Anyload ES-HA Series Product Manual

Anyload ES-HA Series Product Manual

Precision balance

Advertisement

Quick Links

ES-HA
Precision Balance
Product Manual
Anyload Transducer Co. Ltd
Website: www.anyload.com
Email: info@anyload.com
Fax: +1 866 612 9088
North America Toll Free: 1-855-ANYLOAD (269 5623)
Series
(V1611)

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ES-HA Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Anyload ES-HA Series

  • Page 1 ES-HA Series Precision Balance Product Manual (V1611) Anyload Transducer Co. Ltd Website: www.anyload.com Email: info@anyload.com Fax: +1 866 612 9088 North America Toll Free: 1-855-ANYLOAD (269 5623)
  • Page 2: Table Of Contents

    5.9 RS232C Connection 5.9.1 Baud Rate 5.9.2 Data Form Output 5.9.3 Wiring Method 5.9.4 Demonstrating communication software 5.9.5 Order Input 6. Care and Maintenance 7. Error Codes 8. Troubleshooting 9. Guaranty 10. Specifications ANYLOAD ES-HA Series Precision Balance Product Manual (V1611)
  • Page 3: Introduction

    Introduction Thank you for choosing the ES-HA Series Precision Balance Scale. The ES-HA series precision balances do weighing and measuring with high precision load cell, microcomputer and big liquid crystal display screen. It has the advantage of simple operation, speedy weighing, accuracy, stability, and so on. It is widely used in factories, mines, institutions, agricultures, traffic, building, water conservancy, medicine, food, and so on.
  • Page 4: Construction And Installation

    Packing List (1) Electronic balance 1 pc. (2) AC adapter 1 pc. (3) Operation Instruction 1 copy (4) Balance (Its division is 0.01g or higher) with one weight and a wind cover. ANYLOAD ES-HA Series Precision Balance Product Manual (V1611)
  • Page 5: Operation

    After that, the balance will display “0". In a few seconds, the balance will display the calibrating weight value. At this time, put the corresponding calibrating weight on the center of the platform according to the shown weight value. ANYLOAD ES-HA Series Precision Balance Product Manual (V1611)
  • Page 6: Linear Calibration

    NOTE: If you want to restart parts counting, repeat the operation steps above. If “PC Err” is displayed, it means the sample is too light to provide accurate results within the error range set by the balance. ANYLOAD ES-HA Series Precision Balance Product Manual (V1611)
  • Page 7: Unit Selection

    9.9.2 Data form output by ES series electronic balances + (-) XXX. XX UNIT< CR> All the segments above belong to ASCII. Of these, the unit can be set by ES-HA series electronic balances. 5.9.3 Wiring method ANYLOAD ES-HA Series Precision Balance Product Manual (V1611)
  • Page 8: Demonstrating Communication Software

    The balance must be sent to the manufacturer to be serviced. Err 1 The sample being weighed exceeds the capacity of the balance. If the error occurs when the sample is within the balance capacity, maybe the balance is not correctly calibrated. ANYLOAD ES-HA Series Precision Balance Product Manual (V1611)
  • Page 9: Troubleshooting

    10-35 C temperature rang Power source AC adapter(supplied with balance); or 6 AA Batteries (not included) Pan size (mm) Housing 251 (L) ×172 (W) × 58 (H) dimension (mm) Net weight 1.05kg ANYLOAD ES-HA Series Precision Balance Product Manual (V1611)

Table of Contents