KERN YKI-01 Installation Manual page 11

Ethernet interface
Table of Contents

Advertisement

The "Local TCP Port" setting is important here. The standard value is 4001. The
server is accessible at this port number. This figure must be entered in the
application to be used. You can change the number to one of your choice.
Confirm the open dialogue box with "OK".
In the "Configuration" dialogue box, go to the "Serial" tab. This will show the
current RS-232 settings of the adapter. These must match the settings of the
connected measuring device.
As before, you can adapt the settings of the adapter by activating the check box
and clicking on "Settings".
The adapter is now set up for operation as a "TCP Server".
Depending on the application, you may now be able to retrieve the measurement data at
the IP address and port number used.
You can run a quick test using a terminal program such as "telnet" (pre-installed up to
Microsoft Windows XP) or "PuTTY":
YKI-01-BA-e-2111
11