracy assumes
that the screw terminals are at the same temperature as the CJC sensor.
Connect thermocouples to the OMB-AI-EXP32 such that they are floating with respect to GND (pins 17, 18,
49, 50).
When configuring thermocouple sensors, keep any stray capacitance relative to GND (pins 17, 18, 49, 50) as
small as possible to avoid settling time and accuracy errors.
Conditions
15 °C to 35 °C
0 °C to 55 °C
Table 3. Channel configurations
Table 4. Compatible sensor type specifications
12
Specification
Automatically enabled when the channel is configured for a
thermocouple sensor.
±0.15 °C typical
±0.5 °C maximum
Specifications
Specification
16 differential channels
16 differential channels