Calibration Guide
Technology at Work®
August 2020
© 2020 Janam Technologies LLC. All rights reserved.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GT2 and is the answer not in the manual?

Questions and answers

Summary of Contents for Janam GT2

  • Page 1 Calibration Guide Technology at Work® August 2020 © 2020 Janam Technologies LLC. All rights reserved.
  • Page 2 Total = 986 (divide by 10) Average Value = 98.6 Step 2 Using GT2’s temperature reading functionality, take ten temperature measurements of the same single test subject used in Step 1. Record the average value of all ten measurements. Example:...
  • Page 3 Step 4 In the temperature measurement calibration field, enter the difference between the FDA-approved thermometer (Step 1) and GT2’s temperature reading functionality (Step 2). For the example used, the value that should be entered is “0.1”. •...