1. Accuracy parameter inputing method and adjusting method 1.1 About the accuracy parameter: the machine parameter is comforted before leave factory, we sticked a label inside the machine cabinet,the position is showed as the following sketch picture 2-1 1.2 Accuracy parameter inputing method: Please follow the manual instruction, enter the self-calibration menu as below Fig.2-2.
Page 3
When the machine show the calibration menu, please press the key “a+, a-, ALU” ordinal, enter to the precision data value input menu. First, enter the DF data value menu, show as below Fig.2-3, please press key “b+ / b-” to increase or decrease the value. Press key a+, switch to I data value menu, show as below fig.2 - 4 , please press key “b+/b - ”...
Page 4
Remove the weight shelf. Remove nuts 1 and 2 with relative cup springs and washers. Back-off screws 3, 4 and 5 then disassemble the various parts. Reassemble the various parts without tightening the nuts being careful to follow the correct sequence. N.B.
Page 5
Picture-3 showing that complete shaft and machine body form an angle of 90 degre, this is the standard assembly; Picture 3-4 showing that complete shaft and machine body form an angle less-than 90 degree,this is incorrect assembly. Please adjust the nut ②and③. Picture 3-5 showing that complete shaft and machine body form an angle greater than 90 degree,this is also incorrect assembly too.
3. Scale disassembly, assembly, adjusting and maintenance method 3.1 Diameter sensor and distance scale sensor Please check if the following parts with red circle marked loose, check if the screw in the red marked loose . 2, please refer to the photo P1, to check if the cores have been random.
Page 7
When you install the Gauge sensor again, please screw to the Clockwise twist to the limit position, then install, as the following photo.
Page 8
Attention when installing the diameter ruler. when installing diameter guage sensor, the handle bar must be kept as this station which stick the complete shaft tightly, as left photo. 3.2 Width scale sensor assembly / disassembly and adjusting method: First step: Disassemble these four screws...
Page 12
Third step: First step, follow Second step, follow as the following direction,and rotate - following direction,and rotate screw dirver to the end screw - dirver back to 90° angle...
Page 13
4. Dismantle of the photoelectric sensor Pls take off the two screws(remark with red circle part),then the user can take off the photoelectric sensor, show as below picture Fig5-1.(Pls do not take off the blue box screws ,due to the accuracy of the machine will be influenced).
Page 14
Cross round screw can adjust height Adjust this length to change the delicacy of safety hood during open - close Magneti Marelli Aftermarket Spółka z.o.o. Plac Pod Lipami 5, 40-476 Katowice Tel.: + 48 (032) 6036107, Faks: + 48 (032) 603-61-08 e-mail: checkstar@magnetimarelli.com...