EDT directION FE257 User Manual

EDT directION FE257 User Manual

Advanced portable ph meter

Advertisement

FE257
- Advanced portable pH meter - Users Manual
Table of Contents
Section 1 - Introduction and Set Up
Section 2 - pH Calibration & Measurement
Section 3 - mV Calibration & Measurement
Section 4 - Use of the Recorder Output
Section 5 - Troubleshooting
Section 6 - Specification
Appendix 1 - pH Electrodes & Buffers

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FE257 and is the answer not in the manual?

Questions and answers

Summary of Contents for EDT directION FE257

  • Page 1 FE257 - Advanced portable pH meter - Users Manual Table of Contents Section 1 - Introduction and Set Up Section 2 - pH Calibration & Measurement Section 3 - mV Calibration & Measurement Section 4 - Use of the Recorder Output...
  • Page 2 SECTION 1 - Introduction Description The FE257 pH meter is a simple-to-use, portable, battery operated instrument. Features include pH measurement with manual or automatic temperature compensation, automatic calibration with standard buffers, and. with a resolution of 0.1mV, can be used with ion selective electrodes.
  • Page 3 6. The meter is ready for calibration A Note on Electrodes pH, Ion Selective and Redox electrodes may all be used with the FE257. These may be combination or half-cell types. Combination or sensing half-cells should have a BNC terminal and be connected to the socket marked pH/ION input at the back of the meter.
  • Page 4 Section 2 - Calibration and Measurement For accurate results, stir all buffers and samples. Always rinse electrodes and blot dry before transferring from one solution to another to prevent contamination. Ensure that any electrode filling holes are left uncovered during use. Automatic Calibration &...
  • Page 5 Manual Temperature Compensation If the use of a temperature probe is inappropriate, e.g. small sample size then manual temperature compensation can be used. 1. Disconnect temperature probe 2. Press the Mode key to select °C function. Note that the °C flag will flash when no temperature probe is connected.
  • Page 6 Section 5 - Troubleshooting Symptom Probable Cause No display Battery is flat or not installed Power supply Disconnected ‘BAT’ flag displayed Battery Low Wildly erratic readings or Display shows -- in Electrodes disconnected left hand digit (Over-range condition) Electrodes not immersed in solution Reference electrode not filled Reference junction dry Drifting readings...
  • Page 7 Appendix 1 - pH electrodes Before Use Remove the protective cap covering the glass sensing bulb and replace with the protective guard if applicable. Inspect the filling solution for air bubbles and remove by shaking in a downward direction. Soak the electrode in pH storage solution for 30 minutes. Cleaning Soak the electrode in 0.1M HCl for 15 minutes followed by soaking in pH storage solution for 30 minutes...
  • Page 8 For more information on our products please visit our website www.edt.co.uk EDT directION Unit 5 Waldershare Park, Dover. CT15 5DQ. Telephone: 01304 829960 Email: sales@edt.co.uk Company No, 04135318 VAT No. GB 765175410...