Page 1
QP481 Portable Conductivity Meter – Quick User Guide...
Page 2
If you ordered the QP481 Meter only option the package will only contain the QP481 Portable Conductivity Meter and 4 X AA Batteries. If you ordered the QP481 Portable Conductivity Meter Kit the following items are packed in the case: Carry case, QP481 Meter, 4 x AA Batteries, Conductivity Cell with built in ATC, Calibration Solution, user Guide.
Page 3
9. For further samples repeat from 6 above Using different Calibration Standard solutions: The QP481 comes complete with a 1413uS/cm standard solution. This standard is automatically set up in the menu. To use a 12.88mS/cm standard do the following: 1. After switching the instrument on press the right arrow set up key.
Page 4
4. Select “Enter” to finish. Then scroll up to Continue and press READ. Logging and printing The QP481 Portable Conductivity meter has both a log and a print function. You can print and log directly to an external device when connected via the Mini USB output or straight to file.
Page 5
To transfer this data directly into Excel for printing or further manipulation you will need to install the EDT data Wizard. This is provided in the Series 4 Data Kit (order Seperately)which comprises: USB flash drive with the EDT data Wizard Software including Cable. QP481 Portable Conductivity Meter Automatic Temp.Compensation ATC across the entire range Autoranges Select the correct unit range Automatically.
Page 6
EDT data Wizard is installed. Troubleshooting The QP481 will automatically diagnose and indicate errors or issues on the display. Most errors in conductivity measurement will be as a result of a problem with the Cell and not the meter. Please ensure your cell is in good condition, has been cleaned and that before every measurement it is rinsed in deionised water and blotted dry.
Page 7
Related Products: K=0.1 Conductivity Cell (E8072) Series 4 Portable Conductivity Meter Kit A6000 Conductivity Flow Cell 200µL A5019 Polymer Conductivity Cell K=1 QP458 Combined pH and Conductivity A3052 1413µS/cm Conductivity Meter Standard (100ml)
Need help?
Do you have a question about the QP481 and is the answer not in the manual?
Questions and answers