Setting Up
5. Setting Up
5.1 Setting up and initial start-up
Before you set up and connect the device, check that you have received
all components belonging to the product free of mechanical damage.
Device Biosen C_line
Sample tray (Clinic) respectively sample receptacle (GP+)
Bottles for system solution and waste
Power cable
When changing from a cold into a warm environment condensation can
form inside the device. Wait for about one hour before you connect the
device to the mains.
Select a suitable place for setting up the device and make sure to:
Ambient temperature:
humidity:
avoid direct influence of damp
avoid direct sunlight
avoid strong electromagnetic fields
avoid direct influence from ionizing radiation
avoid rapid temperature variations (keep away from heaters,
open windows, ventilators, extraction or air conditioning
systems, etc.)
operate the device in dry rooms on a flat, waterproof surface
free space on the ground required across the whole of the
installation area for ventilation
100-240 Volt and a frequency of 50 or 60 Hz. Set the mains frequency in
the device menu (Section 9.8, page 37) to the correct value of your
region. For further enquiries contact your technical consultant or dealer.
12
Allow the device to reach room temperature!
The device is designed for a mains voltage of
User Manual Biosen C_line Clinic / GP+
15 - 35°C
20 - 85%