Check connection between controller and pump, then press STOP
push-button twice to start the pump.
If the P1 input connector is not in position with the link or the
external interlock connections are open, when the START
pushbutton is pressed the display will be as shown in the following
figure.
Disconnect connector P1 and check the link or the external interlock,
then install the connector to start the pump.
If within each step of the soft start mode the rotational speed of the
pump do not reach the planned value within 15 minutes
if in normal operation (after the starting phase) the current drawn by
the pump is higher than programmed (1.8 A), the pump and the inter-
connected devices are switched off and the display will be as shown
in the following figure.
T
Verify that systems has no leaks than press STOP RESET pushbutton
twice to start the pump.
Check that pump rotor is free to rotate then press the STOP RESET
pushbutton twice to start the pump.
If the upper bearing/pump temperature exceeds 60 °C, the pump is
shut off, and the display will be as shown in the following figure.
Turbo-V Remote Panel User Manual / 87-900-987-01