Front Panel Features - HP E840 A Series User And Service Manual

Table of Contents

Advertisement

Front Panel Features

Section
Indicator / Switch

On Indicator
Power
Standby Indicator
ô
Voltages Indicator
Supply
Temp Indicator
í
Fan Switch
Fans
Fans Indicator
÷
Activity Indicator
Backplane
SYSFAIL Indicator
Reset Switch
û
Diagnostic
Connector
Figure 1-4. HP E8401A and E8403A Basic Monitoring Front Panel
Chapter 1
Green - the mainframe is turned on.
Amber - the mainframe is in standby mode; no power is applied to the VXI modules.
Green - all voltages on the VXI backplane are within ±8% of the VXI specification.
Flashing Amber - one or more voltages is out of specification. The backplane voltages
and allowed variations are listed below:
Voltage
Allowed Variation
+5V
+4.875V to +5.125V
+12V
+11.64V to +12.6V
-12V
-12.6V
+24V
+23.28V to +25.2V
-24V
-25.2V
-5.2V
-5.46V
-2V
-2.1V
Green - the power supply temperature is acceptable.
Flashing Amber - the power supply is approaching thermal shutdown. The power supply
temperature can be measured on pin 11 of the Diagnostic connector.
Full - the power supply and mainframe cooling fans are operating at full speed.
Var(iable) - the power supply and mainframe cooling fans are providing adequate cooling
(default setting). Fan speed is a function of the power supply temperature and the
reference temperature (Diagnostic connector pin 12). The reference temperature is a
function of ambient temperature and load.
Green - the power supply and mainframe cooling fans are operating.
Flashing Amber - the power supply fan or the mainframe cooling fan is not operating. Fan
operation can be checked using pin 13 of the Diagnostic connector.
Green - there is communication between instruments across the backplane. Off - there is
no communication between instruments on the backplane.
Flashing Amber - one or more instruments has asserted its SYSFAIL line due to a
power-on initialization failure, self-test failure, or hardware failure. SYSFAIL is asserted
momentarily at power on and during a system reset.
Pressing the reset button asserts the SYSRESET* line on the VXI backplane. When
asserted (low), this line resets the VXI system; all VXI instruments are rest to their
power-on state.
Refer to "Diagnostic Connector" on page 16.
Description
Diagnostic Pin
1
14
to -11.64V
2
15
to -23.28V
3
to -5.044V
16
to -1.9V
4
Getting Started 15