8.1 Specifications
Meter size: 28", 32", 36", 40", 44", 48", 54", 60", 64", 72", 80", 88", 96", 104", 112", 120"
(700, 800, 900, 1000, 1100, 1200, 1350, 1500, 1600, 1800, 2000, 2200, 2400, 2600,
2800, 3000mm)
Measurement range in terms of flow velocity: 0–1.0 ft/s to 0–32.8 ft/s (0–0.3 m/s to 0–10 m/s)
Accuracy: ± 0.5% of Rate *
* This pulse output error result is established under standard operating conditions at Toshiba's
flow calibration facility, Fuchu Japan.
* Individual meter's measurement error may vary up to ± 0.8 % of Rate at 3.28 ft/s (1.0 m/s)
or more and ± 0.4% of Rate ±0.157 inch/s (4 mm/s) at 3.28 ft/s (1.0m/s) or less.
* Current output: plus ±8μA(0.05% of span.)
* Refer to individual calibration data for each meter's measurement error.
Fluid conductivity: 5 µS/cm minimum
Fluid temperature: 28" to 40" (700 to 1000mm);14 to 140 °F (–10 to +60 °C)
44" (1100mm) or greater;
Ambient temperature: 14 to 140 °F (–10 to +60 °C)
Fluid pressure: -0.1 to the pressure limited by flange standard
Structure: IP 67 and NEMA 4X Watertight
Connection flange standard: AWWA class D
Principal materials:
Case・・・・carbon steel
Lining・・・Chloroprene rubber (std.)
Electrodes・・・316L stainless steel (std.)
Note: See Table 9.2 for optional materials and other related information.
Grounding rings・・・304 stainless steel (std.)
Note: See Table 9.2 for optional materials and other related information.
Measuring tube・・・304 stainless steel
Coating: phthalic acid resin coating (std.), pearl-gray colored
Note: If the optional IP68 and NEMA 6P structure is specified, the coating is
black tar epoxy resin coating 0.5 mm
Option: IP68 and NEMA 6P Submersible type
The coating for this type is black tar epoxy resin coating 0.5 mm.
This type of flowmeter is sub-mersible to 5 m in water.
JIS 10K
14 to 104 °F (–10 to +40 °C)
26