Message Frame /Communication Procedure

RAEPoint Host supports Function Code 0x03 only.
a)  Modbus® RTU
0x03 : Read Holding Registers
The Requesting Message:
Starting
Device
Function
Address High
Address
Code
Byte
Client ID 0x03
addr
Note: The max Quantity value is 48.
The aswering message:
Device Address
Function Code
Client ID
0x03
Registers Table:
(1) 0x03(Read Holding Registers)
RAEPoint Host supports up to 8 monitors by following system specification. The entire registers
space is divided into 8 blocks, one block register space corresponds a monitor data registers.
Monitors' basic address is located from 0x0000 to 0x031F (#1 basic address: 0x0000, #2 basic
address: 0x0060, #3 basic address: 0x00C0, #4 basic address: 0x0120, #5 basic address: 0x0180,
#6 basic address: 0x01E0, #7 basic address: 0x0240, #8 basic address: 0x02A0, address
length:0x300), step value is 0x60.