CONTENTS Introduction and Intended Use ......................2 Important Information on Blood Pressure and its Measurement ............3 Components of your blood pressure monitor ..................5 Using your Monitor for the First Time ....................6 Measurement Procedure ........................8 Care and Maintenance ........................15 Warranty/Service..........................
1 Introduction and Intended Use It enables reliable measurement of systolic and diastolic blood pressure as well as pulse through the oscillometric method. Before using, please read this instruction manual carefully and then keep it in a safe place. 1.1 Remember… Only a health-care professional is qualified to interpret blood pressure measurements.
Warning: Don’t place the cuff over wound part. Warning: Pressurization of the CUFF can temporarily cause loss of function of simultaneously used monitoring ME EQUIPMENT on the same limb. Caution: To avoid any possibility of accidental strangulation, keep this unit away from children and do not drape tubing around your neck.
Page 5
Picture-01 There are six grids in the display of device. Please refer to the picture-01-01. Different grids represent different interval scales of WHO. Blood pressure value WHO grids in device WHO Classification DIA<80 & SYS<120 Optimal blood pressure DIA<85 & SYS<130 Normal blood pressure DIA<90 &...
17. Systolic blood pressure 18. Month symbol 19. Hour symbol 20. Diastolic blood pressure 3.3 Features of C02 / FB160 1. Cuff self-checking function 2. Double users: 2 x 120 sets memory 3. Average value function 4. Irregular heartbeat checking 5.
Page 8
2. Install 3 “AA” size batteries and make sure the + (positive) and - (negative) polarities match the polarities of the battery compartment, then close the battery cover. Make sure that the battery cover is securely in position. Picture-04 Battery replacement Low Battery Indicator 1.
days later. Take the battery out won’t lead to a record missing. 5 Measurement Procedure 5.1 Before measurement: Avoid eating and smoking as well as all forms of exertion directly before measurement. These factors influence the measurement result. Find time to relax by sitting in an armchair in a quiet atmosphere for about ten minutes before taking a measurement.
Page 10
Picture-05 5.3 Measure Procedure Refer to picture-06 1. Sit comfortably in a chair with your feet flat on the floor. 2. Select your User ID (A or B). 3. Stretch your arm forward on the desk and keep relaxing, make sure the palm of hand is upturned. Make sure arm is in correct position, to avoid body movement.
Page 12
Irregular Heartbeat Detector This symbol - indicates that certain pulse irregularities were detected during the measurement. In this case, the result may deviate from your normal basal blood pressure – repeat the measurement. Information for the physician on frequent appearance of the Irregular Heartbeat Symbol. This instrument is an oscillometric blood pressure monitor device that also analyzes pulse frequency during measurement.
Page 14
SYMBOL DESCRIPTIONS The following symbols may appear in this manual, on the Digital Blood Pressure Monitor C02 / FB160, or on its accessories. Some of the symbols represent standards and compliances associated with the Digital Blood Pressure Monitor C02 / FB160 and its use.
Page 15
DISPOSAL: Do not dispose this product as unsorted municipal waste. Collection of such waste separately for special treatment is necessary. Follow instructions for use Put up Fragile Afraid of the rain Fear of the sun Class ll equipment Handle gently Temperature range No Sterilize requirement Not category AP / APG equipment...
the Start/Stop button can be pressed at any time. The device then immediately lowers the cuff pressure automatically. 5.7 Using the AC Adapter You may also operate this monitor using a CE approved AC adapter (output 5V DC1A with Type C connector).
Device fulfills the stipulations of the International standard IEC60601-1-2 The declaration of conformity is available on: http://DOC.hesdo.com/FB160-DOC.pdf 9 Technical Specifications Model: C02 / FB160 Weight: 326.4g(Batteries are included) Display: 62.5 * 63.5 mm LED Digital Display Size: 123 (W) x 82 (L) x 25 (H) mm Accessories: 1×Main Device, 1×Cuff, 1×Users manual, 1x case, 3x AA batteries...
Page 18
adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used Guidance and manufacture’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of device should assure that it is used in such an environment.
Page 19
Guidance and manufacture’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment. IEC 60601 test Electromagnetic Immunity test Compliance level...
Page 20
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
Page 21
0.01 0.12 0.12 0.23 0.38 0.38 0.73 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Page 22
LTE Band 7 5240 5100- WLAN 802.11 Pulse Modulation 5500 5800 217 Hz 5785 NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
Page 23
Shenzhen Jamr Technology Co., Ltd. 2nd Floor, A-building, No.2 Guiyuan Road, Guihua community, Guanlan town, Longhua new district, 518100 Shenzhen, PEOPLE‟S REPUBLIC OF CHINA Shanghai International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537 Hamburg, Germany. shholding@hotmail.com Imported by: Commaxx BV Wiebachstraat 37 6466 NG Kerkrade The Netherlands...
Need help?
Do you have a question about the C02 and is the answer not in the manual?
Questions and answers